Index: firmware/App/Modes/PreTreatmentRecirc.c =================================================================== diff -u -raec0fd5f5f4ba9845d8bad38a782389fdc42368b -r814dbe3aeff17a111e2b8cf43c8e7711c421d5c1 --- firmware/App/Modes/PreTreatmentRecirc.c (.../PreTreatmentRecirc.c) (revision aec0fd5f5f4ba9845d8bad38a782389fdc42368b) +++ firmware/App/Modes/PreTreatmentRecirc.c (.../PreTreatmentRecirc.c) (revision 814dbe3aeff17a111e2b8cf43c8e7711c421d5c1) @@ -32,7 +32,7 @@ // ********** private definitions ********** #define BLOOD_PUMP_RECIRC_FLOW_RATE 100 ///< Blood pump flow rate during recirculation in mL/min. -#define DIALYSATE_PUMP_RECIRC_FLOW_RATE 100 ///< Dialysate pump flow rate during recirculation in mL/min. +#define DIALYSATE_PUMP_RECIRC_FLOW_RATE 250 ///< Dialysate pump flow rate during recirculation in mL/min. // ********** private data ********** @@ -165,7 +165,7 @@ startAirTrapControl(); setBloodPumpTargetFlowRate( BLOOD_PUMP_RECIRC_FLOW_RATE, MOTOR_DIR_FORWARD, PUMP_CONTROL_MODE_OPEN_LOOP ); - setDialInPumpTargetFlowRate( DIALYSATE_PUMP_RECIRC_FLOW_RATE, MOTOR_DIR_FORWARD, PUMP_CONTROL_MODE_CLOSED_LOOP ); + setDialInPumpTargetFlowRate( DIALYSATE_PUMP_RECIRC_FLOW_RATE, MOTOR_DIR_FORWARD, PUMP_CONTROL_MODE_OPEN_LOOP ); signalDialOutPumpHardStop(); }