Index: firmware/App/Modes/Prime.c =================================================================== diff -u -r06eb4485aab0b6f160610358395d8a568f239ed1 -r69ccdb2725e7aed091392a5b7d8d59afcd7afd33 --- firmware/App/Modes/Prime.c (.../Prime.c) (revision 06eb4485aab0b6f160610358395d8a568f239ed1) +++ firmware/App/Modes/Prime.c (.../Prime.c) (revision 69ccdb2725e7aed091392a5b7d8d59afcd7afd33) @@ -606,7 +606,7 @@ signalBloodPumpHardStop(); setDialInPumpTargetFlowRate( DIALYSATE_PUMP_PRIME_FLOW_RATE_ML_MIN, MOTOR_DIR_FORWARD, PUMP_CONTROL_MODE_OPEN_LOOP ); - setDialOutPumpTargetRate( DIALYSATE_PUMP_PRIME_FLOW_RATE_ML_MIN, MOTOR_DIR_FORWARD, PUMP_CONTROL_MODE_OPEN_LOOP ); + setDialOutPumpTargetRate( 225, MOTOR_DIR_FORWARD, PUMP_CONTROL_MODE_OPEN_LOOP ); // 225 target flow rate matches PWM duty cycle for DPi pump @ 300. // Calculate the time out value that must passed prior to checking for the steady state volume in the reservoir primeDialysateDialyzerTimeLimit = (U32)( ( ( DIALYSATE_DIALYZER_TUBE_VOLUME_ML + dialyzerDialysateVolume ) * SEC_PER_MIN * MS_PER_SECOND ) / DIALYSATE_PUMP_PRIME_FLOW_RATE_ML_MIN );