Index: firmware/App/Modes/Prime.c =================================================================== diff -u -r12708a04ae3365e11180182fc52d005c229a7bc6 -r1b2b351dc93462a6a0e1e4e80e7ea923c65ab892 --- firmware/App/Modes/Prime.c (.../Prime.c) (revision 12708a04ae3365e11180182fc52d005c229a7bc6) +++ firmware/App/Modes/Prime.c (.../Prime.c) (revision 1b2b351dc93462a6a0e1e4e80e7ea923c65ab892) @@ -61,7 +61,8 @@ #define LOAD_CELL_STEADY_VOLUME_SAMPLING_TIME ( 1 * MS_PER_SECOND ) ///< Time load cell reading steady state detection sampling time in seconds. #define PRIME_DIALYSATE_BYPASS_TIME_LIMIT ( 8 * MS_PER_SECOND ) ///< Time limit for priming dialysate bypass circuit. #define STEADY_VOLUME_COUNT_SEC ( 10000 / LOAD_CELL_STEADY_VOLUME_SAMPLING_TIME ) ///< Counter must be greater than 10 seconds before steady volume is true. -#define STEADY_VOLUME_TIME_DEADLINE_MS ( 55 * MS_PER_SECOND ) ///< Time in msec for the steady volume deadline time out. +#define STEADY_VOLUME_DIALYSATE_PRIME_TIME_LIMIT_MS ( 55 * MS_PER_SECOND ) ///< Time in msec for the reservoir volume to stabilize during dialysate prime state. +#define STEADY_VOLUME_BYPASS_PRIME_TIME_LIMIT_MS ( 25 * MS_PER_SECOND ) ///< Time in msec for the reservoir volume to stabilize during bypass dialysate prime state. #define VENOUS_PRESSURE_BUBBLE_CLEAR_MAX_MMHG ( 200.0F ) ///< Maximum venous pressure reading (in mmHg) for bubble clear. #define BUBBLE_CLEAR_WAIT_TIME_INITIAL_MS ( 10 * MS_PER_SECOND ) ///< Time in msec to wait for initial bubble clear pressure. @@ -778,7 +779,7 @@ steadyVolumeSamplingStartTime = getMSTimerCount(); } - if ( TRUE == didTimeout( primeDialysateDialyzerStartTime, primeDialysateDialyzerTimeLimit + STEADY_VOLUME_TIME_DEADLINE_MS ) ) + if ( TRUE == didTimeout( primeDialysateDialyzerStartTime, primeDialysateDialyzerTimeLimit + STEADY_VOLUME_DIALYSATE_PRIME_TIME_LIMIT_MS ) ) { SET_ALARM_WITH_1_U32_DATA( ALARM_ID_HD_PRIME_DIALYSATE_DIALYZER_TIME_OUT, primeDialysateDialyzerTimeLimit ); } @@ -942,7 +943,7 @@ steadyVolumeSamplingStartTime = getMSTimerCount(); } - if ( TRUE == didTimeout( primeDialysateBypassStartTime, PRIME_DIALYSATE_BYPASS_TIME_LIMIT + STEADY_VOLUME_TIME_DEADLINE_MS ) ) // allocate 55 seconds for the steady state logic to do its job before timing out + if ( TRUE == didTimeout( primeDialysateBypassStartTime, PRIME_DIALYSATE_BYPASS_TIME_LIMIT + STEADY_VOLUME_BYPASS_PRIME_TIME_LIMIT_MS ) ) // allocate 25 seconds for the steady state logic to do its job before timing out { SET_ALARM_WITH_1_U32_DATA( ALARM_ID_HD_PRIME_DIALYSATE_BYPASS_TIME_OUT, PRIME_DIALYSATE_BYPASS_TIME_LIMIT ); }