Index: firmware/App/Modes/Prime.c =================================================================== diff -u -r19a8bf98a7154e24c35da25225d4b55bf70ddd09 -r9bf209fc5813b7e806f51f408ece77174a336c9b --- firmware/App/Modes/Prime.c (.../Prime.c) (revision 19a8bf98a7154e24c35da25225d4b55bf70ddd09) +++ firmware/App/Modes/Prime.c (.../Prime.c) (revision 9bf209fc5813b7e806f51f408ece77174a336c9b) @@ -166,6 +166,8 @@ setAlarmUserActionEnabled( ALARM_USER_ACTION_RINSEBACK, FALSE ); setAlarmUserActionEnabled( ALARM_USER_ACTION_END_TREATMENT, TRUE ); + doorClosedRequired( TRUE, TRUE ); + // Pumps should be off signalBloodPumpHardStop(); signalDialInPumpHardStop(); @@ -594,7 +596,7 @@ // Calculate the time out value that must passed prior to checking for the steady state volume in the reservoir primeDialysateDialyzerTimeLimit = (U32)( ( ( DIALYSATE_DIALYZER_TUBE_VOLUME_ML + dialyzerDialysateVolume ) * SEC_PER_MIN * MS_PER_SECOND ) / DIALYSATE_PUMP_PRIME_FLOW_RATE_ML_MIN ); - minimumReservoirVolume = getLoadCellWeight( LOAD_CELL_RESERVOIR_2_PRIMARY ); + minimumReservoirVolume = getLoadCellWeight( LOAD_CELL_RESERVOIR_1_PRIMARY ); steadyVolumeCount = 0; steadyVolumeSamplingStartTime = getMSTimerCount(); primeDialysateDialyzerStartTime = getMSTimerCount();