Index: firmware/App/Modes/Prime.c =================================================================== diff -u -r2638aed3d414e7b14f309bdf05fa57d4141eb948 -r66da0a331c118550341c81098447d9fb8b7237e9 --- firmware/App/Modes/Prime.c (.../Prime.c) (revision 2638aed3d414e7b14f309bdf05fa57d4141eb948) +++ firmware/App/Modes/Prime.c (.../Prime.c) (revision 66da0a331c118550341c81098447d9fb8b7237e9) @@ -45,7 +45,7 @@ #define DIALYSATE_PUMP_PRIME_FLOW_RATE_ML_MIN 300 ///< Dialysate pump flow rate during priming fluid path. #define LOAD_CELL_VOLUME_NOISE_TOLERANCE 0.05 ///< Allow 5% tolerance on load cell readings. -#define PRIME_SALINE_DIALYZER_STOP_WEIGHT 200.0 // 10.0 try 200.0 11/19/2021 Sean ///< Load cell readings of 10 grams ( 10 mL ) used in prime saline dialyzer state +#define PRIME_SALINE_DIALYZER_STOP_WEIGHT 50.0 // 10.0 try 200.0 11/19/2021 Sean ///< Load cell readings of 10 grams ( 10 mL ) used in prime saline dialyzer state #define NO_AIR_DETECTED_COUNT ( 20 * MS_PER_SECOND ) ///< No air detected time period count. #define PURGE_AIR_TIME_OUT_COUNT ( 60 * MS_PER_SECOND ) ///< Time period count for purge air time out. @@ -829,6 +829,11 @@ state = HD_PRIME_SALINE_SETUP_STATE; break; + case HD_PRIME_SALINE_DIALYZER_SETUP_STATE: + case HD_PRIME_SALINE_DIALYZER_STATE: + state = HD_PRIME_SALINE_DIALYZER_SETUP_STATE; + break; + case HD_PRIME_RESERVOIR_ONE_FILL_COMPLETE_STATE: case HD_PRIME_DIALYSATE_DIALYZER_STATE: state = HD_PRIME_RESERVOIR_ONE_FILL_COMPLETE_STATE;