Index: firmware/App/Modes/Prime.c =================================================================== diff -u -r75e1ae332d1446dddf9b8d4ce6e8317449c57d67 -r364631fbb6ffd3ab52271e1cc5b71ee9e2cc777f --- firmware/App/Modes/Prime.c (.../Prime.c) (revision 75e1ae332d1446dddf9b8d4ce6e8317449c57d67) +++ firmware/App/Modes/Prime.c (.../Prime.c) (revision 364631fbb6ffd3ab52271e1cc5b71ee9e2cc777f) @@ -46,6 +46,7 @@ #define BLOOD_PUMP_FAST_FLOW_RATE_CIRC_BLOOD_CIRCUIT_ML_MIN 300 ///< Blood pump fast flow rate during prime recirculate blood circuit state. #define BLOOD_PUMP_FLOW_RATE_SALINE_DIALYZER_ML_MIN 300 ///< Blood pump flow rate during prime the saline dialyzer dialysate state. #define DIALYSATE_PUMP_PRIME_FLOW_RATE_ML_MIN 300 ///< Dialysate pump flow rate during priming fluid path. +#define DIALYSATE_PUMP_FAST_PRIME_FLOW_RATE_ML_MIN 600 ///< Dialysate pump flow rate during priming fluid path. #define DPO_PUMP_PRIME_FLOW_RATE_ML_MIN 225 ///< Dialysate outlet pump flow rate needed to match PWM with dialysate inlet pump @ 300 mL/min in open loop mode. #define DIALYSATE_DIALYZER_TUBE_VOLUME_ML 115 ///< This total tube volume is used to calculate the Dpi & Dpo time out in the dialysate dialyzer state. @@ -877,15 +878,15 @@ { DG_SWITCH_RSRVRS_CMD_T rsrvrCmd; - rsrvrCmd.reservoirID = (U32)DG_RESERVOIR_1; + rsrvrCmd.reservoirID = (U32)DG_RESERVOIR_2; rsrvrCmd.useLastTrimmerHeaterDC = FALSE; cmdSetDGActiveReservoir( &rsrvrCmd ); if ( TRUE == hasDGCompletedReservoirSwitch() ) { signalBloodPumpHardStop(); - setDialInPumpTargetFlowRate( DIALYSATE_PUMP_PRIME_FLOW_RATE_ML_MIN, MOTOR_DIR_FORWARD, PUMP_CONTROL_MODE_OPEN_LOOP ); + setDialInPumpTargetFlowRate( DIALYSATE_PUMP_FAST_PRIME_FLOW_RATE_ML_MIN, MOTOR_DIR_FORWARD, PUMP_CONTROL_MODE_OPEN_LOOP ); setValvePosition( VDI, VALVE_POSITION_C_CLOSE ); setValvePosition( VDO, VALVE_POSITION_C_CLOSE );