Index: firmware/App/Modes/Prime.c =================================================================== diff -u -raceb6e5388e905d119c5c379bc97ced0720aa270 -r72e4c11140e7a0d186c3a6e58ef724b7e126bab8 --- firmware/App/Modes/Prime.c (.../Prime.c) (revision aceb6e5388e905d119c5c379bc97ced0720aa270) +++ firmware/App/Modes/Prime.c (.../Prime.c) (revision 72e4c11140e7a0d186c3a6e58ef724b7e126bab8) @@ -53,6 +53,7 @@ #define NO_AIR_DETECTED_COUNT ( 20 * MS_PER_SECOND ) ///< No air detected time period count. #define PURGE_AIR_TIME_OUT_COUNT ( 60 * MS_PER_SECOND ) ///< Time period count for purge air time out. +#define PRIME_SALINE_DIALYZER_TIME_OUT_COUNT ( 60 * MS_PER_SECOND ) ///< Time period count for prime saline dialyzer time out. #define LOAD_CELL_STEADY_VOLUME_SAMPLING_TIME ( 1 * MS_PER_SECOND ) ///< Time load cell reading steady state detection sampling time in seconds. #define PRIME_DIALYSATE_BYPASS_TIME_LIMIT ( 15 * MS_PER_SECOND ) ///< Time limit for priming dialysate bypass circuit. #define STEADY_VOLUME_COUNT_SEC ( 10000 / LOAD_CELL_STEADY_VOLUME_SAMPLING_TIME ) ///< Counter must be greater than 10 seconds before steady volume is true. @@ -500,7 +501,7 @@ if ( TRUE == didTimeout( purgeAirTimeOutStartTime, PURGE_AIR_TIME_OUT_COUNT ) ) { - activateAlarmNoData( ALARM_ID_HD_PRIME_PURGE_AIR_TIME_OUT ); + activateAlarmNoData( ALARM_ID_HD_PRIME_SALINE_PURGE_AIR_TIME_OUT ); } if ( AIR_TRAP_LEVEL_FLUID == getAirTrapLevel( AIR_TRAP_LEVEL_SENSOR_LOWER ) ) @@ -732,6 +733,11 @@ state = HD_PRIME_RESERVOIR_TWO_FILL_COMPLETE_STATE; } + if ( TRUE == didTimeout( primeSalineDialyzerStartTime, PRIME_SALINE_DIALYZER_TIME_OUT_COUNT ) ) + { + activateAlarmNoData( ALARM_ID_PRIME_SALINE_DIALYZER_TIME_OUT ); // Trigger HD prime saline dialyzer time out alarm. + } + if ( TRUE == doesAlarmStatusIndicateStop() ) { setupForPrimePause();