Index: firmware/App/Modes/Prime.c =================================================================== diff -u -rb2e2a47fef76897387c630a2508b6bc1521fff34 -rb1f086e7cd292d5a97a7265075400274d60d4fbf --- firmware/App/Modes/Prime.c (.../Prime.c) (revision b2e2a47fef76897387c630a2508b6bc1521fff34) +++ firmware/App/Modes/Prime.c (.../Prime.c) (revision b1f086e7cd292d5a97a7265075400274d60d4fbf) @@ -166,6 +166,8 @@ setAlarmUserActionEnabled( ALARM_USER_ACTION_RINSEBACK, FALSE ); setAlarmUserActionEnabled( ALARM_USER_ACTION_END_TREATMENT, TRUE ); + doorClosedRequired( TRUE, TRUE ); + // Pumps should be off signalBloodPumpHardStop(); signalDialInPumpHardStop(); @@ -594,7 +596,7 @@ // Calculate the time out value that must passed prior to checking for the steady state volume in the reservoir primeDialysateDialyzerTimeLimit = (U32)( ( ( DIALYSATE_DIALYZER_TUBE_VOLUME_ML + dialyzerDialysateVolume ) * SEC_PER_MIN * MS_PER_SECOND ) / DIALYSATE_PUMP_PRIME_FLOW_RATE_ML_MIN ); - minimumReservoirVolume = getLoadCellWeight( LOAD_CELL_RESERVOIR_2_PRIMARY ); + minimumReservoirVolume = getLoadCellWeight( LOAD_CELL_RESERVOIR_1_PRIMARY ); steadyVolumeCount = 0; steadyVolumeSamplingStartTime = getMSTimerCount(); primeDialysateDialyzerStartTime = getMSTimerCount();