Index: firmware/App/Modes/Prime.c =================================================================== diff -u -re54806da466a8c52e19fd272356dbbfde04171fd -re654b368ff8f4d194678db88f1cda57b04e7d947 --- firmware/App/Modes/Prime.c (.../Prime.c) (revision e54806da466a8c52e19fd272356dbbfde04171fd) +++ firmware/App/Modes/Prime.c (.../Prime.c) (revision e654b368ff8f4d194678db88f1cda57b04e7d947) @@ -500,7 +500,7 @@ if ( TRUE == didTimeout( purgeAirTimeOutStartTime, PURGE_AIR_TIME_OUT_COUNT ) ) { - activateAlarmNoData( ALARM_ID_HD_PRIME_SALINE_PURGE_AIR_TIME_OUT ); + SET_ALARM_WITH_1_U32_DATA( ALARM_ID_HD_PRIME_SALINE_PURGE_AIR_TIME_OUT, PURGE_AIR_TIME_OUT_COUNT ); } if ( AIR_TRAP_LEVEL_FLUID == getAirTrapLevel( AIR_TRAP_LEVEL_SENSOR_LOWER ) ) @@ -734,7 +734,7 @@ if ( TRUE == didTimeout( primeSalineDialyzerStartTime, PRIME_SALINE_DIALYZER_TIME_OUT_COUNT ) ) { - activateAlarmNoData( ALARM_ID_PRIME_SALINE_DIALYZER_TIME_OUT ); // Trigger HD prime saline dialyzer time out alarm. + SET_ALARM_WITH_1_U32_DATA( ALARM_ID_PRIME_SALINE_DIALYZER_TIME_OUT, PRIME_SALINE_DIALYZER_TIME_OUT_COUNT ); // Trigger HD prime saline dialyzer time out alarm. } if ( TRUE == doesAlarmStatusIndicateStop() )