Index: firmware/App/Modes/Rinseback.c =================================================================== diff -u -r3fa496b611a3a84c8aaa882a1b27762c6bc0024c -r0b36ae4b591ce21b1971e00a5c2102d68f68d0a6 --- firmware/App/Modes/Rinseback.c (.../Rinseback.c) (revision 3fa496b611a3a84c8aaa882a1b27762c6bc0024c) +++ firmware/App/Modes/Rinseback.c (.../Rinseback.c) (revision 0b36ae4b591ce21b1971e00a5c2102d68f68d0a6) @@ -416,7 +416,7 @@ result = RINSEBACK_STOP_STATE; } // Check for empty saline bag - if ( TRUE == isSalineBagEmpty() ) + else if ( TRUE == isSalineBagEmpty() ) { SET_ALARM_WITH_1_F32_DATA( ALARM_ID_HD_EMPTY_SALINE_BAG, getMeasuredArterialPressure() ); setupForRinsebackStopOrPause();