Index: firmware/App/Modes/SelfTests.c =================================================================== diff -u -r7c94ce646c60b4052d9d6c8d212cfa1096bfece9 -rc851d640df6889c1ac5b8b1a66340376463d0702 --- firmware/App/Modes/SelfTests.c (.../SelfTests.c) (revision 7c94ce646c60b4052d9d6c8d212cfa1096bfece9) +++ firmware/App/Modes/SelfTests.c (.../SelfTests.c) (revision c851d640df6889c1ac5b8b1a66340376463d0702) @@ -74,7 +74,8 @@ #define WET_SELF_TEST_INTEGRATED_VOLUME_PCT_TOLERANCE 0.05F ///< Tolerance on integrated volume as a percentage (5%). #define WET_SELF_TEST_INTEGRATED_VOLUME_TOLERANCE 20.0F ///< Tolerance on integrated volume in grams. #define WET_SELF_TEST_DISPLACEMENT_TOLERANCE_G 20.0F ///< Tolerance in the load cell readings of the displacement in grams. -#define WET_SELF_TEST_DISPLACEMENT_TIME_MS ( SEC_PER_MIN * MS_PER_SECOND ) ///< Time to displace dialysate in wet self-test in ms. +#define WET_SELF_TEST_FIRST_DISPLACEMENT_TIME_MS ( SEC_PER_MIN * MS_PER_SECOND ) ///< Time to displace dialysate in wet self-test in ms (first). +#define WET_SELF_TEST_SECOND_DISPLACEMENT_TIME_MS ( 30 * MS_PER_SECOND ) ///< Time to displace dialysate in wet self-test in ms (second). #define RESERVOIR_SETTLE_TIME_MS ( 5 * MS_PER_SECOND ) ///< Time allotted for reservoir to settle in ms. #define MAX_NO_CARTRIDGE_SELF_TEST_TIME 30 ///< Maximum no cartridge self-test time (in seconds). @@ -360,7 +361,7 @@ } else { - SET_ALARM_WITH_1_U32_DATA( ALARM_ID_HD_NO_CART_SELF_TEST_TIMEOUT, currentNoCartSelfTestsState ); +// SET_ALARM_WITH_1_U32_DATA( ALARM_ID_HD_NO_CART_SELF_TEST_TIMEOUT, currentNoCartSelfTestsState ); } } } @@ -592,7 +593,8 @@ WET_SELF_TESTS_STATE_T priorSubState = currentWetSelfTestsState; #ifndef _RELEASE_ - if ( getSoftwareConfigStatus( SW_CONFIG_DISABLE_WET_SELF_TEST ) == SW_CONFIG_ENABLE_VALUE ) + if ( ( SW_CONFIG_ENABLE_VALUE == getSoftwareConfigStatus( SW_CONFIG_DISABLE_PRIMING ) ) || + ( SW_CONFIG_ENABLE_VALUE == getSoftwareConfigStatus( SW_CONFIG_DISABLE_WET_SELF_TEST ) ) ) { currentWetSelfTestsState = WET_SELF_TESTS_COMPLETE_STATE; } @@ -1043,7 +1045,7 @@ else { #ifndef _RELEASE_ - if ( SW_CONFIG_DISABLE_VALUE != getSoftwareConfigStatus( SW_CONFIG_DISABLE_AIR_PUMP ) ) + if ( SW_CONFIG_ENABLE_VALUE == getSoftwareConfigStatus( SW_CONFIG_DISABLE_AIR_PUMP ) ) { // Wait for reservoirs to drain before starting this test @@ -1755,10 +1757,10 @@ // This is not a volume for now and it is a flow rate. fmdIntegratedVolume += getMeasuredDialInFlowRate(); - if ( TRUE == didTimeout( displacementStartTime, WET_SELF_TEST_DISPLACEMENT_TIME_MS ) ) + if ( TRUE == didTimeout( displacementStartTime, WET_SELF_TEST_FIRST_DISPLACEMENT_TIME_MS ) ) { signalDialInPumpHardStop(); - fmdIntegratedVolume = ( ( fmdIntegratedVolume * TASK_GENERAL_INTERVAL ) / ( WET_SELF_TEST_DISPLACEMENT_TIME_MS ) ); + fmdIntegratedVolume = ( ( fmdIntegratedVolume * TASK_GENERAL_INTERVAL ) / ( WET_SELF_TEST_FIRST_DISPLACEMENT_TIME_MS ) ); settleStartTime = getMSTimerCount(); state = WET_SELF_TESTS_FIRST_DISPLACEMENT_VERIFY_STATE; } @@ -1897,10 +1899,10 @@ { fmdIntegratedVolume += getMeasuredDialInFlowRate(); - if ( TRUE == didTimeout( displacementStartTime, WET_SELF_TEST_DISPLACEMENT_TIME_MS ) ) + if ( TRUE == didTimeout( displacementStartTime, WET_SELF_TEST_SECOND_DISPLACEMENT_TIME_MS ) ) { signalDialInPumpHardStop(); - fmdIntegratedVolume = ( ( fmdIntegratedVolume * TASK_GENERAL_INTERVAL ) / ( WET_SELF_TEST_DISPLACEMENT_TIME_MS ) ); + fmdIntegratedVolume = ( ( fmdIntegratedVolume * TASK_GENERAL_INTERVAL ) / ( WET_SELF_TEST_SECOND_DISPLACEMENT_TIME_MS ) ); settleStartTime = getMSTimerCount(); state = WET_SELF_TESTS_SECOND_DISPLACEMENT_VERIFY_STATE; }