Index: firmware/App/Modes/SelfTests.c =================================================================== diff -u -rb156b12139fea40401c94b512503a2ceced69e22 -r533dc50613389ef694c7bcdf3edc0d9c0c44e14c --- firmware/App/Modes/SelfTests.c (.../SelfTests.c) (revision b156b12139fea40401c94b512503a2ceced69e22) +++ firmware/App/Modes/SelfTests.c (.../SelfTests.c) (revision 533dc50613389ef694c7bcdf3edc0d9c0c44e14c) @@ -221,6 +221,7 @@ signalBloodPumpHardStop(); signalDialInPumpHardStop(); signalDialOutPumpHardStop(); + cmdStopDGTrimmerHeater(); if ( TRUE == isAlarmActive( ALARM_ID_HD_SYRINGE_PUMP_NOT_ENOUGH_HEPARIN_ALARM ) ) { @@ -1782,6 +1783,7 @@ if ( DG_CMD_REQUEST_REJECT_REASON_NONE == cmdResp.rejectCode ) { setDialInPumpTargetFlowRate( DIP_FLOW_RATE_FIRST_DISPLACEMENT_ML_MIN, MOTOR_DIR_FORWARD, PUMP_CONTROL_MODE_OPEN_LOOP ); + cmdStartDGTrimmerHeater(); settleStartTime = getMSTimerCount(); state = WET_SELF_TESTS_FIRST_DISPLACEMENT_STATE; } @@ -1816,6 +1818,7 @@ if ( TRUE == didTimeout( displacementStartTime, WET_SELF_TEST_FIRST_DISPLACEMENT_TIME_MS ) ) { signalDialInPumpHardStop(); + cmdStopDGTrimmerHeater(); settleStartTime = getMSTimerCount(); state = WET_SELF_TESTS_FIRST_DISPLACEMENT_VERIFY_STATE; } @@ -1925,6 +1928,7 @@ if ( DG_CMD_REQUEST_REJECT_REASON_NONE == cmdResp.rejectCode ) { setDialInPumpTargetFlowRate( DIP_FLOW_RATE_SECOND_DISPLACEMENT_ML_MIN, MOTOR_DIR_FORWARD, PUMP_CONTROL_MODE_OPEN_LOOP ); + cmdStartDGTrimmerHeater(); settleStartTime = getMSTimerCount(); state = WET_SELF_TESTS_SECOND_DISPLACEMENT_STATE; } @@ -1967,6 +1971,7 @@ if ( TRUE == didTimeout( displacementStartTime, WET_SELF_TEST_SECOND_DISPLACEMENT_TIME_MS ) ) { signalDialInPumpHardStop(); + cmdStopDGTrimmerHeater(); settleStartTime = getMSTimerCount(); state = WET_SELF_TESTS_SECOND_DISPLACEMENT_VERIFY_STATE; }