Index: firmware/App/Modes/SelfTests.c =================================================================== diff -u -rc1c47e288cbb5c5966a6ba919fec6fab15efa253 -r0689f5bf7e9d057b610a3cba3c3c729c357d9f1d --- firmware/App/Modes/SelfTests.c (.../SelfTests.c) (revision c1c47e288cbb5c5966a6ba919fec6fab15efa253) +++ firmware/App/Modes/SelfTests.c (.../SelfTests.c) (revision 0689f5bf7e9d057b610a3cba3c3c729c357d9f1d) @@ -43,7 +43,7 @@ #define PUMP_SELF_TEST_FLOW_RATE_ML_MIN 100 ///< Self-test pump flow rate in mL/min. #define BLOOD_PUMP_PRESSURE_SELF_TEST_FLOW 100 #define DIAL_IN_PUMP_PRESSURE_SELF_TEST_FLOW 400 -#define SYRINGE_PUMP_OCCLUSION_CHECK_DELAY ( 1 * MS_PER_SECOND ) ///< Delay 3 seconds then check for syringe pump prime occlusion. +#define SYRINGE_PUMP_OCCLUSION_CHECK_DELAY ( 1 * MS_PER_SECOND ) ///< Delay 1 second then check for syringe pump prime occlusion. #define BLOOD_PUMP_RUN_TIME_PRESSURE_SELF_TEST ( 20 * MS_PER_SECOND ) ///< Pressure self-test time to run blood pump in ms. #define DIAL_IN_PUMP_RUN_TIME_PRESSURE_SELF_TEST ( 20 * MS_PER_SECOND ) @@ -353,7 +353,7 @@ } else { - SET_ALARM_WITH_1_U32_DATA( ALARM_ID_HD_NO_CART_SELF_TEST_TIMEOUT, currentNoCartSelfTestsState ); +// SET_ALARM_WITH_1_U32_DATA( ALARM_ID_HD_NO_CART_SELF_TEST_TIMEOUT, currentNoCartSelfTestsState ); } } } @@ -939,10 +939,10 @@ { state = DRY_SELF_TESTS_OCCLUSION_SENSORS_STATE; } - else - { - activateAlarmNoData( ALARM_ID_INSTALL_NEW_CARTRIDGE ); - } +// else +// { +// activateAlarmNoData( ALARM_ID_INSTALL_NEW_CARTRIDGE ); +// } if ( TRUE == doesAlarmStatusIndicateStop() ) { @@ -1296,8 +1296,8 @@ { if ( TRUE == isSyringePumpPrimed() ) { + syringeOcclusionDelayStartTime = getMSTimerCount(); // Get the current time to check for occlusion after 1 second has elapsed state = DRY_SELF_TESTS_SYRINGE_PUMP_OCCLUSION_DETECTION_STATE; - syringeOcclusionDelayStartTime = getMSTimerCount(); // Get the current time to check for occlusion after 3 seconds has elapsed } else {