Index: firmware/App/Modes/SelfTests.c =================================================================== diff -u -rc1e2dfcd544f59788ad1a58e1e8339a475ba4266 -r2725d0a9743f3209581e5e1098258c43c9bf20b6 --- firmware/App/Modes/SelfTests.c (.../SelfTests.c) (revision c1e2dfcd544f59788ad1a58e1e8339a475ba4266) +++ firmware/App/Modes/SelfTests.c (.../SelfTests.c) (revision 2725d0a9743f3209581e5e1098258c43c9bf20b6) @@ -946,7 +946,6 @@ state = DRY_SELF_TESTS_WAIT_FOR_DOOR_CLOSE_STATE; } } - else { // doorClosedRequired() will trigger alarms if pumpTrack is open @@ -1062,7 +1061,7 @@ pressureSelfTestPreNormalStartTime = getMSTimerCount(); state = DRY_SELF_TESTS_PRESSURE_SENSORS_NORMAL_SETUP_STATE; - if ( TRUE == getTestConfigStatus( TEST_CONFIG_USE_WORN_CARTRIDGE ) ) + if ( TRUE == getTestConfigStatus( TEST_CONFIG_USE_WET_CARTRIDGE ) ) { dryPressureTestsCompleted = TRUE; state = DRY_SELF_TESTS_SYRINGE_PUMP_PRIME_STATE;