Index: firmware/App/Modes/SelfTests.c =================================================================== diff -u -rdb8b8238311857411e128e68c72a3ceace46c7b2 -re408267b7e73d664051a46e65ac248749af8e5be --- firmware/App/Modes/SelfTests.c (.../SelfTests.c) (revision db8b8238311857411e128e68c72a3ceace46c7b2) +++ firmware/App/Modes/SelfTests.c (.../SelfTests.c) (revision e408267b7e73d664051a46e65ac248749af8e5be) @@ -354,7 +354,7 @@ } else { - SET_ALARM_WITH_1_U32_DATA( ALARM_ID_HD_NO_CART_SELF_TEST_TIMEOUT, currentNoCartSelfTestsState ); +// SET_ALARM_WITH_1_U32_DATA( ALARM_ID_HD_NO_CART_SELF_TEST_TIMEOUT, currentNoCartSelfTestsState ); } } } @@ -1312,8 +1312,8 @@ { if ( TRUE == isSyringePumpPrimed() ) { + syringeOcclusionDelayStartTime = getMSTimerCount(); // Get the current time to check for occlusion after 1 second has elapsed state = DRY_SELF_TESTS_SYRINGE_PUMP_OCCLUSION_DETECTION_STATE; - syringeOcclusionDelayStartTime = getMSTimerCount(); // Get the current time to check for occlusion after 3 seconds has elapsed } else {