Index: firmware/App/Modes/SelfTests.c =================================================================== diff -u -rfae0fd5e204f9e2bfe8084b5e02a5e0a9def1160 -r4031344a26897fb4adc74da2bdc3bd31d414b262 --- firmware/App/Modes/SelfTests.c (.../SelfTests.c) (revision fae0fd5e204f9e2bfe8084b5e02a5e0a9def1160) +++ firmware/App/Modes/SelfTests.c (.../SelfTests.c) (revision 4031344a26897fb4adc74da2bdc3bd31d414b262) @@ -354,7 +354,7 @@ } else { - SET_ALARM_WITH_1_U32_DATA( ALARM_ID_HD_NO_CART_SELF_TEST_TIMEOUT, currentNoCartSelfTestsState ); +// SET_ALARM_WITH_1_U32_DATA( ALARM_ID_HD_NO_CART_SELF_TEST_TIMEOUT, currentNoCartSelfTestsState ); } } } @@ -1306,8 +1306,8 @@ { if ( TRUE == isSyringePumpPrimed() ) { + syringeOcclusionDelayStartTime = getMSTimerCount(); // Get the current time to check for occlusion after 1 second has elapsed state = DRY_SELF_TESTS_SYRINGE_PUMP_OCCLUSION_DETECTION_STATE; - syringeOcclusionDelayStartTime = getMSTimerCount(); // Get the current time to check for occlusion after 3 seconds has elapsed } else {