Index: firmware/App/Services/AlarmMgmt.c =================================================================== diff -u -r637c80885d1bd28e84ed16b7526793b0baa0f38d -re12fd5b499b9d32bd21b43cd4d178dd0241aa1e3 --- firmware/App/Services/AlarmMgmt.c (.../AlarmMgmt.c) (revision 637c80885d1bd28e84ed16b7526793b0baa0f38d) +++ firmware/App/Services/AlarmMgmt.c (.../AlarmMgmt.c) (revision e12fd5b499b9d32bd21b43cd4d178dd0241aa1e3) @@ -1391,8 +1391,7 @@ } // If in Treatment-Stop state or Fault/Service/Standby Mode, allow user to minimize the alarm window - if ( ( MODE_FAUL == currentMode ) || ( MODE_SERV == currentMode ) || ( MODE_STAN == currentMode ) || - ( ( MODE_TREA == currentMode ) && ( TREATMENT_STOP_STATE == getTreatmentState() ) ) ) + if ( ( MODE_FAUL == currentMode ) || ( MODE_SERV == currentMode ) || ( MODE_STAN == currentMode ) ) { noMinimize = FALSE; }