Index: firmware/App/Services/AlarmMgmt.c =================================================================== diff -u -r82327a32b9c299351ba3ab344f3b1f9116a204f8 -red0444d18e586808fd59cf6fc118f8885eebfe22 --- firmware/App/Services/AlarmMgmt.c (.../AlarmMgmt.c) (revision 82327a32b9c299351ba3ab344f3b1f9116a204f8) +++ firmware/App/Services/AlarmMgmt.c (.../AlarmMgmt.c) (revision ed0444d18e586808fd59cf6fc118f8885eebfe22) @@ -7,8 +7,8 @@ * * @file AlarmMgmt.c * -* @author (last) Dara Navaei -* @date (last) 05-Mar-2024 +* @author (last) Vinayakam Mani +* @date (last) 01-Apr-2024 * * @author (original) Sean Nash * @date (original) 07-Nov-2019 @@ -409,7 +409,8 @@ // if the block timer is 0 OR we have an unblockable alarm if ( ( ALARM_NOT_BLOCKED == alarmsBlockedTimer ) || ( ALARM_ID_HD_AC_POWER_LOST == alarm ) - || ( ALARM_ID_HD_AC_POWER_LOST_IN_TREATMENT == alarm ) ) + || ( ALARM_ID_HD_AC_POWER_LOST_IN_TREATMENT == alarm ) + || ( ALARM_ID_HD_DG_RESTARTED_FAULT == alarm ) ) { // do not re-trigger alarm if blocked by property if ( ( FALSE == alarmNoRetrigger ) || ( ( ALARM_TABLE[ alarm ].alarmNoRetrigOnRB != TRUE ) && ( ALARM_TABLE[ alarm ].alarmNoRetrigOnEndTx != TRUE ) ) )