Index: vectorcast/Hercules_RM46_HD_Project/environment/MODETREATMENT/MODETREATMENT.tst =================================================================== diff -u -r88dd475deace8ca8c34d131ad8c594b090daabfc -r04c4e781429058e257880a573abbb7531bb4664e --- vectorcast/Hercules_RM46_HD_Project/environment/MODETREATMENT/MODETREATMENT.tst (.../MODETREATMENT.tst) (revision 88dd475deace8ca8c34d131ad8c594b090daabfc) +++ vectorcast/Hercules_RM46_HD_Project/environment/MODETREATMENT/MODETREATMENT.tst (.../MODETREATMENT.tst) (revision 04c4e781429058e257880a573abbb7531bb4664e) @@ -2826,462 +2826,158 @@ -- Subprogram: verifyBloodAndDialysateRateSettingsChange --- Test Case: verifyBloodAndDialysateRateSettingsChange_Accepted +-- Test Case: verifyBloodAndDialysateRateSettingsChange_01_not_in_Tx TEST.UNIT:ModeTreatment TEST.SUBPROGRAM:verifyBloodAndDialysateRateSettingsChange TEST.NEW -TEST.NAME:verifyBloodAndDialysateRateSettingsChange_Accepted -TEST.IMPORT_FAILURES: -(E) Errors from previous script import(s) - >>> (E) @LINE: 2862 TEST.EXPECTED:uut_prototype_stubs.setDialysisParams.maxUFVol:2400.0 - >>> >>> Could not find function setDialysisParams - >>> >>> in unit uut_prototype_stubs. - >>> >>> Value Line Error - Command Ignored - >>> (E) @LINE: 2863 TEST.EXPECTED:uut_prototype_stubs.setDialysisParams.uFRate:10.0 - >>> >>> Could not find function setDialysisParams - >>> >>> in unit uut_prototype_stubs. - >>> >>> Value Line Error - Command Ignored - >>> (E) @LINE: 764 TEST.VALUE:ModeTreatment.<>.presBloodFlowRate:200 - >>> >>> Unknown parameter/object name presBloodFlowRate - >>> >>> Value Line Error - Command Ignored - >>> (E) @LINE: 765 TEST.VALUE:ModeTreatment.<>.presDialysateFlowRate:300 - >>> >>> Unknown parameter/object name presDialysateFlowRate - >>> >>> Value Line Error - Command Ignored - >>> (E) @LINE: 773 TEST.EXPECTED:ModeTreatment.<>.presBloodFlowRate:300 - >>> >>> Unknown parameter/object name presBloodFlowRate - >>> >>> Value Line Error - Command Ignored - >>> (E) @LINE: 774 TEST.EXPECTED:ModeTreatment.<>.presDialysateFlowRate:400 - >>> >>> Unknown parameter/object name presDialysateFlowRate - >>> >>> Value Line Error - Command Ignored -TEST.END_IMPORT_FAILURES: -TEST.STUB:ModeTreatment.broadcastTreatmentSettingsRanges -TEST.VALUE:ModeTreatment.<>.currentTreatmentState:TREATMENT_DIALYSIS_STATE -TEST.VALUE:ModeTreatment.<>.presTreatmentTimeSecs:14400 -TEST.VALUE:ModeTreatment.<>.presMaxUFVolumeML:2400.0 -TEST.VALUE:ModeTreatment.<>.presUFRate:10.0 -TEST.VALUE:ModeTreatment.verifyBloodAndDialysateRateSettingsChange.bloodRate:300 -TEST.VALUE:ModeTreatment.verifyBloodAndDialysateRateSettingsChange.dialRate:400 -TEST.VALUE:uut_prototype_stubs.getCurrentOperationMode.return:MODE_TREA -TEST.VALUE:uut_prototype_stubs.sendChangeBloodDialysateRateChangeResponse.return:1 -TEST.EXPECTED:ModeTreatment.<>.presTreatmentTimeSecs:14400 -TEST.EXPECTED:ModeTreatment.<>.presMaxUFVolumeML:2400.0 -TEST.EXPECTED:ModeTreatment.<>.presUFRate:10.0 -TEST.EXPECTED:ModeTreatment.verifyBloodAndDialysateRateSettingsChange.return:1 -TEST.EXPECTED:uut_prototype_stubs.sendChangeBloodDialysateRateChangeResponse.accepted:1 -TEST.EXPECTED:uut_prototype_stubs.sendChangeBloodDialysateRateChangeResponse.reason:0 +TEST.NAME:verifyBloodAndDialysateRateSettingsChange_01_not_in_Tx +TEST.VALUE:ModeTreatment.<>.presTreatmentTimeSecs:60 +TEST.VALUE:ModeTreatment.<>.presUFRate:2.0 +TEST.VALUE:ModeTreatment.verifyBloodAndDialysateRateSettingsChange.bloodRate:100 +TEST.VALUE:ModeTreatment.verifyBloodAndDialysateRateSettingsChange.dialRate:150 +TEST.VALUE:uut_prototype_stubs.getUltrafiltrationReferenceVolume.return:2.0 +TEST.VALUE:uut_prototype_stubs.getCurrentOperationMode.return:MODE_SERV +TEST.VALUE:uut_prototype_stubs.getTreatmentParameterU32.return:150 +TEST.VALUE:uut_prototype_stubs.getU32TreatmentParamLowerRangeLimit.return:100 +TEST.VALUE:uut_prototype_stubs.getU32TreatmentParamUpperRangeLimit.return:500 +TEST.VALUE:uut_prototype_stubs.getF32TreatmentParamLowerRangeLimit.return:30.0 +TEST.VALUE:uut_prototype_stubs.getF32TreatmentParamUpperRangeLimit.return:35.0 +TEST.EXPECTED:ModeTreatment.verifyBloodAndDialysateRateSettingsChange.return:0 TEST.END --- Test Case: verifyBloodAndDialysateRateSettingsChange_Accepted.001 +-- Test Case: verifyBloodAndDialysateRateSettingsChange_02_in_Tx_blood_flow_below_min TEST.UNIT:ModeTreatment TEST.SUBPROGRAM:verifyBloodAndDialysateRateSettingsChange TEST.NEW -TEST.NAME:verifyBloodAndDialysateRateSettingsChange_Accepted.001 -TEST.IMPORT_FAILURES: -(E) Errors from previous script import(s) - >>> (E) @LINE: 2890 TEST.EXPECTED:uut_prototype_stubs.setDialysisParams.maxUFVol:2400.0 - >>> >>> Could not find function setDialysisParams - >>> >>> in unit uut_prototype_stubs. - >>> >>> Value Line Error - Command Ignored - >>> (E) @LINE: 2891 TEST.EXPECTED:uut_prototype_stubs.setDialysisParams.uFRate:10.0 - >>> >>> Could not find function setDialysisParams - >>> >>> in unit uut_prototype_stubs. - >>> >>> Value Line Error - Command Ignored -TEST.END_IMPORT_FAILURES: -TEST.STUB:ModeTreatment.broadcastTreatmentSettingsRanges -TEST.VALUE:ModeTreatment.<>.currentTreatmentState:TREATMENT_DIALYSIS_STATE -TEST.VALUE:ModeTreatment.<>.presTreatmentTimeSecs:14400 -TEST.VALUE:ModeTreatment.<>.presMaxUFVolumeML:2400.0 -TEST.VALUE:ModeTreatment.<>.presUFRate:10.0 -TEST.VALUE:ModeTreatment.verifyBloodAndDialysateRateSettingsChange.bloodRate:300 -TEST.VALUE:ModeTreatment.verifyBloodAndDialysateRateSettingsChange.dialRate:400 +TEST.NAME:verifyBloodAndDialysateRateSettingsChange_02_in_Tx_blood_flow_below_min +TEST.VALUE:ModeTreatment.<>.presTreatmentTimeSecs:60 +TEST.VALUE:ModeTreatment.<>.presUFRate:2.0 +TEST.VALUE:ModeTreatment.verifyBloodAndDialysateRateSettingsChange.bloodRate:22 +TEST.VALUE:ModeTreatment.verifyBloodAndDialysateRateSettingsChange.dialRate:150 +TEST.VALUE:uut_prototype_stubs.getUltrafiltrationReferenceVolume.return:2.0 TEST.VALUE:uut_prototype_stubs.getCurrentOperationMode.return:MODE_TREA -TEST.VALUE:uut_prototype_stubs.getTreatmentParameterU32.return:1 -TEST.VALUE:uut_prototype_stubs.sendChangeBloodDialysateRateChangeResponse.return:1 -TEST.EXPECTED:ModeTreatment.<>.currentTreatmentState:TREATMENT_DIALYSIS_STATE -TEST.EXPECTED:ModeTreatment.<>.presTreatmentTimeSecs:14400 -TEST.EXPECTED:ModeTreatment.<>.presMaxUFVolumeML:2400.0 -TEST.EXPECTED:ModeTreatment.<>.presUFRate:10.0 -TEST.EXPECTED:ModeTreatment.verifyBloodAndDialysateRateSettingsChange.bloodRate:300 -TEST.EXPECTED:ModeTreatment.verifyBloodAndDialysateRateSettingsChange.dialRate:400 -TEST.EXPECTED:ModeTreatment.verifyBloodAndDialysateRateSettingsChange.return:1 -TEST.EXPECTED:uut_prototype_stubs.sendChangeBloodDialysateRateChangeResponse.accepted:1 -TEST.EXPECTED:uut_prototype_stubs.sendChangeBloodDialysateRateChangeResponse.reason:0 -TEST.END - --- Test Case: verifyBloodAndDialysateRateSettingsChange_Accepted.002 -TEST.UNIT:ModeTreatment -TEST.SUBPROGRAM:verifyBloodAndDialysateRateSettingsChange -TEST.NEW -TEST.NAME:verifyBloodAndDialysateRateSettingsChange_Accepted.002 -TEST.IMPORT_FAILURES: -(E) Errors from previous script import(s) - >>> (E) @LINE: 2917 TEST.EXPECTED:uut_prototype_stubs.setDialysisParams.maxUFVol:2400.0 - >>> >>> Could not find function setDialysisParams - >>> >>> in unit uut_prototype_stubs. - >>> >>> Value Line Error - Command Ignored - >>> (E) @LINE: 2918 TEST.EXPECTED:uut_prototype_stubs.setDialysisParams.uFRate:10.0 - >>> >>> Could not find function setDialysisParams - >>> >>> in unit uut_prototype_stubs. - >>> >>> Value Line Error - Command Ignored -TEST.END_IMPORT_FAILURES: -TEST.STUB:ModeTreatment.broadcastTreatmentSettingsRanges -TEST.VALUE:ModeTreatment.<>.currentTreatmentState:TREATMENT_DIALYSIS_STATE -TEST.VALUE:ModeTreatment.<>.presTreatmentTimeSecs:14400 -TEST.VALUE:ModeTreatment.<>.presMaxUFVolumeML:2400.0 -TEST.VALUE:ModeTreatment.<>.presUFRate:10.0 -TEST.VALUE:ModeTreatment.verifyBloodAndDialysateRateSettingsChange.bloodRate:300 -TEST.VALUE:ModeTreatment.verifyBloodAndDialysateRateSettingsChange.dialRate:400 -TEST.VALUE:uut_prototype_stubs.getCurrentOperationMode.return:MODE_TREA -TEST.VALUE:uut_prototype_stubs.getTreatmentParameterU32.return:300 -TEST.VALUE:uut_prototype_stubs.sendChangeBloodDialysateRateChangeResponse.bloodRate:0 -TEST.VALUE:uut_prototype_stubs.sendChangeBloodDialysateRateChangeResponse.return:1 -TEST.EXPECTED:ModeTreatment.<>.presTreatmentTimeSecs:14400 -TEST.EXPECTED:ModeTreatment.<>.presMaxUFVolumeML:2400.0 -TEST.EXPECTED:ModeTreatment.<>.presUFRate:10.0 -TEST.EXPECTED:ModeTreatment.verifyBloodAndDialysateRateSettingsChange.return:1 -TEST.EXPECTED:uut_prototype_stubs.sendChangeBloodDialysateRateChangeResponse.accepted:1 -TEST.EXPECTED:uut_prototype_stubs.sendChangeBloodDialysateRateChangeResponse.reason:0 -TEST.END - --- Test Case: verifyBloodAndDialysateRateSettingsChange_Accepted.003 -TEST.UNIT:ModeTreatment -TEST.SUBPROGRAM:verifyBloodAndDialysateRateSettingsChange -TEST.NEW -TEST.NAME:verifyBloodAndDialysateRateSettingsChange_Accepted.003 -TEST.STUB:ModeTreatment.broadcastTreatmentSettingsRanges -TEST.VALUE:ModeTreatment.<>.currentTreatmentState:TREATMENT_DIALYSIS_STATE -TEST.VALUE:ModeTreatment.<>.presTreatmentTimeSecs:14400 -TEST.VALUE:ModeTreatment.<>.presMaxUFVolumeML:2400.0 -TEST.VALUE:ModeTreatment.<>.presUFRate:10.0 -TEST.VALUE:ModeTreatment.verifyBloodAndDialysateRateSettingsChange.bloodRate:300 -TEST.VALUE:ModeTreatment.verifyBloodAndDialysateRateSettingsChange.dialRate:1 -TEST.VALUE:uut_prototype_stubs.getCurrentOperationMode.return:MODE_TREA -TEST.VALUE:uut_prototype_stubs.getTreatmentParameterU32.param:TREATMENT_PARAM_BLOOD_FLOW -TEST.VALUE:uut_prototype_stubs.getTreatmentParameterU32.return:1 -TEST.VALUE:uut_prototype_stubs.sendChangeBloodDialysateRateChangeResponse.bloodRate:0 -TEST.VALUE:uut_prototype_stubs.sendChangeBloodDialysateRateChangeResponse.dialRate:1 -TEST.VALUE:uut_prototype_stubs.sendChangeBloodDialysateRateChangeResponse.return:1 -TEST.EXPECTED:ModeTreatment.<>.currentTreatmentState:TREATMENT_DIALYSIS_STATE -TEST.EXPECTED:ModeTreatment.<>.presTreatmentTimeSecs:14400 -TEST.EXPECTED:ModeTreatment.<>.presMaxUFVolumeML:2400.0 -TEST.EXPECTED:ModeTreatment.<>.presUFRate:10.0 -TEST.EXPECTED:ModeTreatment.verifyBloodAndDialysateRateSettingsChange.bloodRate:300 -TEST.EXPECTED:ModeTreatment.verifyBloodAndDialysateRateSettingsChange.dialRate:1 -TEST.EXPECTED:uut_prototype_stubs.getTreatmentParameterU32.param:TREATMENT_PARAM_DIALYSATE_FLOW,TREATMENT_PARAM_BLOOD_FLOW -TEST.EXPECTED:uut_prototype_stubs.sendChangeBloodDialysateRateChangeResponse.bloodRate:1 -TEST.EXPECTED:uut_prototype_stubs.sendChangeBloodDialysateRateChangeResponse.dialRate:1 -TEST.END - --- Test Case: verifyBloodAndDialysateRateSettingsChange_Accepted.004 -TEST.UNIT:ModeTreatment -TEST.SUBPROGRAM:verifyBloodAndDialysateRateSettingsChange -TEST.NEW -TEST.NAME:verifyBloodAndDialysateRateSettingsChange_Accepted.004 -TEST.STUB:ModeTreatment.broadcastTreatmentSettingsRanges -TEST.VALUE:ModeTreatment.<>.currentTreatmentState:TREATMENT_DIALYSIS_STATE -TEST.VALUE:ModeTreatment.<>.presTreatmentTimeSecs:14400 -TEST.VALUE:ModeTreatment.<>.presMaxUFVolumeML:2400.0 -TEST.VALUE:ModeTreatment.<>.presUFRate:10.0 -TEST.VALUE:ModeTreatment.verifyBloodAndDialysateRateSettingsChange.bloodRate:1 -TEST.VALUE:ModeTreatment.verifyBloodAndDialysateRateSettingsChange.dialRate:400 -TEST.VALUE:uut_prototype_stubs.getCurrentOperationMode.return:MODE_TREA -TEST.VALUE:uut_prototype_stubs.sendChangeBloodDialysateRateChangeResponse.return:1 -TEST.EXPECTED:ModeTreatment.<>.currentTreatmentState:TREATMENT_DIALYSIS_STATE -TEST.EXPECTED:ModeTreatment.<>.presTreatmentTimeSecs:14400 -TEST.EXPECTED:ModeTreatment.<>.presMaxUFVolumeML:2400.0 -TEST.EXPECTED:ModeTreatment.<>.presUFRate:10.0 -TEST.EXPECTED:ModeTreatment.verifyBloodAndDialysateRateSettingsChange.bloodRate:1 -TEST.EXPECTED:ModeTreatment.verifyBloodAndDialysateRateSettingsChange.dialRate:400 +TEST.VALUE:uut_prototype_stubs.getTreatmentParameterU32.return:150 +TEST.VALUE:uut_prototype_stubs.getU32TreatmentParamLowerRangeLimit.return:100 +TEST.VALUE:uut_prototype_stubs.getU32TreatmentParamUpperRangeLimit.return:500 +TEST.VALUE:uut_prototype_stubs.getF32TreatmentParamLowerRangeLimit.return:30.0 +TEST.VALUE:uut_prototype_stubs.getF32TreatmentParamUpperRangeLimit.return:35.0 TEST.EXPECTED:ModeTreatment.verifyBloodAndDialysateRateSettingsChange.return:0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeBloodDialysateRateChangeResponse.accepted:0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeBloodDialysateRateChangeResponse.reason:8 TEST.END --- Test Case: verifyBloodAndDialysateRateSettingsChange_Accepted.005 +-- Test Case: verifyBloodAndDialysateRateSettingsChange_03_in_Tx_blood_flow_above_max TEST.UNIT:ModeTreatment TEST.SUBPROGRAM:verifyBloodAndDialysateRateSettingsChange TEST.NEW -TEST.NAME:verifyBloodAndDialysateRateSettingsChange_Accepted.005 -TEST.STUB:ModeTreatment.broadcastTreatmentSettingsRanges -TEST.VALUE:ModeTreatment.<>.currentTreatmentState:TREATMENT_DIALYSIS_STATE -TEST.VALUE:ModeTreatment.<>.presTreatmentTimeSecs:14400 -TEST.VALUE:ModeTreatment.<>.presMaxUFVolumeML:2400.0 -TEST.VALUE:ModeTreatment.<>.presUFRate:10.0 -TEST.VALUE:ModeTreatment.verifyBloodAndDialysateRateSettingsChange.bloodRate:300 -TEST.VALUE:ModeTreatment.verifyBloodAndDialysateRateSettingsChange.dialRate:<> +TEST.NAME:verifyBloodAndDialysateRateSettingsChange_03_in_Tx_blood_flow_above_max +TEST.VALUE:ModeTreatment.<>.presTreatmentTimeSecs:60 +TEST.VALUE:ModeTreatment.<>.presUFRate:2.0 +TEST.VALUE:ModeTreatment.verifyBloodAndDialysateRateSettingsChange.bloodRate:650 +TEST.VALUE:ModeTreatment.verifyBloodAndDialysateRateSettingsChange.dialRate:150 +TEST.VALUE:uut_prototype_stubs.getUltrafiltrationReferenceVolume.return:2.0 TEST.VALUE:uut_prototype_stubs.getCurrentOperationMode.return:MODE_TREA -TEST.VALUE:uut_prototype_stubs.getTreatmentParameterU32.return:<> -TEST.VALUE:uut_prototype_stubs.sendChangeBloodDialysateRateChangeResponse.bloodRate:0 -TEST.VALUE:uut_prototype_stubs.sendChangeBloodDialysateRateChangeResponse.dialRate:<> -TEST.VALUE:uut_prototype_stubs.sendChangeBloodDialysateRateChangeResponse.return:1 -TEST.EXPECTED:ModeTreatment.<>.currentTreatmentState:TREATMENT_DIALYSIS_STATE -TEST.EXPECTED:ModeTreatment.<>.presTreatmentTimeSecs:14400 -TEST.EXPECTED:ModeTreatment.<>.presMaxUFVolumeML:2400.0 -TEST.EXPECTED:ModeTreatment.<>.presUFRate:10.0 -TEST.EXPECTED:ModeTreatment.verifyBloodAndDialysateRateSettingsChange.bloodRate:300 -TEST.EXPECTED:ModeTreatment.verifyBloodAndDialysateRateSettingsChange.dialRate:4294967295 +TEST.VALUE:uut_prototype_stubs.getTreatmentParameterU32.return:150 +TEST.VALUE:uut_prototype_stubs.getU32TreatmentParamLowerRangeLimit.return:100 +TEST.VALUE:uut_prototype_stubs.getU32TreatmentParamUpperRangeLimit.return:500 +TEST.VALUE:uut_prototype_stubs.getF32TreatmentParamLowerRangeLimit.return:30.0 +TEST.VALUE:uut_prototype_stubs.getF32TreatmentParamUpperRangeLimit.return:35.0 TEST.EXPECTED:ModeTreatment.verifyBloodAndDialysateRateSettingsChange.return:0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeBloodDialysateRateChangeResponse.accepted:0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeBloodDialysateRateChangeResponse.reason:9 -TEST.EXPECTED:uut_prototype_stubs.sendChangeBloodDialysateRateChangeResponse.bloodRate:4294967295 -TEST.EXPECTED:uut_prototype_stubs.sendChangeBloodDialysateRateChangeResponse.dialRate:4294967295 TEST.END --- Test Case: verifyBloodAndDialysateRateSettingsChange_Accepted.006 +-- Test Case: verifyBloodAndDialysateRateSettingsChange_04_in_Tx_dial_flow_below_min TEST.UNIT:ModeTreatment TEST.SUBPROGRAM:verifyBloodAndDialysateRateSettingsChange TEST.NEW -TEST.NAME:verifyBloodAndDialysateRateSettingsChange_Accepted.006 -TEST.IMPORT_FAILURES: -(E) Errors from previous script import(s) - >>> (E) @LINE: 3027 TEST.EXPECTED:uut_prototype_stubs.setDialysisParams.maxUFVol:2400.0 - >>> >>> Could not find function setDialysisParams - >>> >>> in unit uut_prototype_stubs. - >>> >>> Value Line Error - Command Ignored - >>> (E) @LINE: 3028 TEST.EXPECTED:uut_prototype_stubs.setDialysisParams.uFRate:10.0 - >>> >>> Could not find function setDialysisParams - >>> >>> in unit uut_prototype_stubs. - >>> >>> Value Line Error - Command Ignored -TEST.END_IMPORT_FAILURES: -TEST.STUB:ModeTreatment.broadcastTreatmentSettingsRanges -TEST.VALUE:ModeTreatment.<>.currentTreatmentState:TREATMENT_DIALYSIS_STATE -TEST.VALUE:ModeTreatment.<>.presTreatmentTimeSecs:14400 -TEST.VALUE:ModeTreatment.<>.presMaxUFVolumeML:2400.0 -TEST.VALUE:ModeTreatment.<>.presUFRate:10.0 +TEST.NAME:verifyBloodAndDialysateRateSettingsChange_04_in_Tx_dial_flow_below_min +TEST.VALUE:ModeTreatment.<>.presTreatmentTimeSecs:60 +TEST.VALUE:ModeTreatment.<>.presUFRate:2.0 TEST.VALUE:ModeTreatment.verifyBloodAndDialysateRateSettingsChange.bloodRate:300 -TEST.VALUE:ModeTreatment.verifyBloodAndDialysateRateSettingsChange.dialRate:300 +TEST.VALUE:ModeTreatment.verifyBloodAndDialysateRateSettingsChange.dialRate:22 +TEST.VALUE:uut_prototype_stubs.getUltrafiltrationReferenceVolume.return:2.0 TEST.VALUE:uut_prototype_stubs.getCurrentOperationMode.return:MODE_TREA -TEST.VALUE:uut_prototype_stubs.getTreatmentParameterU32.return:300 -TEST.VALUE:uut_prototype_stubs.sendChangeBloodDialysateRateChangeResponse.bloodRate:0 -TEST.VALUE:uut_prototype_stubs.sendChangeBloodDialysateRateChangeResponse.return:1 -TEST.EXPECTED:ModeTreatment.<>.presTreatmentTimeSecs:14400 -TEST.EXPECTED:ModeTreatment.<>.presMaxUFVolumeML:2400.0 -TEST.EXPECTED:ModeTreatment.<>.presUFRate:10.0 -TEST.EXPECTED:ModeTreatment.verifyBloodAndDialysateRateSettingsChange.return:1 -TEST.EXPECTED:uut_prototype_stubs.sendChangeBloodDialysateRateChangeResponse.accepted:1 -TEST.EXPECTED:uut_prototype_stubs.sendChangeBloodDialysateRateChangeResponse.reason:0 -TEST.END - --- Test Case: verifyBloodAndDialysateRateSettingsChange_BloodOutOfRange1 -TEST.UNIT:ModeTreatment -TEST.SUBPROGRAM:verifyBloodAndDialysateRateSettingsChange -TEST.NEW -TEST.NAME:verifyBloodAndDialysateRateSettingsChange_BloodOutOfRange1 -TEST.IMPORT_FAILURES: -(E) Errors from previous script import(s) - >>> (E) @LINE: 795 TEST.VALUE:ModeTreatment.<>.presBloodFlowRate:200 - >>> >>> Unknown parameter/object name presBloodFlowRate - >>> >>> Value Line Error - Command Ignored - >>> (E) @LINE: 796 TEST.VALUE:ModeTreatment.<>.presDialysateFlowRate:300 - >>> >>> Unknown parameter/object name presDialysateFlowRate - >>> >>> Value Line Error - Command Ignored - >>> (E) @LINE: 804 TEST.EXPECTED:ModeTreatment.<>.presBloodFlowRate:200 - >>> >>> Unknown parameter/object name presBloodFlowRate - >>> >>> Value Line Error - Command Ignored - >>> (E) @LINE: 805 TEST.EXPECTED:ModeTreatment.<>.presDialysateFlowRate:300 - >>> >>> Unknown parameter/object name presDialysateFlowRate - >>> >>> Value Line Error - Command Ignored -TEST.END_IMPORT_FAILURES: -TEST.STUB:ModeTreatment.broadcastTreatmentSettingsRanges -TEST.VALUE:ModeTreatment.<>.currentTreatmentState:TREATMENT_DIALYSIS_STATE -TEST.VALUE:ModeTreatment.<>.presTreatmentTimeSecs:14400 -TEST.VALUE:ModeTreatment.<>.presMaxUFVolumeML:2400.0 -TEST.VALUE:ModeTreatment.<>.presUFRate:10.0 -TEST.VALUE:ModeTreatment.verifyBloodAndDialysateRateSettingsChange.bloodRate:550 -TEST.VALUE:ModeTreatment.verifyBloodAndDialysateRateSettingsChange.dialRate:400 -TEST.VALUE:uut_prototype_stubs.getCurrentOperationMode.return:MODE_TREA -TEST.VALUE:uut_prototype_stubs.sendChangeBloodDialysateRateChangeResponse.return:1 -TEST.EXPECTED:ModeTreatment.<>.presTreatmentTimeSecs:14400 -TEST.EXPECTED:ModeTreatment.<>.presMaxUFVolumeML:2400.0 -TEST.EXPECTED:ModeTreatment.<>.presUFRate:10.0 +TEST.VALUE:uut_prototype_stubs.getTreatmentParameterU32.return:150 +TEST.VALUE:uut_prototype_stubs.getU32TreatmentParamLowerRangeLimit.return:100 +TEST.VALUE:uut_prototype_stubs.getU32TreatmentParamUpperRangeLimit.return:500 +TEST.VALUE:uut_prototype_stubs.getF32TreatmentParamLowerRangeLimit.return:30.0 +TEST.VALUE:uut_prototype_stubs.getF32TreatmentParamUpperRangeLimit.return:35.0 TEST.EXPECTED:ModeTreatment.verifyBloodAndDialysateRateSettingsChange.return:0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeBloodDialysateRateChangeResponse.accepted:0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeBloodDialysateRateChangeResponse.reason:8 TEST.END --- Test Case: verifyBloodAndDialysateRateSettingsChange_BloodOutOfRange2 +-- Test Case: verifyBloodAndDialysateRateSettingsChange_05_in_Tx_dial_flow_above_max TEST.UNIT:ModeTreatment TEST.SUBPROGRAM:verifyBloodAndDialysateRateSettingsChange TEST.NEW -TEST.NAME:verifyBloodAndDialysateRateSettingsChange_BloodOutOfRange2 -TEST.IMPORT_FAILURES: -(E) Errors from previous script import(s) - >>> (E) @LINE: 822 TEST.VALUE:ModeTreatment.<>.presBloodFlowRate:200 - >>> >>> Unknown parameter/object name presBloodFlowRate - >>> >>> Value Line Error - Command Ignored - >>> (E) @LINE: 823 TEST.VALUE:ModeTreatment.<>.presDialysateFlowRate:300 - >>> >>> Unknown parameter/object name presDialysateFlowRate - >>> >>> Value Line Error - Command Ignored - >>> (E) @LINE: 831 TEST.EXPECTED:ModeTreatment.<>.presBloodFlowRate:200 - >>> >>> Unknown parameter/object name presBloodFlowRate - >>> >>> Value Line Error - Command Ignored - >>> (E) @LINE: 832 TEST.EXPECTED:ModeTreatment.<>.presDialysateFlowRate:300 - >>> >>> Unknown parameter/object name presDialysateFlowRate - >>> >>> Value Line Error - Command Ignored -TEST.END_IMPORT_FAILURES: -TEST.STUB:ModeTreatment.broadcastTreatmentSettingsRanges -TEST.VALUE:ModeTreatment.<>.currentTreatmentState:TREATMENT_DIALYSIS_STATE -TEST.VALUE:ModeTreatment.<>.presTreatmentTimeSecs:14400 -TEST.VALUE:ModeTreatment.<>.presMaxUFVolumeML:2400.0 -TEST.VALUE:ModeTreatment.<>.presUFRate:10.0 -TEST.VALUE:ModeTreatment.verifyBloodAndDialysateRateSettingsChange.bloodRate:90 -TEST.VALUE:ModeTreatment.verifyBloodAndDialysateRateSettingsChange.dialRate:400 -TEST.VALUE:uut_prototype_stubs.getCurrentOperationMode.return:MODE_TREA -TEST.VALUE:uut_prototype_stubs.sendChangeBloodDialysateRateChangeResponse.return:1 -TEST.EXPECTED:ModeTreatment.<>.presTreatmentTimeSecs:14400 -TEST.EXPECTED:ModeTreatment.<>.presMaxUFVolumeML:2400.0 -TEST.EXPECTED:ModeTreatment.<>.presUFRate:10.0 -TEST.EXPECTED:ModeTreatment.verifyBloodAndDialysateRateSettingsChange.return:0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeBloodDialysateRateChangeResponse.accepted:0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeBloodDialysateRateChangeResponse.reason:8 -TEST.END - --- Test Case: verifyBloodAndDialysateRateSettingsChange_DialOutOfRange1 -TEST.UNIT:ModeTreatment -TEST.SUBPROGRAM:verifyBloodAndDialysateRateSettingsChange -TEST.NEW -TEST.NAME:verifyBloodAndDialysateRateSettingsChange_DialOutOfRange1 -TEST.IMPORT_FAILURES: -(E) Errors from previous script import(s) - >>> (E) @LINE: 849 TEST.VALUE:ModeTreatment.<>.presBloodFlowRate:200 - >>> >>> Unknown parameter/object name presBloodFlowRate - >>> >>> Value Line Error - Command Ignored - >>> (E) @LINE: 850 TEST.VALUE:ModeTreatment.<>.presDialysateFlowRate:300 - >>> >>> Unknown parameter/object name presDialysateFlowRate - >>> >>> Value Line Error - Command Ignored - >>> (E) @LINE: 858 TEST.EXPECTED:ModeTreatment.<>.presBloodFlowRate:200 - >>> >>> Unknown parameter/object name presBloodFlowRate - >>> >>> Value Line Error - Command Ignored - >>> (E) @LINE: 859 TEST.EXPECTED:ModeTreatment.<>.presDialysateFlowRate:300 - >>> >>> Unknown parameter/object name presDialysateFlowRate - >>> >>> Value Line Error - Command Ignored -TEST.END_IMPORT_FAILURES: -TEST.STUB:ModeTreatment.broadcastTreatmentSettingsRanges -TEST.VALUE:ModeTreatment.<>.currentTreatmentState:TREATMENT_DIALYSIS_STATE -TEST.VALUE:ModeTreatment.<>.presTreatmentTimeSecs:14400 -TEST.VALUE:ModeTreatment.<>.presMaxUFVolumeML:2400.0 -TEST.VALUE:ModeTreatment.<>.presUFRate:10.0 +TEST.NAME:verifyBloodAndDialysateRateSettingsChange_05_in_Tx_dial_flow_above_max +TEST.VALUE:ModeTreatment.<>.presTreatmentTimeSecs:60 +TEST.VALUE:ModeTreatment.<>.presUFRate:2.0 TEST.VALUE:ModeTreatment.verifyBloodAndDialysateRateSettingsChange.bloodRate:300 TEST.VALUE:ModeTreatment.verifyBloodAndDialysateRateSettingsChange.dialRate:650 +TEST.VALUE:uut_prototype_stubs.getUltrafiltrationReferenceVolume.return:2.0 TEST.VALUE:uut_prototype_stubs.getCurrentOperationMode.return:MODE_TREA -TEST.VALUE:uut_prototype_stubs.sendChangeBloodDialysateRateChangeResponse.return:1 -TEST.EXPECTED:ModeTreatment.<>.presTreatmentTimeSecs:14400 -TEST.EXPECTED:ModeTreatment.<>.presMaxUFVolumeML:2400.0 -TEST.EXPECTED:ModeTreatment.<>.presUFRate:10.0 +TEST.VALUE:uut_prototype_stubs.getTreatmentParameterU32.return:150 +TEST.VALUE:uut_prototype_stubs.getU32TreatmentParamLowerRangeLimit.return:100 +TEST.VALUE:uut_prototype_stubs.getU32TreatmentParamUpperRangeLimit.return:500 +TEST.VALUE:uut_prototype_stubs.getF32TreatmentParamLowerRangeLimit.return:30.0 +TEST.VALUE:uut_prototype_stubs.getF32TreatmentParamUpperRangeLimit.return:35.0 TEST.EXPECTED:ModeTreatment.verifyBloodAndDialysateRateSettingsChange.return:0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeBloodDialysateRateChangeResponse.accepted:0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeBloodDialysateRateChangeResponse.reason:9 TEST.END --- Test Case: verifyBloodAndDialysateRateSettingsChange_DialOutOfRange2 +-- Test Case: verifyBloodAndDialysateRateSettingsChange_06_in_Tx_dial_volume_above_range TEST.UNIT:ModeTreatment TEST.SUBPROGRAM:verifyBloodAndDialysateRateSettingsChange TEST.NEW -TEST.NAME:verifyBloodAndDialysateRateSettingsChange_DialOutOfRange2 -TEST.IMPORT_FAILURES: -(E) Errors from previous script import(s) - >>> (E) @LINE: 876 TEST.VALUE:ModeTreatment.<>.presBloodFlowRate:200 - >>> >>> Unknown parameter/object name presBloodFlowRate - >>> >>> Value Line Error - Command Ignored - >>> (E) @LINE: 877 TEST.VALUE:ModeTreatment.<>.presDialysateFlowRate:300 - >>> >>> Unknown parameter/object name presDialysateFlowRate - >>> >>> Value Line Error - Command Ignored - >>> (E) @LINE: 885 TEST.EXPECTED:ModeTreatment.<>.presBloodFlowRate:200 - >>> >>> Unknown parameter/object name presBloodFlowRate - >>> >>> Value Line Error - Command Ignored - >>> (E) @LINE: 886 TEST.EXPECTED:ModeTreatment.<>.presDialysateFlowRate:300 - >>> >>> Unknown parameter/object name presDialysateFlowRate - >>> >>> Value Line Error - Command Ignored -TEST.END_IMPORT_FAILURES: -TEST.STUB:ModeTreatment.broadcastTreatmentSettingsRanges -TEST.VALUE:ModeTreatment.<>.currentTreatmentState:TREATMENT_DIALYSIS_STATE -TEST.VALUE:ModeTreatment.<>.presTreatmentTimeSecs:14400 -TEST.VALUE:ModeTreatment.<>.presMaxUFVolumeML:2400.0 -TEST.VALUE:ModeTreatment.<>.presUFRate:10.0 +TEST.NAME:verifyBloodAndDialysateRateSettingsChange_06_in_Tx_dial_volume_above_range +TEST.VALUE:ModeTreatment.<>.presTreatmentTimeSecs:1000000 +TEST.VALUE:ModeTreatment.<>.presUFRate:2.0 TEST.VALUE:ModeTreatment.verifyBloodAndDialysateRateSettingsChange.bloodRate:300 -TEST.VALUE:ModeTreatment.verifyBloodAndDialysateRateSettingsChange.dialRate:90 +TEST.VALUE:ModeTreatment.verifyBloodAndDialysateRateSettingsChange.dialRate:300 +TEST.VALUE:uut_prototype_stubs.getUltrafiltrationReferenceVolume.return:2.0 TEST.VALUE:uut_prototype_stubs.getCurrentOperationMode.return:MODE_TREA -TEST.VALUE:uut_prototype_stubs.sendChangeBloodDialysateRateChangeResponse.return:1 -TEST.EXPECTED:ModeTreatment.<>.presTreatmentTimeSecs:14400 -TEST.EXPECTED:ModeTreatment.<>.presMaxUFVolumeML:2400.0 -TEST.EXPECTED:ModeTreatment.<>.presUFRate:10.0 +TEST.VALUE:uut_prototype_stubs.getTreatmentParameterU32.return:150 +TEST.VALUE:uut_prototype_stubs.getU32TreatmentParamLowerRangeLimit.return:100 +TEST.VALUE:uut_prototype_stubs.getU32TreatmentParamUpperRangeLimit.return:500 +TEST.VALUE:uut_prototype_stubs.getF32TreatmentParamLowerRangeLimit.return:30.0 +TEST.VALUE:uut_prototype_stubs.getF32TreatmentParamUpperRangeLimit.return:35.0 TEST.EXPECTED:ModeTreatment.verifyBloodAndDialysateRateSettingsChange.return:0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeBloodDialysateRateChangeResponse.accepted:0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeBloodDialysateRateChangeResponse.reason:9 TEST.END --- Test Case: verifyBloodAndDialysateRateSettingsChange_DialVolOutOfRange +-- Test Case: verifyBloodAndDialysateRateSettingsChange_07_in_Tx_blood_dial_rate_not_eq_to_param TEST.UNIT:ModeTreatment TEST.SUBPROGRAM:verifyBloodAndDialysateRateSettingsChange TEST.NEW -TEST.NAME:verifyBloodAndDialysateRateSettingsChange_DialVolOutOfRange -TEST.IMPORT_FAILURES: -(E) Errors from previous script import(s) - >>> (E) @LINE: 903 TEST.VALUE:ModeTreatment.<>.presBloodFlowRate:200 - >>> >>> Unknown parameter/object name presBloodFlowRate - >>> >>> Value Line Error - Command Ignored - >>> (E) @LINE: 904 TEST.VALUE:ModeTreatment.<>.presDialysateFlowRate:300 - >>> >>> Unknown parameter/object name presDialysateFlowRate - >>> >>> Value Line Error - Command Ignored - >>> (E) @LINE: 912 TEST.EXPECTED:ModeTreatment.<>.presBloodFlowRate:200 - >>> >>> Unknown parameter/object name presBloodFlowRate - >>> >>> Value Line Error - Command Ignored - >>> (E) @LINE: 913 TEST.EXPECTED:ModeTreatment.<>.presDialysateFlowRate:300 - >>> >>> Unknown parameter/object name presDialysateFlowRate - >>> >>> Value Line Error - Command Ignored -TEST.END_IMPORT_FAILURES: -TEST.STUB:ModeTreatment.broadcastTreatmentSettingsRanges -TEST.VALUE:ModeTreatment.<>.currentTreatmentState:TREATMENT_DIALYSIS_STATE -TEST.VALUE:ModeTreatment.<>.presTreatmentTimeSecs:28800 -TEST.VALUE:ModeTreatment.<>.presMaxUFVolumeML:2400.0 -TEST.VALUE:ModeTreatment.<>.presUFRate:10.0 +TEST.NAME:verifyBloodAndDialysateRateSettingsChange_07_in_Tx_blood_dial_rate_not_eq_to_param +TEST.VALUE:ModeTreatment.<>.presTreatmentTimeSecs:60 +TEST.VALUE:ModeTreatment.<>.presUFRate:2.0 TEST.VALUE:ModeTreatment.verifyBloodAndDialysateRateSettingsChange.bloodRate:300 -TEST.VALUE:ModeTreatment.verifyBloodAndDialysateRateSettingsChange.dialRate:400 +TEST.VALUE:ModeTreatment.verifyBloodAndDialysateRateSettingsChange.dialRate:300 +TEST.VALUE:uut_prototype_stubs.getUltrafiltrationReferenceVolume.return:2.0 TEST.VALUE:uut_prototype_stubs.getCurrentOperationMode.return:MODE_TREA -TEST.VALUE:uut_prototype_stubs.sendChangeBloodDialysateRateChangeResponse.return:1 -TEST.EXPECTED:ModeTreatment.<>.presTreatmentTimeSecs:28800 -TEST.EXPECTED:ModeTreatment.<>.presMaxUFVolumeML:2400.0 -TEST.EXPECTED:ModeTreatment.<>.presUFRate:10.0 -TEST.EXPECTED:ModeTreatment.verifyBloodAndDialysateRateSettingsChange.return:0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeBloodDialysateRateChangeResponse.accepted:0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeBloodDialysateRateChangeResponse.reason:10 +TEST.VALUE:uut_prototype_stubs.setTreatmentParameterU32.return:1 +TEST.VALUE:uut_prototype_stubs.getTreatmentParameterU32.return:150 +TEST.VALUE:uut_prototype_stubs.getU32TreatmentParamLowerRangeLimit.return:100 +TEST.VALUE:uut_prototype_stubs.getU32TreatmentParamUpperRangeLimit.return:500 +TEST.VALUE:uut_prototype_stubs.getF32TreatmentParamLowerRangeLimit.return:30.0 +TEST.VALUE:uut_prototype_stubs.getF32TreatmentParamUpperRangeLimit.return:35.0 +TEST.EXPECTED:ModeTreatment.verifyBloodAndDialysateRateSettingsChange.return:1 TEST.END --- Test Case: verifyBloodAndDialysateRateSettingsChange_NotInTreatmentMode +-- Test Case: verifyBloodAndDialysateRateSettingsChange_08_in_Tx_blood_dial_rate_eq_to_param TEST.UNIT:ModeTreatment TEST.SUBPROGRAM:verifyBloodAndDialysateRateSettingsChange TEST.NEW -TEST.NAME:verifyBloodAndDialysateRateSettingsChange_NotInTreatmentMode -TEST.IMPORT_FAILURES: -(E) Errors from previous script import(s) - >>> (E) @LINE: 930 TEST.VALUE:ModeTreatment.<>.presBloodFlowRate:200 - >>> >>> Unknown parameter/object name presBloodFlowRate - >>> >>> Value Line Error - Command Ignored - >>> (E) @LINE: 931 TEST.VALUE:ModeTreatment.<>.presDialysateFlowRate:300 - >>> >>> Unknown parameter/object name presDialysateFlowRate - >>> >>> Value Line Error - Command Ignored - >>> (E) @LINE: 939 TEST.EXPECTED:ModeTreatment.<>.presBloodFlowRate:200 - >>> >>> Unknown parameter/object name presBloodFlowRate - >>> >>> Value Line Error - Command Ignored - >>> (E) @LINE: 940 TEST.EXPECTED:ModeTreatment.<>.presDialysateFlowRate:300 - >>> >>> Unknown parameter/object name presDialysateFlowRate - >>> >>> Value Line Error - Command Ignored -TEST.END_IMPORT_FAILURES: -TEST.STUB:ModeTreatment.broadcastTreatmentSettingsRanges -TEST.VALUE:ModeTreatment.<>.currentTreatmentState:TREATMENT_DIALYSIS_STATE -TEST.VALUE:ModeTreatment.<>.presTreatmentTimeSecs:14400 -TEST.VALUE:ModeTreatment.<>.presMaxUFVolumeML:2400.0 -TEST.VALUE:ModeTreatment.<>.presUFRate:10.0 +TEST.NAME:verifyBloodAndDialysateRateSettingsChange_08_in_Tx_blood_dial_rate_eq_to_param +TEST.VALUE:ModeTreatment.<>.presTreatmentTimeSecs:60 +TEST.VALUE:ModeTreatment.<>.presUFRate:2.0 TEST.VALUE:ModeTreatment.verifyBloodAndDialysateRateSettingsChange.bloodRate:300 -TEST.VALUE:ModeTreatment.verifyBloodAndDialysateRateSettingsChange.dialRate:400 -TEST.VALUE:uut_prototype_stubs.getCurrentOperationMode.return:MODE_STAN -TEST.VALUE:uut_prototype_stubs.sendChangeBloodDialysateRateChangeResponse.return:1 -TEST.EXPECTED:ModeTreatment.<>.presTreatmentTimeSecs:14400 -TEST.EXPECTED:ModeTreatment.<>.presMaxUFVolumeML:2400.0 -TEST.EXPECTED:ModeTreatment.<>.presUFRate:10.0 -TEST.EXPECTED:ModeTreatment.verifyBloodAndDialysateRateSettingsChange.return:0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeBloodDialysateRateChangeResponse.accepted:0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeBloodDialysateRateChangeResponse.reason:3 +TEST.VALUE:ModeTreatment.verifyBloodAndDialysateRateSettingsChange.dialRate:300 +TEST.VALUE:uut_prototype_stubs.getUltrafiltrationReferenceVolume.return:2.0 +TEST.VALUE:uut_prototype_stubs.getCurrentOperationMode.return:MODE_TREA +TEST.VALUE:uut_prototype_stubs.setTreatmentParameterU32.return:1 +TEST.VALUE:uut_prototype_stubs.getTreatmentParameterU32.return:300 +TEST.VALUE:uut_prototype_stubs.getU32TreatmentParamLowerRangeLimit.return:100 +TEST.VALUE:uut_prototype_stubs.getU32TreatmentParamUpperRangeLimit.return:500 +TEST.VALUE:uut_prototype_stubs.getF32TreatmentParamLowerRangeLimit.return:30.0 +TEST.VALUE:uut_prototype_stubs.getF32TreatmentParamUpperRangeLimit.return:35.0 +TEST.EXPECTED:ModeTreatment.verifyBloodAndDialysateRateSettingsChange.return:1 TEST.END -- Subprogram: verifyPressureLimitsChange @@ -3812,448 +3508,55 @@ -- Subprogram: verifyTreatmentDurationSettingChange --- Test Case: verifyTreatmentDurationSettingChange.001 +-- Test Case: verifyTreatmentDurationSettingChange.01_not_in_Tx TEST.UNIT:ModeTreatment TEST.SUBPROGRAM:verifyTreatmentDurationSettingChange TEST.NEW -TEST.NAME:verifyTreatmentDurationSettingChange.001 -TEST.VALUE:ModeTreatment.<>.currentTreatmentState:TREATMENT_DIALYSIS_STATE -TEST.VALUE:ModeTreatment.<>.presTreatmentTimeSecs:3600 -TEST.VALUE:ModeTreatment.<>.presMaxUFVolumeML:1500.0 -TEST.VALUE:ModeTreatment.<>.presUFRate:10.0 -TEST.VALUE:ModeTreatment.<>.treatmentTimeMS:3000000 -TEST.VALUE:ModeTreatment.verifyTreatmentDurationSettingChange.treatmentTime:480 -TEST.VALUE:ModeTreatment.verifyTreatmentDurationSettingChange.return:0 -TEST.VALUE:uut_prototype_stubs.getUltrafiltrationVolumeCollected.return:1000.0 -TEST.VALUE:uut_prototype_stubs.getCurrentOperationMode.return:MODE_TREA -TEST.VALUE:uut_prototype_stubs.getTreatmentParameterU32.return:160000 -TEST.VALUE:uut_prototype_stubs.sendChangeTreatmentDurationResponse.accepted:0 -TEST.VALUE:uut_prototype_stubs.sendChangeTreatmentDurationResponse.reason:10 -TEST.VALUE:uut_prototype_stubs.sendChangeTreatmentDurationResponse.time_min:60 -TEST.VALUE:uut_prototype_stubs.sendChangeTreatmentDurationResponse.volume_mL:1500.0 -TEST.VALUE:uut_prototype_stubs.sendChangeTreatmentDurationResponse.return:1 +TEST.NAME:verifyTreatmentDurationSettingChange.01_not_in_Tx +TEST.VALUE:ModeTreatment.<>.presTreatmentTimeSecs:60 +TEST.VALUE:ModeTreatment.<>.presUFRate:2.0 +TEST.VALUE:ModeTreatment.<>.treatmentParamRangesPublishInterval.data:50 +TEST.VALUE:ModeTreatment.<>.treatmentParamRangesPublishInterval.override:0 +TEST.VALUE:uut_prototype_stubs.getUltrafiltrationReferenceVolume.return:2.0 +TEST.VALUE:uut_prototype_stubs.getCurrentOperationMode.return:MODE_SERV +TEST.VALUE:uut_prototype_stubs.getTreatmentParameterU32.return:150 TEST.EXPECTED:ModeTreatment.verifyTreatmentDurationSettingChange.return:0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeTreatmentDurationResponse.accepted:0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeTreatmentDurationResponse.reason:10 -TEST.EXPECTED:uut_prototype_stubs.sendChangeTreatmentDurationResponse.time_min:60 -TEST.EXPECTED:uut_prototype_stubs.sendChangeTreatmentDurationResponse.volume_mL:1500.0 TEST.END --- Test Case: verifyTreatmentDurationSettingChange_Accepted +-- Test Case: verifyTreatmentDurationSettingChange.02_in_Tx_not_in_Tx_state TEST.UNIT:ModeTreatment TEST.SUBPROGRAM:verifyTreatmentDurationSettingChange TEST.NEW -TEST.NAME:verifyTreatmentDurationSettingChange_Accepted -TEST.IMPORT_FAILURES: -(E) Errors from previous script import(s) - >>> (E) @LINE: 3840 TEST.EXPECTED:uut_prototype_stubs.setDialysisParams.uFRate:10.0 - >>> >>> Could not find function setDialysisParams - >>> >>> in unit uut_prototype_stubs. - >>> >>> Value Line Error - Command Ignored - >>> (E) @LINE: 1139 TEST.VALUE:ModeTreatment.<>.presBloodFlowRate:200 - >>> >>> Unknown parameter/object name presBloodFlowRate - >>> >>> Value Line Error - Command Ignored - >>> (E) @LINE: 1140 TEST.VALUE:ModeTreatment.<>.presDialysateFlowRate:300 - >>> >>> Unknown parameter/object name presDialysateFlowRate - >>> >>> Value Line Error - Command Ignored - >>> (E) @LINE: 1149 TEST.EXPECTED:ModeTreatment.<>.presBloodFlowRate:200 - >>> >>> Unknown parameter/object name presBloodFlowRate - >>> >>> Value Line Error - Command Ignored - >>> (E) @LINE: 1150 TEST.EXPECTED:ModeTreatment.<>.presDialysateFlowRate:300 - >>> >>> Unknown parameter/object name presDialysateFlowRate - >>> >>> Value Line Error - Command Ignored -TEST.END_IMPORT_FAILURES: -TEST.VALUE:ModeTreatment.<>.currentTreatmentState:TREATMENT_DIALYSIS_STATE -TEST.VALUE:ModeTreatment.<>.presTreatmentTimeSecs:3600 -TEST.VALUE:ModeTreatment.<>.presMaxUFVolumeML:1500.0 -TEST.VALUE:ModeTreatment.<>.presUFRate:10.0 -TEST.VALUE:ModeTreatment.<>.treatmentTimeMS:3000000 -TEST.VALUE:ModeTreatment.verifyTreatmentDurationSettingChange.treatmentTime:120 -TEST.VALUE:uut_prototype_stubs.getUltrafiltrationVolumeCollected.return:1000.0 +TEST.NAME:verifyTreatmentDurationSettingChange.02_in_Tx_not_in_Tx_state +TEST.VALUE:ModeTreatment.<>.currentTreatmentState:NUM_OF_TREATMENT_STATES +TEST.VALUE:ModeTreatment.<>.presTreatmentTimeSecs:60 +TEST.VALUE:ModeTreatment.<>.presUFRate:2.0 +TEST.VALUE:ModeTreatment.<>.treatmentParamRangesPublishInterval.data:50 +TEST.VALUE:ModeTreatment.<>.treatmentParamRangesPublishInterval.override:0 +TEST.VALUE:uut_prototype_stubs.getUltrafiltrationReferenceVolume.return:2.0 TEST.VALUE:uut_prototype_stubs.getCurrentOperationMode.return:MODE_TREA -TEST.VALUE:uut_prototype_stubs.getTreatmentParameterU32.return:1 -TEST.VALUE:uut_prototype_stubs.sendChangeTreatmentDurationResponse.return:1 -TEST.EXPECTED:uut_prototype_stubs.sendChangeTreatmentDurationResponse.accepted:1 -TEST.EXPECTED:uut_prototype_stubs.sendChangeTreatmentDurationResponse.reason:0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeTreatmentDurationResponse.time_min:120 +TEST.VALUE:uut_prototype_stubs.getTreatmentParameterU32.return:150 +TEST.EXPECTED:ModeTreatment.verifyTreatmentDurationSettingChange.return:0 TEST.END --- Test Case: verifyTreatmentDurationSettingChange_DialVolOutOfRange +-- Test Case: verifyTreatmentDurationSettingChange.02_in_Tx_not_in_Tx_state.001 TEST.UNIT:ModeTreatment TEST.SUBPROGRAM:verifyTreatmentDurationSettingChange TEST.NEW -TEST.NAME:verifyTreatmentDurationSettingChange_DialVolOutOfRange -TEST.IMPORT_FAILURES: -(I) @LINE: 3957 - >>> 'REQUEST_REJECT_REASON_NONE' was specified as a macro, but it is in the symbol dictionary. -(E) Errors from previous script import(s) - >>> (I) @LINE: 3923 - >>> >>> 'REQUEST_REJECT_REASON_NONE' was specified as a macro, but it is in the symbol dictionary. - >>> (I) @LINE: 3921 - >>> >>> 'REQUEST_REJECT_REASON_NONE' was specified as a macro, but it is in the symbol dictionary. - >>> (I) @LINE: 90 - >>> >>> 'REQUEST_REJECT_REASON_NONE' was specified as a macro, but it is in the symbol dictionary. - >>> (I) @LINE: 3906 - >>> >>> 'REQUEST_REJECT_REASON_NONE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 3902 ->>> >>> 'REQUEST_REJECT_REASON_NONE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 3902 ->>> >>> 'REQUEST_REJECT_REASON_NONE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 3865 ->>> >>> 'REQUEST_REJECT_REASON_NONE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 3862 ->>> >>> 'REQUEST_REJECT_REASON_NONE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 3623 ->>> >>> 'REQUEST_REJECT_REASON_NONE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 3244 ->>> >>> 'REQUEST_REJECT_REASON_NONE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 3242 ->>> >>> 'REQUEST_REJECT_REASON_NONE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 3240 ->>> >>> 'REQUEST_REJECT_REASON_NONE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 3335 ->>> >>> 'REQUEST_REJECT_REASON_NONE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 3321 ->>> >>> 'REQUEST_REJECT_REASON_NONE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 3319 ->>> >>> 'REQUEST_REJECT_REASON_NONE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 3317 ->>> >>> 'REQUEST_REJECT_REASON_NONE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 3299 ->>> >>> 'REQUEST_REJECT_REASON_NONE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 3266 ->>> >>> 'REQUEST_REJECT_REASON_NONE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 3264 ->>> >>> 'REQUEST_REJECT_REASON_NONE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 3262 ->>> >>> 'REQUEST_REJECT_REASON_NONE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 3213 ->>> >>> 'REQUEST_REJECT_REASON_NONE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 3211 ->>> >>> 'REQUEST_REJECT_REASON_NONE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 3182 ->>> >>> 'REQUEST_REJECT_REASON_NONE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 3120 ->>> >>> 'REQUEST_REJECT_REASON_NONE' was specified as a macro, but it is in the symbol dictionary. ->>> (E) @LINE: 1172 TEST.VALUE:ModeTreatment.<>.presBloodFlowRate:200 ->>> >>> Unknown parameter/object name presBloodFlowRate ->>> >>> Value Line Error - Command Ignored ->>> (E) @LINE: 1173 TEST.VALUE:ModeTreatment.<>.presDialysateFlowRate:400 ->>> >>> Unknown parameter/object name presDialysateFlowRate ->>> >>> Value Line Error - Command Ignored ->>> (E) @LINE: 1182 TEST.EXPECTED:ModeTreatment.<>.presBloodFlowRate:200 ->>> >>> Unknown parameter/object name presBloodFlowRate ->>> >>> Value Line Error - Command Ignored ->>> (E) @LINE: 1183 TEST.EXPECTED:ModeTreatment.<>.presDialysateFlowRate:400 ->>> >>> Unknown parameter/object name presDialysateFlowRate ->>> >>> Value Line Error - Command Ignored -TEST.END_IMPORT_FAILURES: +TEST.NAME:verifyTreatmentDurationSettingChange.02_in_Tx_not_in_Tx_state.001 TEST.VALUE:ModeTreatment.<>.currentTreatmentState:TREATMENT_DIALYSIS_STATE -TEST.VALUE:ModeTreatment.<>.presTreatmentTimeSecs:3600 -TEST.VALUE:ModeTreatment.<>.presMaxUFVolumeML:1500.0 -TEST.VALUE:ModeTreatment.<>.presUFRate:10.0 -TEST.VALUE:ModeTreatment.<>.treatmentTimeMS:3000000 -TEST.VALUE:ModeTreatment.verifyTreatmentDurationSettingChange.treatmentTime:480 -TEST.VALUE:ModeTreatment.verifyTreatmentDurationSettingChange.return:1 -TEST.VALUE:uut_prototype_stubs.getUltrafiltrationVolumeCollected.return:1000.0 +TEST.VALUE:ModeTreatment.<>.presTreatmentTimeSecs:60 +TEST.VALUE:ModeTreatment.<>.presUFRate:2.0 +TEST.VALUE:ModeTreatment.<>.treatmentParamRangesPublishInterval.data:50 +TEST.VALUE:ModeTreatment.<>.treatmentParamRangesPublishInterval.override:0 +TEST.VALUE:uut_prototype_stubs.getUltrafiltrationReferenceVolume.return:2.0 TEST.VALUE:uut_prototype_stubs.getCurrentOperationMode.return:MODE_TREA -TEST.VALUE:uut_prototype_stubs.getTreatmentParameterU32.return:1 -TEST.VALUE:uut_prototype_stubs.sendChangeTreatmentDurationResponse.accepted:1 -TEST.VALUE:uut_prototype_stubs.sendChangeTreatmentDurationResponse.reason:MACRO=REQUEST_REJECT_REASON_NONE -TEST.VALUE:uut_prototype_stubs.sendChangeTreatmentDurationResponse.time_min:480 -TEST.VALUE:uut_prototype_stubs.sendChangeTreatmentDurationResponse.volume_mL:4300.0 -TEST.VALUE:uut_prototype_stubs.sendChangeTreatmentDurationResponse.return:1 -TEST.EXPECTED:ModeTreatment.verifyTreatmentDurationSettingChange.return:1 -TEST.EXPECTED:uut_prototype_stubs.sendChangeTreatmentDurationResponse.accepted:1 -TEST.EXPECTED:uut_prototype_stubs.sendChangeTreatmentDurationResponse.time_min:480 -TEST.EXPECTED:uut_prototype_stubs.sendChangeTreatmentDurationResponse.volume_mL:4300.0 +TEST.VALUE:uut_prototype_stubs.getTreatmentParameterU32.return:150 +TEST.EXPECTED:ModeTreatment.verifyTreatmentDurationSettingChange.return:0 TEST.END --- Test Case: verifyTreatmentDurationSettingChange_DurationToLow -TEST.UNIT:ModeTreatment -TEST.SUBPROGRAM:verifyTreatmentDurationSettingChange -TEST.NEW -TEST.NAME:verifyTreatmentDurationSettingChange_DurationToLow -TEST.IMPORT_FAILURES: -(E) Errors from previous script import(s) - >>> (E) @LINE: 1201 TEST.VALUE:ModeTreatment.<>.presBloodFlowRate:200 - >>> >>> Unknown parameter/object name presBloodFlowRate - >>> >>> Value Line Error - Command Ignored - >>> (E) @LINE: 1202 TEST.VALUE:ModeTreatment.<>.presDialysateFlowRate:400 - >>> >>> Unknown parameter/object name presDialysateFlowRate - >>> >>> Value Line Error - Command Ignored - >>> (E) @LINE: 1211 TEST.EXPECTED:ModeTreatment.<>.presBloodFlowRate:200 - >>> >>> Unknown parameter/object name presBloodFlowRate - >>> >>> Value Line Error - Command Ignored - >>> (E) @LINE: 1212 TEST.EXPECTED:ModeTreatment.<>.presDialysateFlowRate:400 - >>> >>> Unknown parameter/object name presDialysateFlowRate - >>> >>> Value Line Error - Command Ignored -TEST.END_IMPORT_FAILURES: -TEST.VALUE:ModeTreatment.<>.currentTreatmentState:TREATMENT_DIALYSIS_STATE -TEST.VALUE:ModeTreatment.<>.presTreatmentTimeSecs:3600 -TEST.VALUE:ModeTreatment.<>.presMaxUFVolumeML:1500.0 -TEST.VALUE:ModeTreatment.<>.presUFRate:10.0 -TEST.VALUE:ModeTreatment.<>.treatmentTimeMS:3000000 -TEST.VALUE:ModeTreatment.verifyTreatmentDurationSettingChange.treatmentTime:59 -TEST.VALUE:uut_prototype_stubs.getUltrafiltrationVolumeCollected.return:1000.0 -TEST.VALUE:uut_prototype_stubs.getCurrentOperationMode.return:MODE_TREA -TEST.VALUE:uut_prototype_stubs.getTreatmentParameterU32.return:1 -TEST.VALUE:uut_prototype_stubs.sendChangeTreatmentDurationResponse.return:1 -TEST.EXPECTED:uut_prototype_stubs.sendChangeTreatmentDurationResponse.accepted:0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeTreatmentDurationResponse.reason:13 -TEST.EXPECTED:uut_prototype_stubs.sendChangeTreatmentDurationResponse.time_min:60 -TEST.EXPECTED:uut_prototype_stubs.sendChangeTreatmentDurationResponse.volume_mL:1500.0 -TEST.END - --- Test Case: verifyTreatmentDurationSettingChange_InvalidTreatmentState1 -TEST.UNIT:ModeTreatment -TEST.SUBPROGRAM:verifyTreatmentDurationSettingChange -TEST.NEW -TEST.NAME:verifyTreatmentDurationSettingChange_InvalidTreatmentState1 -TEST.IMPORT_FAILURES: -(E) Errors from previous script import(s) - >>> (E) @LINE: 1230 TEST.VALUE:ModeTreatment.<>.presBloodFlowRate:200 - >>> >>> Unknown parameter/object name presBloodFlowRate - >>> >>> Value Line Error - Command Ignored - >>> (E) @LINE: 1231 TEST.VALUE:ModeTreatment.<>.presDialysateFlowRate:400 - >>> >>> Unknown parameter/object name presDialysateFlowRate - >>> >>> Value Line Error - Command Ignored - >>> (E) @LINE: 1240 TEST.EXPECTED:ModeTreatment.<>.presBloodFlowRate:200 - >>> >>> Unknown parameter/object name presBloodFlowRate - >>> >>> Value Line Error - Command Ignored - >>> (E) @LINE: 1241 TEST.EXPECTED:ModeTreatment.<>.presDialysateFlowRate:400 - >>> >>> Unknown parameter/object name presDialysateFlowRate - >>> >>> Value Line Error - Command Ignored -TEST.END_IMPORT_FAILURES: -TEST.VALUE:ModeTreatment.<>.currentTreatmentState:TREATMENT_START_STATE -TEST.VALUE:ModeTreatment.<>.presTreatmentTimeSecs:3600 -TEST.VALUE:ModeTreatment.<>.presMaxUFVolumeML:1500.0 -TEST.VALUE:ModeTreatment.<>.presUFRate:10.0 -TEST.VALUE:ModeTreatment.<>.treatmentTimeMS:3000000 -TEST.VALUE:ModeTreatment.verifyTreatmentDurationSettingChange.treatmentTime:120 -TEST.VALUE:uut_prototype_stubs.getUltrafiltrationVolumeCollected.return:1000.0 -TEST.VALUE:uut_prototype_stubs.getCurrentOperationMode.return:MODE_TREA -TEST.VALUE:uut_prototype_stubs.getTreatmentParameterU32.return:1 -TEST.VALUE:uut_prototype_stubs.sendChangeTreatmentDurationResponse.return:1 -TEST.EXPECTED:uut_prototype_stubs.sendChangeTreatmentDurationResponse.accepted:0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeTreatmentDurationResponse.reason:4 -TEST.EXPECTED:uut_prototype_stubs.sendChangeTreatmentDurationResponse.time_min:60 -TEST.EXPECTED:uut_prototype_stubs.sendChangeTreatmentDurationResponse.volume_mL:1500.0 -TEST.END - --- Test Case: verifyTreatmentDurationSettingChange_InvalidTreatmentState2 -TEST.UNIT:ModeTreatment -TEST.SUBPROGRAM:verifyTreatmentDurationSettingChange -TEST.NEW -TEST.NAME:verifyTreatmentDurationSettingChange_InvalidTreatmentState2 -TEST.IMPORT_FAILURES: -(E) Errors from previous script import(s) - >>> (E) @LINE: 1259 TEST.VALUE:ModeTreatment.<>.presBloodFlowRate:200 - >>> >>> Unknown parameter/object name presBloodFlowRate - >>> >>> Value Line Error - Command Ignored - >>> (E) @LINE: 1260 TEST.VALUE:ModeTreatment.<>.presDialysateFlowRate:400 - >>> >>> Unknown parameter/object name presDialysateFlowRate - >>> >>> Value Line Error - Command Ignored - >>> (E) @LINE: 1269 TEST.EXPECTED:ModeTreatment.<>.presBloodFlowRate:200 - >>> >>> Unknown parameter/object name presBloodFlowRate - >>> >>> Value Line Error - Command Ignored - >>> (E) @LINE: 1270 TEST.EXPECTED:ModeTreatment.<>.presDialysateFlowRate:400 - >>> >>> Unknown parameter/object name presDialysateFlowRate - >>> >>> Value Line Error - Command Ignored -TEST.END_IMPORT_FAILURES: -TEST.VALUE:ModeTreatment.<>.currentTreatmentState:TREATMENT_END_STATE -TEST.VALUE:ModeTreatment.<>.presTreatmentTimeSecs:3600 -TEST.VALUE:ModeTreatment.<>.presMaxUFVolumeML:1500.0 -TEST.VALUE:ModeTreatment.<>.presUFRate:10.0 -TEST.VALUE:ModeTreatment.<>.treatmentTimeMS:3000000 -TEST.VALUE:ModeTreatment.verifyTreatmentDurationSettingChange.treatmentTime:120 -TEST.VALUE:uut_prototype_stubs.getUltrafiltrationVolumeCollected.return:1000.0 -TEST.VALUE:uut_prototype_stubs.getCurrentOperationMode.return:MODE_TREA -TEST.VALUE:uut_prototype_stubs.getTreatmentParameterU32.return:1 -TEST.VALUE:uut_prototype_stubs.sendChangeTreatmentDurationResponse.return:1 -TEST.EXPECTED:uut_prototype_stubs.sendChangeTreatmentDurationResponse.accepted:0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeTreatmentDurationResponse.reason:4 -TEST.EXPECTED:uut_prototype_stubs.sendChangeTreatmentDurationResponse.time_min:60 -TEST.EXPECTED:uut_prototype_stubs.sendChangeTreatmentDurationResponse.volume_mL:1500.0 -TEST.END - --- Test Case: verifyTreatmentDurationSettingChange_LessThanElapsed -TEST.UNIT:ModeTreatment -TEST.SUBPROGRAM:verifyTreatmentDurationSettingChange -TEST.NEW -TEST.NAME:verifyTreatmentDurationSettingChange_LessThanElapsed -TEST.IMPORT_FAILURES: -(E) Errors from previous script import(s) - >>> (E) @LINE: 1288 TEST.VALUE:ModeTreatment.<>.presBloodFlowRate:200 - >>> >>> Unknown parameter/object name presBloodFlowRate - >>> >>> Value Line Error - Command Ignored - >>> (E) @LINE: 1289 TEST.VALUE:ModeTreatment.<>.presDialysateFlowRate:400 - >>> >>> Unknown parameter/object name presDialysateFlowRate - >>> >>> Value Line Error - Command Ignored - >>> (E) @LINE: 1298 TEST.EXPECTED:ModeTreatment.<>.presBloodFlowRate:200 - >>> >>> Unknown parameter/object name presBloodFlowRate - >>> >>> Value Line Error - Command Ignored - >>> (E) @LINE: 1299 TEST.EXPECTED:ModeTreatment.<>.presDialysateFlowRate:400 - >>> >>> Unknown parameter/object name presDialysateFlowRate - >>> >>> Value Line Error - Command Ignored -TEST.END_IMPORT_FAILURES: -TEST.VALUE:ModeTreatment.<>.currentTreatmentState:TREATMENT_DIALYSIS_STATE -TEST.VALUE:ModeTreatment.<>.presTreatmentTimeSecs:7200 -TEST.VALUE:ModeTreatment.<>.presMaxUFVolumeML:1500.0 -TEST.VALUE:ModeTreatment.<>.presUFRate:10.0 -TEST.VALUE:ModeTreatment.<>.treatmentTimeMS:3700000 -TEST.VALUE:ModeTreatment.verifyTreatmentDurationSettingChange.treatmentTime:60 -TEST.VALUE:uut_prototype_stubs.getUltrafiltrationVolumeCollected.return:1000.0 -TEST.VALUE:uut_prototype_stubs.getCurrentOperationMode.return:MODE_TREA -TEST.VALUE:uut_prototype_stubs.getTreatmentParameterU32.return:1 -TEST.VALUE:uut_prototype_stubs.sendChangeTreatmentDurationResponse.return:1 -TEST.EXPECTED:uut_prototype_stubs.sendChangeTreatmentDurationResponse.accepted:0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeTreatmentDurationResponse.reason:7 -TEST.EXPECTED:uut_prototype_stubs.sendChangeTreatmentDurationResponse.time_min:120 -TEST.EXPECTED:uut_prototype_stubs.sendChangeTreatmentDurationResponse.volume_mL:1500.0 -TEST.END - --- Test Case: verifyTreatmentDurationSettingChange_NotInTreatmentMode -TEST.UNIT:ModeTreatment -TEST.SUBPROGRAM:verifyTreatmentDurationSettingChange -TEST.NEW -TEST.NAME:verifyTreatmentDurationSettingChange_NotInTreatmentMode -TEST.IMPORT_FAILURES: -(E) Errors from previous script import(s) - >>> (E) @LINE: 1317 TEST.VALUE:ModeTreatment.<>.presBloodFlowRate:200 - >>> >>> Unknown parameter/object name presBloodFlowRate - >>> >>> Value Line Error - Command Ignored - >>> (E) @LINE: 1318 TEST.VALUE:ModeTreatment.<>.presDialysateFlowRate:400 - >>> >>> Unknown parameter/object name presDialysateFlowRate - >>> >>> Value Line Error - Command Ignored - >>> (E) @LINE: 1327 TEST.EXPECTED:ModeTreatment.<>.presBloodFlowRate:200 - >>> >>> Unknown parameter/object name presBloodFlowRate - >>> >>> Value Line Error - Command Ignored - >>> (E) @LINE: 1328 TEST.EXPECTED:ModeTreatment.<>.presDialysateFlowRate:400 - >>> >>> Unknown parameter/object name presDialysateFlowRate - >>> >>> Value Line Error - Command Ignored -TEST.END_IMPORT_FAILURES: -TEST.VALUE:ModeTreatment.<>.currentTreatmentState:TREATMENT_DIALYSIS_STATE -TEST.VALUE:ModeTreatment.<>.presTreatmentTimeSecs:3600 -TEST.VALUE:ModeTreatment.<>.presMaxUFVolumeML:1500.0 -TEST.VALUE:ModeTreatment.<>.presUFRate:10.0 -TEST.VALUE:ModeTreatment.<>.treatmentTimeMS:3000000 -TEST.VALUE:ModeTreatment.verifyTreatmentDurationSettingChange.treatmentTime:120 -TEST.VALUE:uut_prototype_stubs.getUltrafiltrationVolumeCollected.return:1000.0 -TEST.VALUE:uut_prototype_stubs.getCurrentOperationMode.return:MODE_STAN -TEST.VALUE:uut_prototype_stubs.getTreatmentParameterU32.return:1 -TEST.VALUE:uut_prototype_stubs.sendChangeTreatmentDurationResponse.return:1 -TEST.EXPECTED:uut_prototype_stubs.sendChangeTreatmentDurationResponse.accepted:0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeTreatmentDurationResponse.reason:3 -TEST.EXPECTED:uut_prototype_stubs.sendChangeTreatmentDurationResponse.time_min:60 -TEST.EXPECTED:uut_prototype_stubs.sendChangeTreatmentDurationResponse.volume_mL:1500.0 -TEST.END - --- Test Case: verifyTreatmentDurationSettingChange_TimeOutOfRange -TEST.UNIT:ModeTreatment -TEST.SUBPROGRAM:verifyTreatmentDurationSettingChange -TEST.NEW -TEST.NAME:verifyTreatmentDurationSettingChange_TimeOutOfRange -TEST.IMPORT_FAILURES: -(E) Errors from previous script import(s) - >>> (E) @LINE: 1346 TEST.VALUE:ModeTreatment.<>.presBloodFlowRate:200 - >>> >>> Unknown parameter/object name presBloodFlowRate - >>> >>> Value Line Error - Command Ignored - >>> (E) @LINE: 1347 TEST.VALUE:ModeTreatment.<>.presDialysateFlowRate:300 - >>> >>> Unknown parameter/object name presDialysateFlowRate - >>> >>> Value Line Error - Command Ignored - >>> (E) @LINE: 1356 TEST.EXPECTED:ModeTreatment.<>.presBloodFlowRate:200 - >>> >>> Unknown parameter/object name presBloodFlowRate - >>> >>> Value Line Error - Command Ignored - >>> (E) @LINE: 1357 TEST.EXPECTED:ModeTreatment.<>.presDialysateFlowRate:300 - >>> >>> Unknown parameter/object name presDialysateFlowRate - >>> >>> Value Line Error - Command Ignored -TEST.END_IMPORT_FAILURES: -TEST.VALUE:ModeTreatment.<>.currentTreatmentState:TREATMENT_DIALYSIS_STATE -TEST.VALUE:ModeTreatment.<>.presTreatmentTimeSecs:3600 -TEST.VALUE:ModeTreatment.<>.presMaxUFVolumeML:1500.0 -TEST.VALUE:ModeTreatment.<>.presUFRate:10.0 -TEST.VALUE:ModeTreatment.<>.treatmentTimeMS:3000000 -TEST.VALUE:ModeTreatment.verifyTreatmentDurationSettingChange.treatmentTime:485 -TEST.VALUE:uut_prototype_stubs.getUltrafiltrationVolumeCollected.return:1000.0 -TEST.VALUE:uut_prototype_stubs.getCurrentOperationMode.return:MODE_TREA -TEST.VALUE:uut_prototype_stubs.getTreatmentParameterU32.return:1 -TEST.VALUE:uut_prototype_stubs.sendChangeTreatmentDurationResponse.return:1 -TEST.EXPECTED:uut_prototype_stubs.sendChangeTreatmentDurationResponse.accepted:0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeTreatmentDurationResponse.reason:6 -TEST.EXPECTED:uut_prototype_stubs.sendChangeTreatmentDurationResponse.time_min:60 -TEST.EXPECTED:uut_prototype_stubs.sendChangeTreatmentDurationResponse.volume_mL:1500.0 -TEST.END - --- Test Case: verifyTreatmentDurationSettingChange_UFVolOutOfRange -TEST.UNIT:ModeTreatment -TEST.SUBPROGRAM:verifyTreatmentDurationSettingChange -TEST.NEW -TEST.NAME:verifyTreatmentDurationSettingChange_UFVolOutOfRange -TEST.IMPORT_FAILURES: -(E) Errors from previous script import(s) - >>> (E) @LINE: 1375 TEST.VALUE:ModeTreatment.<>.presBloodFlowRate:200 - >>> >>> Unknown parameter/object name presBloodFlowRate - >>> >>> Value Line Error - Command Ignored - >>> (E) @LINE: 1376 TEST.VALUE:ModeTreatment.<>.presDialysateFlowRate:300 - >>> >>> Unknown parameter/object name presDialysateFlowRate - >>> >>> Value Line Error - Command Ignored - >>> (E) @LINE: 1385 TEST.EXPECTED:ModeTreatment.<>.presBloodFlowRate:200 - >>> >>> Unknown parameter/object name presBloodFlowRate - >>> >>> Value Line Error - Command Ignored - >>> (E) @LINE: 1386 TEST.EXPECTED:ModeTreatment.<>.presDialysateFlowRate:300 - >>> >>> Unknown parameter/object name presDialysateFlowRate - >>> >>> Value Line Error - Command Ignored -TEST.END_IMPORT_FAILURES: -TEST.VALUE:ModeTreatment.<>.currentTreatmentState:TREATMENT_DIALYSIS_STATE -TEST.VALUE:ModeTreatment.<>.presTreatmentTimeSecs:3600 -TEST.VALUE:ModeTreatment.<>.presMaxUFVolumeML:1500.0 -TEST.VALUE:ModeTreatment.<>.presUFRate:40.0 -TEST.VALUE:ModeTreatment.<>.treatmentTimeMS:3000000 -TEST.VALUE:ModeTreatment.verifyTreatmentDurationSettingChange.treatmentTime:300 -TEST.VALUE:uut_prototype_stubs.getUltrafiltrationVolumeCollected.return:1000.0 -TEST.VALUE:uut_prototype_stubs.getCurrentOperationMode.return:MODE_TREA -TEST.VALUE:uut_prototype_stubs.getTreatmentParameterU32.return:1 -TEST.VALUE:uut_prototype_stubs.sendChangeTreatmentDurationResponse.return:1 -TEST.EXPECTED:uut_prototype_stubs.sendChangeTreatmentDurationResponse.accepted:0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeTreatmentDurationResponse.reason:11 -TEST.EXPECTED:uut_prototype_stubs.sendChangeTreatmentDurationResponse.time_min:60 -TEST.EXPECTED:uut_prototype_stubs.sendChangeTreatmentDurationResponse.volume_mL:1500.0 -TEST.END - -- Subprogram: verifyUFSettingsChange --- Test Case: verifyUFSettingsChange.001 -TEST.UNIT:ModeTreatment -TEST.SUBPROGRAM:verifyUFSettingsChange -TEST.NEW -TEST.NAME:verifyUFSettingsChange.001 -TEST.VALUE:ModeTreatment.<>.currentTreatmentState:TREATMENT_BLOOD_PRIME_STATE -TEST.VALUE:ModeTreatment.<>.presTreatmentTimeSecs:600 -TEST.VALUE:ModeTreatment.<>.presUFRate:1.0 -TEST.VALUE:ModeTreatment.verifyUFSettingsChange.uFVolume:120.0 -TEST.VALUE:ModeTreatment.verifyUFSettingsChange.return:1 -TEST.VALUE:uut_prototype_stubs.getCurrentOperationMode.return:MODE_TREA -TEST.VALUE:uut_prototype_stubs.getTreatmentParameterU32.return:130000 -TEST.VALUE:uut_prototype_stubs.sendChangeUFSettingsResponse.accepted:1 -TEST.VALUE:uut_prototype_stubs.sendChangeUFSettingsResponse.reason:0 -TEST.VALUE:uut_prototype_stubs.sendChangeUFSettingsResponse.volume_mL:120.0 -TEST.VALUE:uut_prototype_stubs.sendChangeUFSettingsResponse.time_min:0 -TEST.VALUE:uut_prototype_stubs.sendChangeUFSettingsResponse.ufRate_mL_min:12.0 -TEST.VALUE:uut_prototype_stubs.sendChangeUFSettingsResponse.timeDiff:0 -TEST.VALUE:uut_prototype_stubs.sendChangeUFSettingsResponse.rateDiff:11.0 -TEST.VALUE:uut_prototype_stubs.sendChangeUFSettingsResponse.oldUFRate_mL_min:1.0 -TEST.EXPECTED:ModeTreatment.verifyUFSettingsChange.return:1 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsResponse.accepted:1 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsResponse.reason:0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsResponse.volume_mL:120.0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsResponse.time_min:0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsResponse.ufRate_mL_min:12.0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsResponse.timeDiff:0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsResponse.rateDiff:11.0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsResponse.oldUFRate_mL_min:1.0 -TEST.END - -- Test Case: verifyUFSettingsChange.002 TEST.UNIT:ModeTreatment TEST.SUBPROGRAM:verifyUFSettingsChange @@ -4296,37 +3599,6 @@ TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsResponse.oldUFRate_mL_min:0.0 TEST.END --- Test Case: verifyUFSettingsChange.003 -TEST.UNIT:ModeTreatment -TEST.SUBPROGRAM:verifyUFSettingsChange -TEST.NEW -TEST.NAME:verifyUFSettingsChange.003 -TEST.VALUE:ModeTreatment.<>.currentTreatmentState:TREATMENT_BLOOD_PRIME_STATE -TEST.VALUE:ModeTreatment.<>.presTreatmentTimeSecs:600 -TEST.VALUE:ModeTreatment.<>.presUFRate:1.0 -TEST.VALUE:ModeTreatment.verifyUFSettingsChange.uFVolume:120.0 -TEST.VALUE:ModeTreatment.verifyUFSettingsChange.return:1 -TEST.VALUE:uut_prototype_stubs.getCurrentOperationMode.return:MODE_TREA -TEST.VALUE:uut_prototype_stubs.getTreatmentParameterU32.return:130000 -TEST.VALUE:uut_prototype_stubs.sendChangeUFSettingsResponse.accepted:1 -TEST.VALUE:uut_prototype_stubs.sendChangeUFSettingsResponse.reason:0 -TEST.VALUE:uut_prototype_stubs.sendChangeUFSettingsResponse.volume_mL:120.0 -TEST.VALUE:uut_prototype_stubs.sendChangeUFSettingsResponse.time_min:0 -TEST.VALUE:uut_prototype_stubs.sendChangeUFSettingsResponse.ufRate_mL_min:12.0 -TEST.VALUE:uut_prototype_stubs.sendChangeUFSettingsResponse.timeDiff:0 -TEST.VALUE:uut_prototype_stubs.sendChangeUFSettingsResponse.rateDiff:11.0 -TEST.VALUE:uut_prototype_stubs.sendChangeUFSettingsResponse.oldUFRate_mL_min:1.0 -TEST.EXPECTED:ModeTreatment.verifyUFSettingsChange.return:1 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsResponse.accepted:1 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsResponse.reason:0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsResponse.volume_mL:120.0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsResponse.time_min:0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsResponse.ufRate_mL_min:12.0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsResponse.timeDiff:0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsResponse.rateDiff:11.0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsResponse.oldUFRate_mL_min:1.0 -TEST.END - -- Test Case: verifyUFSettingsChange.004 TEST.UNIT:ModeTreatment TEST.SUBPROGRAM:verifyUFSettingsChange @@ -4362,278 +3634,6 @@ TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsResponse.oldUFRate_mL_min:1.0 TEST.END --- Test Case: verifyUFSettingsChange_Accepted -TEST.UNIT:ModeTreatment -TEST.SUBPROGRAM:verifyUFSettingsChange -TEST.NEW -TEST.NAME:verifyUFSettingsChange_Accepted -TEST.IMPORT_FAILURES: -(E) Errors from previous script import(s) - >>> (E) @LINE: 1406 TEST.VALUE:ModeTreatment.<>.presBloodFlowRate:200 - >>> >>> Unknown parameter/object name presBloodFlowRate - >>> >>> Value Line Error - Command Ignored - >>> (E) @LINE: 1407 TEST.VALUE:ModeTreatment.<>.presDialysateFlowRate:300 - >>> >>> Unknown parameter/object name presDialysateFlowRate - >>> >>> Value Line Error - Command Ignored - >>> (E) @LINE: 1422 TEST.EXPECTED:ModeTreatment.<>.presBloodFlowRate:200 - >>> >>> Unknown parameter/object name presBloodFlowRate - >>> >>> Value Line Error - Command Ignored - >>> (E) @LINE: 1423 TEST.EXPECTED:ModeTreatment.<>.presDialysateFlowRate:300 - >>> >>> Unknown parameter/object name presDialysateFlowRate - >>> >>> Value Line Error - Command Ignored -TEST.END_IMPORT_FAILURES: -TEST.VALUE:ModeTreatment.<>.currentTreatmentState:TREATMENT_DIALYSIS_STATE -TEST.VALUE:ModeTreatment.<>.presTreatmentTimeSecs:3600 -TEST.VALUE:ModeTreatment.<>.presMaxUFVolumeML:600.0 -TEST.VALUE:ModeTreatment.<>.presUFRate:10.0 -TEST.VALUE:ModeTreatment.<>.treatmentTimeMS:2900000 -TEST.VALUE:ModeTreatment.<>.pendingParamChangesTimer:0 -TEST.VALUE:ModeTreatment.<>.pendingUFVolumeChange:0.0 -TEST.VALUE:ModeTreatment.<>.pendingUFRateChange:0.0 -TEST.VALUE:ModeTreatment.<>.pendingTreatmentTimeChange:0 -TEST.VALUE:ModeTreatment.verifyUFSettingsChange.uFVolume:1200.0 -TEST.VALUE:uut_prototype_stubs.getDialysisState.return:DIALYSIS_UF_STATE -TEST.VALUE:uut_prototype_stubs.getUltrafiltrationState.return:UF_PAUSED_STATE -TEST.VALUE:uut_prototype_stubs.getCurrentOperationMode.return:MODE_TREA -TEST.VALUE:uut_prototype_stubs.sendChangeUFSettingsResponse.return:1 -TEST.VALUE:uut_prototype_stubs.getMSTimerCount.return:3000000 -TEST.EXPECTED:ModeTreatment.<>.currentTreatmentState:TREATMENT_DIALYSIS_STATE -TEST.EXPECTED:ModeTreatment.<>.presTreatmentTimeSecs:3600 -TEST.EXPECTED:ModeTreatment.<>.presMaxUFVolumeML:600.0 -TEST.EXPECTED:ModeTreatment.<>.presUFRate:10.0 -TEST.EXPECTED:ModeTreatment.<>.treatmentTimeMS:2900000 -TEST.EXPECTED:ModeTreatment.<>.pendingParamChangesTimer:3000000 -TEST.EXPECTED:ModeTreatment.<>.pendingUFVolumeChange:600.0 -TEST.EXPECTED:ModeTreatment.<>.pendingUFRateChange:0.0 -TEST.EXPECTED:ModeTreatment.<>.pendingTreatmentTimeChange:0 -TEST.EXPECTED:ModeTreatment.verifyUFSettingsChange.uFVolume:1200.0 -TEST.EXPECTED:ModeTreatment.verifyUFSettingsChange.return:0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsResponse.accepted:0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsResponse.reason:12 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsResponse.volume_mL:600.0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsResponse.time_min:0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsResponse.ufRate_mL_min:0.0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsResponse.timeDiff:0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsResponse.rateDiff:0.0 -TEST.END - --- Test Case: verifyUFSettingsChange_Accepted_Bolus -TEST.UNIT:ModeTreatment -TEST.SUBPROGRAM:verifyUFSettingsChange -TEST.NEW -TEST.NAME:verifyUFSettingsChange_Accepted_Bolus -TEST.IMPORT_FAILURES: -(E) Errors from previous script import(s) - >>> (E) @LINE: 1447 TEST.VALUE:ModeTreatment.<>.presBloodFlowRate:200 - >>> >>> Unknown parameter/object name presBloodFlowRate - >>> >>> Value Line Error - Command Ignored - >>> (E) @LINE: 1448 TEST.VALUE:ModeTreatment.<>.presDialysateFlowRate:300 - >>> >>> Unknown parameter/object name presDialysateFlowRate - >>> >>> Value Line Error - Command Ignored - >>> (E) @LINE: 1462 TEST.EXPECTED:ModeTreatment.<>.presBloodFlowRate:200 - >>> >>> Unknown parameter/object name presBloodFlowRate - >>> >>> Value Line Error - Command Ignored - >>> (E) @LINE: 1463 TEST.EXPECTED:ModeTreatment.<>.presDialysateFlowRate:300 - >>> >>> Unknown parameter/object name presDialysateFlowRate - >>> >>> Value Line Error - Command Ignored -TEST.END_IMPORT_FAILURES: -TEST.VALUE:ModeTreatment.<>.currentTreatmentState:TREATMENT_DIALYSIS_STATE -TEST.VALUE:ModeTreatment.<>.presTreatmentTimeSecs:3600 -TEST.VALUE:ModeTreatment.<>.presMaxUFVolumeML:600.0 -TEST.VALUE:ModeTreatment.<>.presUFRate:10.0 -TEST.VALUE:ModeTreatment.<>.treatmentTimeMS:2900000 -TEST.VALUE:ModeTreatment.<>.pendingParamChangesTimer:0 -TEST.VALUE:ModeTreatment.<>.pendingUFVolumeChange:0.0 -TEST.VALUE:ModeTreatment.<>.pendingUFRateChange:0.0 -TEST.VALUE:ModeTreatment.<>.pendingTreatmentTimeChange:0 -TEST.VALUE:ModeTreatment.verifyUFSettingsChange.uFVolume:1200.0 -TEST.VALUE:uut_prototype_stubs.getUltrafiltrationState.return:UF_PAUSED_STATE -TEST.VALUE:uut_prototype_stubs.getCurrentOperationMode.return:MODE_TREA -TEST.VALUE:uut_prototype_stubs.sendChangeUFSettingsResponse.return:1 -TEST.VALUE:uut_prototype_stubs.getMSTimerCount.return:3000000 -TEST.EXPECTED:ModeTreatment.<>.currentTreatmentState:TREATMENT_DIALYSIS_STATE -TEST.EXPECTED:ModeTreatment.<>.presTreatmentTimeSecs:3600 -TEST.EXPECTED:ModeTreatment.<>.presMaxUFVolumeML:600.0 -TEST.EXPECTED:ModeTreatment.<>.presUFRate:10.0 -TEST.EXPECTED:ModeTreatment.<>.treatmentTimeMS:2900000 -TEST.EXPECTED:ModeTreatment.<>.pendingParamChangesTimer:3000000 -TEST.EXPECTED:ModeTreatment.<>.pendingUFVolumeChange:600.0 -TEST.EXPECTED:ModeTreatment.<>.pendingUFRateChange:0.0 -TEST.EXPECTED:ModeTreatment.<>.pendingTreatmentTimeChange:0 -TEST.EXPECTED:ModeTreatment.verifyUFSettingsChange.uFVolume:1200.0 -TEST.EXPECTED:ModeTreatment.verifyUFSettingsChange.return:0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsResponse.accepted:0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsResponse.reason:12 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsResponse.volume_mL:600.0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsResponse.time_min:0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsResponse.ufRate_mL_min:0.0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsResponse.timeDiff:0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsResponse.rateDiff:0.0 -TEST.END - --- Test Case: verifyUFSettingsChange_Accepted_NoTime1 -TEST.UNIT:ModeTreatment -TEST.SUBPROGRAM:verifyUFSettingsChange -TEST.NEW -TEST.NAME:verifyUFSettingsChange_Accepted_NoTime1 -TEST.IMPORT_FAILURES: -(E) Errors from previous script import(s) - >>> (E) @LINE: 1487 TEST.VALUE:ModeTreatment.<>.presBloodFlowRate:200 - >>> >>> Unknown parameter/object name presBloodFlowRate - >>> >>> Value Line Error - Command Ignored - >>> (E) @LINE: 1488 TEST.VALUE:ModeTreatment.<>.presDialysateFlowRate:300 - >>> >>> Unknown parameter/object name presDialysateFlowRate - >>> >>> Value Line Error - Command Ignored - >>> (E) @LINE: 1503 TEST.EXPECTED:ModeTreatment.<>.presBloodFlowRate:200 - >>> >>> Unknown parameter/object name presBloodFlowRate - >>> >>> Value Line Error - Command Ignored - >>> (E) @LINE: 1504 TEST.EXPECTED:ModeTreatment.<>.presDialysateFlowRate:300 - >>> >>> Unknown parameter/object name presDialysateFlowRate - >>> >>> Value Line Error - Command Ignored -TEST.END_IMPORT_FAILURES: -TEST.VALUE:ModeTreatment.<>.currentTreatmentState:TREATMENT_DIALYSIS_STATE -TEST.VALUE:ModeTreatment.<>.presTreatmentTimeSecs:3600 -TEST.VALUE:ModeTreatment.<>.presMaxUFVolumeML:600.0 -TEST.VALUE:ModeTreatment.<>.presUFRate:10.0 -TEST.VALUE:ModeTreatment.<>.treatmentTimeMS:1800000 -TEST.VALUE:ModeTreatment.<>.pendingParamChangesTimer:0 -TEST.VALUE:ModeTreatment.<>.pendingUFVolumeChange:0.0 -TEST.VALUE:ModeTreatment.<>.pendingUFRateChange:0.0 -TEST.VALUE:ModeTreatment.<>.pendingTreatmentTimeChange:0 -TEST.VALUE:ModeTreatment.verifyUFSettingsChange.uFVolume:500.0 -TEST.VALUE:uut_prototype_stubs.getDialysisState.return:DIALYSIS_UF_STATE -TEST.VALUE:uut_prototype_stubs.getUltrafiltrationState.return:UF_PAUSED_STATE -TEST.VALUE:uut_prototype_stubs.getCurrentOperationMode.return:MODE_TREA -TEST.VALUE:uut_prototype_stubs.sendChangeUFSettingsResponse.return:1 -TEST.VALUE:uut_prototype_stubs.getMSTimerCount.return:3000000 -TEST.EXPECTED:ModeTreatment.<>.currentTreatmentState:TREATMENT_DIALYSIS_STATE -TEST.EXPECTED:ModeTreatment.<>.presTreatmentTimeSecs:3600 -TEST.EXPECTED:ModeTreatment.<>.presMaxUFVolumeML:600.0 -TEST.EXPECTED:ModeTreatment.<>.presUFRate:10.0 -TEST.EXPECTED:ModeTreatment.<>.treatmentTimeMS:1800000 -TEST.EXPECTED:ModeTreatment.<>.pendingParamChangesTimer:3000000 -TEST.EXPECTED:ModeTreatment.<>.pendingUFVolumeChange:500.0 -TEST.EXPECTED:ModeTreatment.<>.pendingUFRateChange:16.6667 -TEST.EXPECTED:ModeTreatment.<>.pendingTreatmentTimeChange:0 -TEST.EXPECTED:ModeTreatment.verifyUFSettingsChange.uFVolume:500.0 -TEST.EXPECTED:ModeTreatment.verifyUFSettingsChange.return:1 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsResponse.accepted:1 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsResponse.reason:0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsResponse.volume_mL:500.0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsResponse.time_min:0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsResponse.ufRate_mL_min:16.6667 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsResponse.timeDiff:0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsResponse.rateDiff:6.66667 -TEST.END - --- Test Case: verifyUFSettingsChange_Accepted_NoTime2 -TEST.UNIT:ModeTreatment -TEST.SUBPROGRAM:verifyUFSettingsChange -TEST.NEW -TEST.NAME:verifyUFSettingsChange_Accepted_NoTime2 -TEST.IMPORT_FAILURES: -(E) Errors from previous script import(s) - >>> (E) @LINE: 1528 TEST.VALUE:ModeTreatment.<>.presBloodFlowRate:200 - >>> >>> Unknown parameter/object name presBloodFlowRate - >>> >>> Value Line Error - Command Ignored - >>> (E) @LINE: 1529 TEST.VALUE:ModeTreatment.<>.presDialysateFlowRate:600 - >>> >>> Unknown parameter/object name presDialysateFlowRate - >>> >>> Value Line Error - Command Ignored - >>> (E) @LINE: 1544 TEST.EXPECTED:ModeTreatment.<>.presBloodFlowRate:200 - >>> >>> Unknown parameter/object name presBloodFlowRate - >>> >>> Value Line Error - Command Ignored - >>> (E) @LINE: 1545 TEST.EXPECTED:ModeTreatment.<>.presDialysateFlowRate:600 - >>> >>> Unknown parameter/object name presDialysateFlowRate - >>> >>> Value Line Error - Command Ignored -TEST.END_IMPORT_FAILURES: -TEST.VALUE:ModeTreatment.<>.currentTreatmentState:TREATMENT_DIALYSIS_STATE -TEST.VALUE:ModeTreatment.<>.presTreatmentTimeSecs:14400 -TEST.VALUE:ModeTreatment.<>.presMaxUFVolumeML:2400.0 -TEST.VALUE:ModeTreatment.<>.presUFRate:10.0 -TEST.VALUE:ModeTreatment.<>.treatmentTimeMS:10000000 -TEST.VALUE:ModeTreatment.<>.pendingParamChangesTimer:0 -TEST.VALUE:ModeTreatment.<>.pendingUFVolumeChange:0.0 -TEST.VALUE:ModeTreatment.<>.pendingUFRateChange:0.0 -TEST.VALUE:ModeTreatment.<>.pendingTreatmentTimeChange:0 -TEST.VALUE:ModeTreatment.verifyUFSettingsChange.uFVolume:3600.0 -TEST.VALUE:uut_prototype_stubs.getDialysisState.return:DIALYSIS_UF_STATE -TEST.VALUE:uut_prototype_stubs.getUltrafiltrationState.return:UF_PAUSED_STATE -TEST.VALUE:uut_prototype_stubs.getCurrentOperationMode.return:MODE_TREA -TEST.VALUE:uut_prototype_stubs.sendChangeUFSettingsResponse.return:1 -TEST.VALUE:uut_prototype_stubs.getMSTimerCount.return:3000000 -TEST.EXPECTED:ModeTreatment.<>.currentTreatmentState:TREATMENT_DIALYSIS_STATE -TEST.EXPECTED:ModeTreatment.<>.presTreatmentTimeSecs:14400 -TEST.EXPECTED:ModeTreatment.<>.presMaxUFVolumeML:2400.0 -TEST.EXPECTED:ModeTreatment.<>.presUFRate:10.0 -TEST.EXPECTED:ModeTreatment.<>.treatmentTimeMS:10000000 -TEST.EXPECTED:ModeTreatment.<>.pendingParamChangesTimer:3000000 -TEST.EXPECTED:ModeTreatment.<>.pendingUFVolumeChange:2400.0 -TEST.EXPECTED:ModeTreatment.<>.pendingUFRateChange:0.0 -TEST.EXPECTED:ModeTreatment.<>.pendingTreatmentTimeChange:0 -TEST.EXPECTED:ModeTreatment.verifyUFSettingsChange.uFVolume:3600.0 -TEST.EXPECTED:ModeTreatment.verifyUFSettingsChange.return:0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsResponse.accepted:0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsResponse.reason:12 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsResponse.volume_mL:2400.0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsResponse.ufRate_mL_min:0.0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsResponse.rateDiff:0.0 -TEST.END - --- Test Case: verifyUFSettingsChange_Accepted_Stopped -TEST.UNIT:ModeTreatment -TEST.SUBPROGRAM:verifyUFSettingsChange -TEST.NEW -TEST.NAME:verifyUFSettingsChange_Accepted_Stopped -TEST.IMPORT_FAILURES: -(E) Errors from previous script import(s) - >>> (E) @LINE: 1569 TEST.VALUE:ModeTreatment.<>.presBloodFlowRate:200 - >>> >>> Unknown parameter/object name presBloodFlowRate - >>> >>> Value Line Error - Command Ignored - >>> (E) @LINE: 1570 TEST.VALUE:ModeTreatment.<>.presDialysateFlowRate:300 - >>> >>> Unknown parameter/object name presDialysateFlowRate - >>> >>> Value Line Error - Command Ignored - >>> (E) @LINE: 1585 TEST.EXPECTED:ModeTreatment.<>.presBloodFlowRate:200 - >>> >>> Unknown parameter/object name presBloodFlowRate - >>> >>> Value Line Error - Command Ignored - >>> (E) @LINE: 1586 TEST.EXPECTED:ModeTreatment.<>.presDialysateFlowRate:300 - >>> >>> Unknown parameter/object name presDialysateFlowRate - >>> >>> Value Line Error - Command Ignored -TEST.END_IMPORT_FAILURES: -TEST.VALUE:ModeTreatment.<>.currentTreatmentState:TREATMENT_STOP_STATE -TEST.VALUE:ModeTreatment.<>.presTreatmentTimeSecs:3600 -TEST.VALUE:ModeTreatment.<>.presMaxUFVolumeML:600.0 -TEST.VALUE:ModeTreatment.<>.presUFRate:10.0 -TEST.VALUE:ModeTreatment.<>.treatmentTimeMS:2900000 -TEST.VALUE:ModeTreatment.<>.pendingParamChangesTimer:0 -TEST.VALUE:ModeTreatment.<>.pendingUFVolumeChange:0.0 -TEST.VALUE:ModeTreatment.<>.pendingUFRateChange:0.0 -TEST.VALUE:ModeTreatment.<>.pendingTreatmentTimeChange:0 -TEST.VALUE:ModeTreatment.verifyUFSettingsChange.uFVolume:1200.0 -TEST.VALUE:uut_prototype_stubs.getDialysisState.return:DIALYSIS_UF_STATE -TEST.VALUE:uut_prototype_stubs.getUltrafiltrationState.return:UF_PAUSED_STATE -TEST.VALUE:uut_prototype_stubs.getCurrentOperationMode.return:MODE_TREA -TEST.VALUE:uut_prototype_stubs.sendChangeUFSettingsResponse.return:1 -TEST.VALUE:uut_prototype_stubs.getMSTimerCount.return:3000000 -TEST.EXPECTED:ModeTreatment.<>.currentTreatmentState:TREATMENT_STOP_STATE -TEST.EXPECTED:ModeTreatment.<>.presTreatmentTimeSecs:3600 -TEST.EXPECTED:ModeTreatment.<>.presMaxUFVolumeML:600.0 -TEST.EXPECTED:ModeTreatment.<>.presUFRate:10.0 -TEST.EXPECTED:ModeTreatment.<>.treatmentTimeMS:2900000 -TEST.EXPECTED:ModeTreatment.<>.pendingParamChangesTimer:3000000 -TEST.EXPECTED:ModeTreatment.<>.pendingUFVolumeChange:600.0 -TEST.EXPECTED:ModeTreatment.<>.pendingUFRateChange:0.0 -TEST.EXPECTED:ModeTreatment.<>.pendingTreatmentTimeChange:0 -TEST.EXPECTED:ModeTreatment.verifyUFSettingsChange.uFVolume:1200.0 -TEST.EXPECTED:ModeTreatment.verifyUFSettingsChange.return:0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsResponse.accepted:0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsResponse.reason:12 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsResponse.volume_mL:600.0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsResponse.time_min:0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsResponse.ufRate_mL_min:0.0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsResponse.timeDiff:0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsResponse.rateDiff:0.0 -TEST.END - -- Test Case: verifyUFSettingsChange_InvalidTreatmentState1 TEST.UNIT:ModeTreatment TEST.SUBPROGRAM:verifyUFSettingsChange @@ -4792,167 +3792,6 @@ TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsResponse.rateDiff:0.0 TEST.END --- Test Case: verifyUFSettingsChange_TreatmentAlmostDone -TEST.UNIT:ModeTreatment -TEST.SUBPROGRAM:verifyUFSettingsChange -TEST.NEW -TEST.NAME:verifyUFSettingsChange_TreatmentAlmostDone -TEST.IMPORT_FAILURES: -(E) Errors from previous script import(s) - >>> (E) @LINE: 1733 TEST.VALUE:ModeTreatment.<>.presBloodFlowRate:200 - >>> >>> Unknown parameter/object name presBloodFlowRate - >>> >>> Value Line Error - Command Ignored - >>> (E) @LINE: 1734 TEST.VALUE:ModeTreatment.<>.presDialysateFlowRate:300 - >>> >>> Unknown parameter/object name presDialysateFlowRate - >>> >>> Value Line Error - Command Ignored - >>> (E) @LINE: 1749 TEST.EXPECTED:ModeTreatment.<>.presBloodFlowRate:200 - >>> >>> Unknown parameter/object name presBloodFlowRate - >>> >>> Value Line Error - Command Ignored - >>> (E) @LINE: 1750 TEST.EXPECTED:ModeTreatment.<>.presDialysateFlowRate:300 - >>> >>> Unknown parameter/object name presDialysateFlowRate - >>> >>> Value Line Error - Command Ignored -TEST.END_IMPORT_FAILURES: -TEST.VALUE:ModeTreatment.<>.currentTreatmentState:TREATMENT_DIALYSIS_STATE -TEST.VALUE:ModeTreatment.<>.presTreatmentTimeSecs:3600 -TEST.VALUE:ModeTreatment.<>.presMaxUFVolumeML:600.0 -TEST.VALUE:ModeTreatment.<>.presUFRate:10.0 -TEST.VALUE:ModeTreatment.<>.treatmentTimeMS:3010000 -TEST.VALUE:ModeTreatment.<>.pendingParamChangesTimer:0 -TEST.VALUE:ModeTreatment.<>.pendingUFVolumeChange:0.0 -TEST.VALUE:ModeTreatment.<>.pendingUFRateChange:0.0 -TEST.VALUE:ModeTreatment.<>.pendingTreatmentTimeChange:0 -TEST.VALUE:ModeTreatment.verifyUFSettingsChange.uFVolume:1200.0 -TEST.VALUE:uut_prototype_stubs.getDialysisState.return:DIALYSIS_UF_STATE -TEST.VALUE:uut_prototype_stubs.getUltrafiltrationState.return:UF_RUNNING_STATE -TEST.VALUE:uut_prototype_stubs.getCurrentOperationMode.return:MODE_TREA -TEST.VALUE:uut_prototype_stubs.sendChangeUFSettingsResponse.return:1 -TEST.VALUE:uut_prototype_stubs.getMSTimerCount.return:3000000 -TEST.EXPECTED:ModeTreatment.<>.presTreatmentTimeSecs:3600 -TEST.EXPECTED:ModeTreatment.<>.presMaxUFVolumeML:600.0 -TEST.EXPECTED:ModeTreatment.<>.presUFRate:10.0 -TEST.EXPECTED:ModeTreatment.<>.pendingParamChangesTimer:0 -TEST.EXPECTED:ModeTreatment.<>.pendingUFVolumeChange:600.0 -TEST.EXPECTED:ModeTreatment.<>.pendingUFRateChange:10.0 -TEST.EXPECTED:ModeTreatment.<>.pendingTreatmentTimeChange:60 -TEST.EXPECTED:ModeTreatment.verifyUFSettingsChange.return:0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsResponse.accepted:0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsResponse.reason:5 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsResponse.volume_mL:600.0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsResponse.time_min:60 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsResponse.ufRate_mL_min:10.0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsResponse.timeDiff:0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsResponse.rateDiff:0.0 -TEST.END - --- Test Case: verifyUFSettingsChange_UFRateAndTrtTimeOutOfRange -TEST.UNIT:ModeTreatment -TEST.SUBPROGRAM:verifyUFSettingsChange -TEST.NEW -TEST.NAME:verifyUFSettingsChange_UFRateAndTrtTimeOutOfRange -TEST.IMPORT_FAILURES: -(E) Errors from previous script import(s) - >>> (E) @LINE: 1774 TEST.VALUE:ModeTreatment.<>.presBloodFlowRate:200 - >>> >>> Unknown parameter/object name presBloodFlowRate - >>> >>> Value Line Error - Command Ignored - >>> (E) @LINE: 1775 TEST.VALUE:ModeTreatment.<>.presDialysateFlowRate:300 - >>> >>> Unknown parameter/object name presDialysateFlowRate - >>> >>> Value Line Error - Command Ignored - >>> (E) @LINE: 1791 TEST.EXPECTED:ModeTreatment.<>.presBloodFlowRate:200 - >>> >>> Unknown parameter/object name presBloodFlowRate - >>> >>> Value Line Error - Command Ignored - >>> (E) @LINE: 1792 TEST.EXPECTED:ModeTreatment.<>.presDialysateFlowRate:300 - >>> >>> Unknown parameter/object name presDialysateFlowRate - >>> >>> Value Line Error - Command Ignored -TEST.END_IMPORT_FAILURES: -TEST.VALUE:ModeTreatment.<>.currentTreatmentState:TREATMENT_DIALYSIS_STATE -TEST.VALUE:ModeTreatment.<>.presTreatmentTimeSecs:3600 -TEST.VALUE:ModeTreatment.<>.presMaxUFVolumeML:600.0 -TEST.VALUE:ModeTreatment.<>.presUFRate:10.0 -TEST.VALUE:ModeTreatment.<>.treatmentTimeMS:2900000 -TEST.VALUE:ModeTreatment.<>.pendingParamChangesTimer:0 -TEST.VALUE:ModeTreatment.<>.pendingUFVolumeChange:0.0 -TEST.VALUE:ModeTreatment.<>.pendingUFRateChange:0.0 -TEST.VALUE:ModeTreatment.<>.pendingTreatmentTimeChange:0 -TEST.VALUE:ModeTreatment.verifyUFSettingsChange.uFVolume:5000.0 -TEST.VALUE:uut_prototype_stubs.getDialysisState.return:DIALYSIS_UF_STATE -TEST.VALUE:uut_prototype_stubs.getUltrafiltrationState.return:UF_RUNNING_STATE -TEST.VALUE:uut_prototype_stubs.getCurrentOperationMode.return:MODE_TREA -TEST.VALUE:uut_prototype_stubs.sendChangeUFSettingsResponse.return:1 -TEST.VALUE:uut_prototype_stubs.getMSTimerCount.return:3000000 -TEST.EXPECTED:ModeTreatment.<>.currentTreatmentState:TREATMENT_DIALYSIS_STATE -TEST.EXPECTED:ModeTreatment.<>.presTreatmentTimeSecs:3600 -TEST.EXPECTED:ModeTreatment.<>.presMaxUFVolumeML:600.0 -TEST.EXPECTED:ModeTreatment.<>.presUFRate:10.0 -TEST.EXPECTED:ModeTreatment.<>.treatmentTimeMS:2900000 -TEST.EXPECTED:ModeTreatment.<>.pendingParamChangesTimer:3000000 -TEST.EXPECTED:ModeTreatment.<>.pendingUFVolumeChange:600.0 -TEST.EXPECTED:ModeTreatment.<>.pendingUFRateChange:0.0 -TEST.EXPECTED:ModeTreatment.<>.pendingTreatmentTimeChange:0 -TEST.EXPECTED:ModeTreatment.verifyUFSettingsChange.return:0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsResponse.accepted:0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsResponse.reason:12 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsResponse.volume_mL:600.0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsResponse.time_min:0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsResponse.ufRate_mL_min:0.0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsResponse.timeDiff:0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsResponse.rateDiff:0.0 -TEST.END - --- Test Case: verifyUFSettingsChange_UFRateOutOfRange -TEST.UNIT:ModeTreatment -TEST.SUBPROGRAM:verifyUFSettingsChange -TEST.NEW -TEST.NAME:verifyUFSettingsChange_UFRateOutOfRange -TEST.IMPORT_FAILURES: -(E) Errors from previous script import(s) - >>> (E) @LINE: 1817 TEST.VALUE:ModeTreatment.<>.presBloodFlowRate:200 - >>> >>> Unknown parameter/object name presBloodFlowRate - >>> >>> Value Line Error - Command Ignored - >>> (E) @LINE: 1818 TEST.VALUE:ModeTreatment.<>.presDialysateFlowRate:300 - >>> >>> Unknown parameter/object name presDialysateFlowRate - >>> >>> Value Line Error - Command Ignored - >>> (E) @LINE: 1834 TEST.EXPECTED:ModeTreatment.<>.presBloodFlowRate:200 - >>> >>> Unknown parameter/object name presBloodFlowRate - >>> >>> Value Line Error - Command Ignored - >>> (E) @LINE: 1835 TEST.EXPECTED:ModeTreatment.<>.presDialysateFlowRate:300 - >>> >>> Unknown parameter/object name presDialysateFlowRate - >>> >>> Value Line Error - Command Ignored -TEST.END_IMPORT_FAILURES: -TEST.VALUE:ModeTreatment.<>.currentTreatmentState:TREATMENT_DIALYSIS_STATE -TEST.VALUE:ModeTreatment.<>.presTreatmentTimeSecs:3600 -TEST.VALUE:ModeTreatment.<>.presMaxUFVolumeML:600.0 -TEST.VALUE:ModeTreatment.<>.presUFRate:10.0 -TEST.VALUE:ModeTreatment.<>.treatmentTimeMS:2900000 -TEST.VALUE:ModeTreatment.<>.pendingParamChangesTimer:0 -TEST.VALUE:ModeTreatment.<>.pendingUFVolumeChange:0.0 -TEST.VALUE:ModeTreatment.<>.pendingUFRateChange:0.0 -TEST.VALUE:ModeTreatment.<>.pendingTreatmentTimeChange:0 -TEST.VALUE:ModeTreatment.verifyUFSettingsChange.uFVolume:3000.0 -TEST.VALUE:uut_prototype_stubs.getDialysisState.return:DIALYSIS_UF_STATE -TEST.VALUE:uut_prototype_stubs.getUltrafiltrationState.return:UF_RUNNING_STATE -TEST.VALUE:uut_prototype_stubs.getCurrentOperationMode.return:MODE_TREA -TEST.VALUE:uut_prototype_stubs.sendChangeUFSettingsResponse.return:1 -TEST.VALUE:uut_prototype_stubs.getMSTimerCount.return:3000000 -TEST.EXPECTED:ModeTreatment.<>.currentTreatmentState:TREATMENT_DIALYSIS_STATE -TEST.EXPECTED:ModeTreatment.<>.presTreatmentTimeSecs:3600 -TEST.EXPECTED:ModeTreatment.<>.presMaxUFVolumeML:600.0 -TEST.EXPECTED:ModeTreatment.<>.presUFRate:10.0 -TEST.EXPECTED:ModeTreatment.<>.treatmentTimeMS:2900000 -TEST.EXPECTED:ModeTreatment.<>.pendingParamChangesTimer:3000000 -TEST.EXPECTED:ModeTreatment.<>.pendingUFVolumeChange:600.0 -TEST.EXPECTED:ModeTreatment.<>.pendingUFRateChange:0.0 -TEST.EXPECTED:ModeTreatment.<>.pendingTreatmentTimeChange:0 -TEST.EXPECTED:ModeTreatment.verifyUFSettingsChange.uFVolume:3000.0 -TEST.EXPECTED:ModeTreatment.verifyUFSettingsChange.return:0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsResponse.accepted:0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsResponse.reason:12 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsResponse.volume_mL:600.0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsResponse.time_min:0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsResponse.ufRate_mL_min:0.0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsResponse.timeDiff:0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsResponse.rateDiff:0.0 -TEST.END - -- Test Case: verifyUFSettingsChange_UFVolOutOfRange TEST.UNIT:ModeTreatment TEST.SUBPROGRAM:verifyUFSettingsChange @@ -5056,498 +3895,6 @@ -- Subprogram: verifyUFSettingsConfirmation --- Test Case: verifyUFSettingsConfirmation.001 -TEST.UNIT:ModeTreatment -TEST.SUBPROGRAM:verifyUFSettingsConfirmation -TEST.NEW -TEST.NAME:verifyUFSettingsConfirmation.001 -TEST.STUB:ModeTreatment.broadcastTreatmentSettingsRanges -TEST.VALUE:ModeTreatment.<>.currentTreatmentState:TREATMENT_DIALYSIS_STATE -TEST.VALUE:ModeTreatment.<>.presTreatmentTimeSecs:3600 -TEST.VALUE:ModeTreatment.<>.presMaxUFVolumeML:600.0 -TEST.VALUE:ModeTreatment.<>.presUFRate:10.0 -TEST.VALUE:ModeTreatment.<>.pendingParamChangesTimer:1000000 -TEST.VALUE:ModeTreatment.<>.pendingUFVolumeChange:1200.0 -TEST.VALUE:ModeTreatment.<>.pendingUFRateChange:20.0 -TEST.VALUE:ModeTreatment.<>.pendingTreatmentTimeChange:120 -TEST.VALUE:ModeTreatment.verifyUFSettingsConfirmation.uFVolume:1200.0 -TEST.VALUE:ModeTreatment.verifyUFSettingsConfirmation.adjustment:MACRO=UF_ADJ_TREATMENT_TIME -TEST.VALUE:uut_prototype_stubs.getDialysisState.return:DIALYSIS_UF_STATE -TEST.VALUE:uut_prototype_stubs.getUltrafiltrationState.return:UF_RUNNING_STATE -TEST.VALUE:uut_prototype_stubs.getCurrentOperationMode.return:MODE_TREA -TEST.VALUE:uut_prototype_stubs.sendChangeUFSettingsOptionResponse.return:1 -TEST.VALUE:uut_prototype_stubs.getMSTimerCount.return:1059000 -TEST.EXPECTED:ModeTreatment.<>.currentTreatmentState:TREATMENT_DIALYSIS_STATE -TEST.EXPECTED:ModeTreatment.<>.presTreatmentTimeSecs:3600 -TEST.EXPECTED:ModeTreatment.<>.presMaxUFVolumeML:1200.0 -TEST.EXPECTED:ModeTreatment.<>.presUFRate:10.0 -TEST.EXPECTED:ModeTreatment.<>.pendingParamChangesTimer:1000000 -TEST.EXPECTED:ModeTreatment.<>.pendingUFVolumeChange:1200.0 -TEST.EXPECTED:ModeTreatment.<>.pendingUFRateChange:20.0 -TEST.EXPECTED:ModeTreatment.<>.pendingTreatmentTimeChange:120 -TEST.EXPECTED:ModeTreatment.verifyUFSettingsConfirmation.uFVolume:1200.0 -TEST.EXPECTED:ModeTreatment.verifyUFSettingsConfirmation.adjustment:0 -TEST.EXPECTED:ModeTreatment.verifyUFSettingsConfirmation.return:1 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsOptionResponse.accepted:1 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsOptionResponse.reason:0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsOptionResponse.volume_mL:1200.0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsOptionResponse.time_min:60 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsOptionResponse.ufRate_mL_min:10.0 -TEST.END - --- Test Case: verifyUFSettingsConfirmation_Accepted_AdjTime -TEST.UNIT:ModeTreatment -TEST.SUBPROGRAM:verifyUFSettingsConfirmation -TEST.NEW -TEST.NAME:verifyUFSettingsConfirmation_Accepted_AdjTime -TEST.IMPORT_FAILURES: -(I) @LINE: 5185 - >>> 'UF_ADJ_TREATMENT_TIME' was specified as a macro, but it is in the symbol dictionary. -(E) Errors from previous script import(s) - >>> (I) @LINE: 5149 - >>> >>> 'UF_ADJ_TREATMENT_TIME' was specified as a macro, but it is in the symbol dictionary. - >>> (I) @LINE: 5146 - >>> >>> 'UF_ADJ_TREATMENT_TIME' was specified as a macro, but it is in the symbol dictionary. - >>> (I) @LINE: 185 - >>> >>> 'UF_ADJ_TREATMENT_TIME' was specified as a macro, but it is in the symbol dictionary. - >>> (I) @LINE: 5077 - >>> >>> 'UF_ADJ_TREATMENT_TIME' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 5071 ->>> >>> 'UF_ADJ_TREATMENT_TIME' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 5069 ->>> >>> 'UF_ADJ_TREATMENT_TIME' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 5030 ->>> >>> 'UF_ADJ_TREATMENT_TIME' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 5025 ->>> >>> 'UF_ADJ_TREATMENT_TIME' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 4761 ->>> >>> 'UF_ADJ_TREATMENT_TIME' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 4380 ->>> >>> 'UF_ADJ_TREATMENT_TIME' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 4376 ->>> >>> 'UF_ADJ_TREATMENT_TIME' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 4337 ->>> >>> 'UF_ADJ_TREATMENT_TIME' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 4367 ->>> >>> 'UF_ADJ_TREATMENT_TIME' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 4351 ->>> >>> 'UF_ADJ_TREATMENT_TIME' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 4347 ->>> >>> 'UF_ADJ_TREATMENT_TIME' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 4343 ->>> >>> 'UF_ADJ_TREATMENT_TIME' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 4323 ->>> >>> 'UF_ADJ_TREATMENT_TIME' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 4288 ->>> >>> 'UF_ADJ_TREATMENT_TIME' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 4284 ->>> >>> 'UF_ADJ_TREATMENT_TIME' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 4280 ->>> >>> 'UF_ADJ_TREATMENT_TIME' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 4229 ->>> >>> 'UF_ADJ_TREATMENT_TIME' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 4225 ->>> >>> 'UF_ADJ_TREATMENT_TIME' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 4194 ->>> >>> 'UF_ADJ_TREATMENT_TIME' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 4130 ->>> >>> 'UF_ADJ_TREATMENT_TIME' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 4128 ->>> >>> 'UF_ADJ_TREATMENT_TIME' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 3852 ->>> >>> 'UF_ADJ_TREATMENT_TIME' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 3850 ->>> >>> 'UF_ADJ_TREATMENT_TIME' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 5684 ->>> >>> 'UF_ADJ_TREATMENT_TIME' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 6963 ->>> >>> 'UF_ADJ_TREATMENT_TIME' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 6892 ->>> >>> 'UF_ADJ_TREATMENT_TIME' was specified as a macro, but it is in the symbol dictionary. ->>> (E) @LINE: 1904 TEST.VALUE:ModeTreatment.<>.presBloodFlowRate:200 ->>> >>> Unknown parameter/object name presBloodFlowRate ->>> >>> Value Line Error - Command Ignored ->>> (E) @LINE: 1905 TEST.VALUE:ModeTreatment.<>.presDialysateFlowRate:300 ->>> >>> Unknown parameter/object name presDialysateFlowRate ->>> >>> Value Line Error - Command Ignored ->>> (I) @LINE: 1913 ->>> >>> 'UF_ADJ_TREATMENT_TIME' was specified as a macro, but it is in the symbol dictionary. ->>> (E) @LINE: 1920 TEST.EXPECTED:ModeTreatment.<>.presBloodFlowRate:200 ->>> >>> Unknown parameter/object name presBloodFlowRate ->>> >>> Value Line Error - Command Ignored ->>> (E) @LINE: 1921 TEST.EXPECTED:ModeTreatment.<>.presDialysateFlowRate:300 ->>> >>> Unknown parameter/object name presDialysateFlowRate ->>> >>> Value Line Error - Command Ignored -TEST.END_IMPORT_FAILURES: -TEST.STUB:ModeTreatment.broadcastTreatmentSettingsRanges -TEST.VALUE:ModeTreatment.<>.currentTreatmentState:TREATMENT_DIALYSIS_STATE -TEST.VALUE:ModeTreatment.<>.presTreatmentTimeSecs:3600 -TEST.VALUE:ModeTreatment.<>.presMaxUFVolumeML:600.0 -TEST.VALUE:ModeTreatment.<>.presUFRate:10.0 -TEST.VALUE:ModeTreatment.<>.pendingParamChangesTimer:1000000 -TEST.VALUE:ModeTreatment.<>.pendingUFVolumeChange:1200.0 -TEST.VALUE:ModeTreatment.<>.pendingUFRateChange:20.0 -TEST.VALUE:ModeTreatment.<>.pendingTreatmentTimeChange:120 -TEST.VALUE:ModeTreatment.verifyUFSettingsConfirmation.uFVolume:1200.0 -TEST.VALUE:ModeTreatment.verifyUFSettingsConfirmation.adjustment:MACRO=UF_ADJ_TREATMENT_TIME -TEST.VALUE:uut_prototype_stubs.getDialysisState.return:DIALYSIS_UF_STATE -TEST.VALUE:uut_prototype_stubs.getUltrafiltrationState.return:UF_PAUSED_STATE -TEST.VALUE:uut_prototype_stubs.getCurrentOperationMode.return:MODE_TREA -TEST.VALUE:uut_prototype_stubs.sendChangeUFSettingsOptionResponse.return:1 -TEST.VALUE:uut_prototype_stubs.getMSTimerCount.return:1059000 -TEST.EXPECTED:ModeTreatment.<>.currentTreatmentState:TREATMENT_DIALYSIS_STATE -TEST.EXPECTED:ModeTreatment.<>.presTreatmentTimeSecs:3600 -TEST.EXPECTED:ModeTreatment.<>.presMaxUFVolumeML:1200.0 -TEST.EXPECTED:ModeTreatment.<>.presUFRate:10.0 -TEST.EXPECTED:ModeTreatment.<>.pendingParamChangesTimer:1000000 -TEST.EXPECTED:ModeTreatment.<>.pendingUFVolumeChange:1200.0 -TEST.EXPECTED:ModeTreatment.<>.pendingUFRateChange:20.0 -TEST.EXPECTED:ModeTreatment.<>.pendingTreatmentTimeChange:120 -TEST.EXPECTED:ModeTreatment.verifyUFSettingsConfirmation.uFVolume:1200.0 -TEST.EXPECTED:ModeTreatment.verifyUFSettingsConfirmation.adjustment:0 -TEST.EXPECTED:ModeTreatment.verifyUFSettingsConfirmation.return:1 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsOptionResponse.accepted:1 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsOptionResponse.reason:0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsOptionResponse.volume_mL:1200.0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsOptionResponse.time_min:60 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsOptionResponse.ufRate_mL_min:10.0 -TEST.END - --- Test Case: verifyUFSettingsConfirmation_Accepted_AdjUFRate -TEST.UNIT:ModeTreatment -TEST.SUBPROGRAM:verifyUFSettingsConfirmation -TEST.NEW -TEST.NAME:verifyUFSettingsConfirmation_Accepted_AdjUFRate -TEST.IMPORT_FAILURES: -(I) @LINE: 5305 - >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. -(E) Errors from previous script import(s) - >>> (I) @LINE: 5267 - >>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. - >>> (I) @LINE: 5262 - >>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. - >>> (I) @LINE: 298 - >>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. - >>> (I) @LINE: 5189 - >>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 5181 ->>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 5177 ->>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 5136 ->>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 5129 ->>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 4856 ->>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 4473 ->>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 4467 ->>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 4426 ->>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 4454 ->>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 4436 ->>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 4430 ->>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 4424 ->>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 4402 ->>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 4365 ->>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 4359 ->>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 4353 ->>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 4300 ->>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 4294 ->>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 4261 ->>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 4195 ->>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 4191 ->>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 3913 ->>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 3909 ->>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 5741 ->>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 7017 ->>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 6943 ->>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (E) @LINE: 1939 TEST.VALUE:ModeTreatment.<>.presBloodFlowRate:200 ->>> >>> Unknown parameter/object name presBloodFlowRate ->>> >>> Value Line Error - Command Ignored ->>> (E) @LINE: 1940 TEST.VALUE:ModeTreatment.<>.presDialysateFlowRate:300 ->>> >>> Unknown parameter/object name presDialysateFlowRate ->>> >>> Value Line Error - Command Ignored ->>> (I) @LINE: 1948 ->>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (E) @LINE: 1950 TEST.VALUE:uut_prototype_stubs.getUltrafiltrationState.return:UF_OFF_STATE ->>> >>> Value for 'Enumeral' out of range ->>> >>> Read "UF_OFF_STATE" ->>> >>> Command Ignored ->>> (E) @LINE: 1955 TEST.EXPECTED:ModeTreatment.<>.presBloodFlowRate:200 ->>> >>> Unknown parameter/object name presBloodFlowRate ->>> >>> Value Line Error - Command Ignored ->>> (E) @LINE: 1956 TEST.EXPECTED:ModeTreatment.<>.presDialysateFlowRate:300 ->>> >>> Unknown parameter/object name presDialysateFlowRate ->>> >>> Value Line Error - Command Ignored -TEST.END_IMPORT_FAILURES: -TEST.STUB:ModeTreatment.broadcastTreatmentSettingsRanges -TEST.VALUE:ModeTreatment.<>.currentTreatmentState:TREATMENT_DIALYSIS_STATE -TEST.VALUE:ModeTreatment.<>.presTreatmentTimeSecs:3600 -TEST.VALUE:ModeTreatment.<>.presMaxUFVolumeML:600.0 -TEST.VALUE:ModeTreatment.<>.presUFRate:10.0 -TEST.VALUE:ModeTreatment.<>.pendingParamChangesTimer:1000000 -TEST.VALUE:ModeTreatment.<>.pendingUFVolumeChange:1200.0 -TEST.VALUE:ModeTreatment.<>.pendingUFRateChange:20.0 -TEST.VALUE:ModeTreatment.<>.pendingTreatmentTimeChange:120 -TEST.VALUE:ModeTreatment.verifyUFSettingsConfirmation.uFVolume:1200.0 -TEST.VALUE:ModeTreatment.verifyUFSettingsConfirmation.adjustment:MACRO=UF_ADJ_UF_RATE -TEST.VALUE:uut_prototype_stubs.getDialysisState.return:DIALYSIS_UF_STATE -TEST.VALUE:uut_prototype_stubs.getCurrentOperationMode.return:MODE_TREA -TEST.VALUE:uut_prototype_stubs.sendChangeUFSettingsOptionResponse.return:1 -TEST.VALUE:uut_prototype_stubs.getMSTimerCount.return:1059000 -TEST.EXPECTED:ModeTreatment.<>.presTreatmentTimeSecs:3600 -TEST.EXPECTED:ModeTreatment.<>.presMaxUFVolumeML:1200.0 -TEST.EXPECTED:ModeTreatment.<>.presUFRate:20.0 -TEST.EXPECTED:ModeTreatment.verifyUFSettingsConfirmation.return:1 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsOptionResponse.accepted:1 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsOptionResponse.reason:0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsOptionResponse.volume_mL:1200.0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsOptionResponse.time_min:60 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsOptionResponse.ufRate_mL_min:20.0 -TEST.END - --- Test Case: verifyUFSettingsConfirmation_Accepted_NoResume_Bolus -TEST.UNIT:ModeTreatment -TEST.SUBPROGRAM:verifyUFSettingsConfirmation -TEST.NEW -TEST.NAME:verifyUFSettingsConfirmation_Accepted_NoResume_Bolus -TEST.IMPORT_FAILURES: -(I) @LINE: 5413 - >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. -(E) Errors from previous script import(s) - >>> (I) @LINE: 5373 - >>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. - >>> (I) @LINE: 5366 - >>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. - >>> (I) @LINE: 399 - >>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. - >>> (I) @LINE: 5289 - >>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 5279 ->>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 5273 ->>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 5230 ->>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 5221 ->>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 4946 ->>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 4561 ->>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 4553 ->>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 4510 ->>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 4536 ->>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 4516 ->>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 4508 ->>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 4500 ->>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 4476 ->>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 4437 ->>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 4429 ->>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 4421 ->>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 4366 ->>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 4358 ->>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 4323 ->>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 4255 ->>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 4249 ->>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 3969 ->>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 3963 ->>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 5793 ->>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 7066 ->>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 6989 ->>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (E) @LINE: 1974 TEST.VALUE:ModeTreatment.<>.presBloodFlowRate:200 ->>> >>> Unknown parameter/object name presBloodFlowRate ->>> >>> Value Line Error - Command Ignored ->>> (E) @LINE: 1975 TEST.VALUE:ModeTreatment.<>.presDialysateFlowRate:300 ->>> >>> Unknown parameter/object name presDialysateFlowRate ->>> >>> Value Line Error - Command Ignored ->>> (I) @LINE: 1983 ->>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (E) @LINE: 1989 TEST.EXPECTED:ModeTreatment.<>.presBloodFlowRate:200 ->>> >>> Unknown parameter/object name presBloodFlowRate ->>> >>> Value Line Error - Command Ignored ->>> (E) @LINE: 1990 TEST.EXPECTED:ModeTreatment.<>.presDialysateFlowRate:300 ->>> >>> Unknown parameter/object name presDialysateFlowRate ->>> >>> Value Line Error - Command Ignored -TEST.END_IMPORT_FAILURES: -TEST.STUB:ModeTreatment.broadcastTreatmentSettingsRanges -TEST.VALUE:ModeTreatment.<>.currentTreatmentState:TREATMENT_DIALYSIS_STATE -TEST.VALUE:ModeTreatment.<>.presTreatmentTimeSecs:3600 -TEST.VALUE:ModeTreatment.<>.presMaxUFVolumeML:600.0 -TEST.VALUE:ModeTreatment.<>.presUFRate:10.0 -TEST.VALUE:ModeTreatment.<>.pendingParamChangesTimer:1000000 -TEST.VALUE:ModeTreatment.<>.pendingUFVolumeChange:1200.0 -TEST.VALUE:ModeTreatment.<>.pendingUFRateChange:20.0 -TEST.VALUE:ModeTreatment.<>.pendingTreatmentTimeChange:120 -TEST.VALUE:ModeTreatment.verifyUFSettingsConfirmation.uFVolume:1200.0 -TEST.VALUE:ModeTreatment.verifyUFSettingsConfirmation.adjustment:MACRO=UF_ADJ_UF_RATE -TEST.VALUE:uut_prototype_stubs.getUltrafiltrationState.return:UF_PAUSED_STATE -TEST.VALUE:uut_prototype_stubs.getCurrentOperationMode.return:MODE_TREA -TEST.VALUE:uut_prototype_stubs.sendChangeUFSettingsOptionResponse.return:1 -TEST.VALUE:uut_prototype_stubs.getMSTimerCount.return:1059000 -TEST.EXPECTED:ModeTreatment.<>.presTreatmentTimeSecs:3600 -TEST.EXPECTED:ModeTreatment.<>.presMaxUFVolumeML:1200.0 -TEST.EXPECTED:ModeTreatment.<>.presUFRate:20.0 -TEST.EXPECTED:ModeTreatment.verifyUFSettingsConfirmation.return:1 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsOptionResponse.accepted:1 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsOptionResponse.reason:0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsOptionResponse.volume_mL:1200.0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsOptionResponse.time_min:60 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsOptionResponse.ufRate_mL_min:20.0 -TEST.END - --- Test Case: verifyUFSettingsConfirmation_Accepted_NoResume_Stopped -TEST.UNIT:ModeTreatment -TEST.SUBPROGRAM:verifyUFSettingsConfirmation -TEST.NEW -TEST.NAME:verifyUFSettingsConfirmation_Accepted_NoResume_Stopped -TEST.IMPORT_FAILURES: -(I) @LINE: 5521 - >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. -(E) Errors from previous script import(s) - >>> (I) @LINE: 5479 - >>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. - >>> (I) @LINE: 5470 - >>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. - >>> (I) @LINE: 500 - >>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. - >>> (I) @LINE: 5389 - >>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 5377 ->>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 5369 ->>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 5324 ->>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 5313 ->>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 5036 ->>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 4649 ->>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 4639 ->>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 4594 ->>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 4618 ->>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 4596 ->>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 4586 ->>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 4576 ->>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 4550 ->>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 4509 ->>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 4499 ->>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 4489 ->>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 4432 ->>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 4422 ->>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 4385 ->>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 4315 ->>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 4307 ->>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 4025 ->>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 4017 ->>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 5845 ->>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 7115 ->>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (I) @LINE: 7035 ->>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (E) @LINE: 2008 TEST.VALUE:ModeTreatment.<>.presBloodFlowRate:200 ->>> >>> Unknown parameter/object name presBloodFlowRate ->>> >>> Value Line Error - Command Ignored ->>> (E) @LINE: 2009 TEST.VALUE:ModeTreatment.<>.presDialysateFlowRate:300 ->>> >>> Unknown parameter/object name presDialysateFlowRate ->>> >>> Value Line Error - Command Ignored ->>> (I) @LINE: 2017 ->>> >>> 'UF_ADJ_UF_RATE' was specified as a macro, but it is in the symbol dictionary. ->>> (E) @LINE: 2024 TEST.EXPECTED:ModeTreatment.<>.presBloodFlowRate:200 ->>> >>> Unknown parameter/object name presBloodFlowRate ->>> >>> Value Line Error - Command Ignored ->>> (E) @LINE: 2025 TEST.EXPECTED:ModeTreatment.<>.presDialysateFlowRate:300 ->>> >>> Unknown parameter/object name presDialysateFlowRate ->>> >>> Value Line Error - Command Ignored -TEST.END_IMPORT_FAILURES: -TEST.STUB:ModeTreatment.broadcastTreatmentSettingsRanges -TEST.VALUE:ModeTreatment.<>.currentTreatmentState:TREATMENT_STOP_STATE -TEST.VALUE:ModeTreatment.<>.presTreatmentTimeSecs:3600 -TEST.VALUE:ModeTreatment.<>.presMaxUFVolumeML:600.0 -TEST.VALUE:ModeTreatment.<>.presUFRate:10.0 -TEST.VALUE:ModeTreatment.<>.pendingParamChangesTimer:1000000 -TEST.VALUE:ModeTreatment.<>.pendingUFVolumeChange:1200.0 -TEST.VALUE:ModeTreatment.<>.pendingUFRateChange:20.0 -TEST.VALUE:ModeTreatment.<>.pendingTreatmentTimeChange:120 -TEST.VALUE:ModeTreatment.verifyUFSettingsConfirmation.uFVolume:1200.0 -TEST.VALUE:ModeTreatment.verifyUFSettingsConfirmation.adjustment:MACRO=UF_ADJ_UF_RATE -TEST.VALUE:uut_prototype_stubs.getDialysisState.return:DIALYSIS_UF_STATE -TEST.VALUE:uut_prototype_stubs.getUltrafiltrationState.return:UF_RUNNING_STATE -TEST.VALUE:uut_prototype_stubs.getCurrentOperationMode.return:MODE_TREA -TEST.VALUE:uut_prototype_stubs.sendChangeUFSettingsOptionResponse.return:1 -TEST.VALUE:uut_prototype_stubs.getMSTimerCount.return:1059000 -TEST.EXPECTED:ModeTreatment.<>.presTreatmentTimeSecs:3600 -TEST.EXPECTED:ModeTreatment.<>.presMaxUFVolumeML:1200.0 -TEST.EXPECTED:ModeTreatment.<>.presUFRate:20.0 -TEST.EXPECTED:ModeTreatment.verifyUFSettingsConfirmation.return:1 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsOptionResponse.accepted:1 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsOptionResponse.reason:0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsOptionResponse.volume_mL:1200.0 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsOptionResponse.time_min:60 -TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsOptionResponse.ufRate_mL_min:20.0 -TEST.END - -- Test Case: verifyUFSettingsConfirmation_ConfirmationTimeout TEST.UNIT:ModeTreatment TEST.SUBPROGRAM:verifyUFSettingsConfirmation @@ -5767,3 +4114,139 @@ TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsOptionResponse.time_min:60 TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsOptionResponse.ufRate_mL_min:10.0 TEST.END + +-- Test Case: verifyUFSettingsConfirmation_adjust_not_tx_dialysis_state_UF_state +TEST.UNIT:ModeTreatment +TEST.SUBPROGRAM:verifyUFSettingsConfirmation +TEST.NEW +TEST.NAME:verifyUFSettingsConfirmation_adjust_not_tx_dialysis_state_UF_state +TEST.STUB:ModeTreatment.broadcastTreatmentSettingsRanges +TEST.VALUE:ModeTreatment.<>.currentTreatmentState:TREATMENT_DIALYSIS_STATE +TEST.VALUE:ModeTreatment.<>.presTreatmentTimeSecs:3600 +TEST.VALUE:ModeTreatment.<>.presMaxUFVolumeML:600.0 +TEST.VALUE:ModeTreatment.<>.presUFRate:10.0 +TEST.VALUE:ModeTreatment.<>.pendingParamChangesTimer:1000000 +TEST.VALUE:ModeTreatment.<>.pendingUFVolumeChange:1200.0 +TEST.VALUE:ModeTreatment.<>.pendingUFRateChange:20.0 +TEST.VALUE:ModeTreatment.<>.pendingTreatmentTimeChange:120 +TEST.VALUE:ModeTreatment.verifyUFSettingsConfirmation.uFVolume:1200.0 +TEST.VALUE:ModeTreatment.verifyUFSettingsConfirmation.adjustment:MACRO=UF_ADJ_UF_RATE +TEST.VALUE:uut_prototype_stubs.getDialysisState.return:DIALYSIS_SALINE_BOLUS_STATE +TEST.VALUE:uut_prototype_stubs.getUltrafiltrationState.return:UF_PAUSED_STATE +TEST.VALUE:uut_prototype_stubs.getCurrentOperationMode.return:MODE_TREA +TEST.VALUE:uut_prototype_stubs.sendChangeUFSettingsOptionResponse.return:1 +TEST.VALUE:uut_prototype_stubs.getMSTimerCount.return:1061000 +TEST.VALUE:uut_prototype_stubs.didTimeout.return:0 +TEST.EXPECTED:ModeTreatment.<>.presTreatmentTimeSecs:3600 +TEST.EXPECTED:ModeTreatment.<>.presMaxUFVolumeML:1200.0 +TEST.EXPECTED:ModeTreatment.<>.presUFRate:20.0 +TEST.EXPECTED:ModeTreatment.verifyUFSettingsConfirmation.return:0 +TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsOptionResponse.accepted:0 +TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsOptionResponse.reason:0 +TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsOptionResponse.volume_mL:1200.0 +TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsOptionResponse.time_min:60 +TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsOptionResponse.ufRate_mL_min:20.0 +TEST.EXPECTED:uut_prototype_stubs.didTimeout.timeoutPeriod:60000 +TEST.END + +-- Test Case: verifyUFSettingsConfirmation_adjust_tx_dialysis_not_UF_paused_state +TEST.UNIT:ModeTreatment +TEST.SUBPROGRAM:verifyUFSettingsConfirmation +TEST.NEW +TEST.NAME:verifyUFSettingsConfirmation_adjust_tx_dialysis_not_UF_paused_state +TEST.STUB:ModeTreatment.broadcastTreatmentSettingsRanges +TEST.VALUE:ModeTreatment.<>.currentTreatmentState:TREATMENT_DIALYSIS_STATE +TEST.VALUE:ModeTreatment.<>.presTreatmentTimeSecs:3600 +TEST.VALUE:ModeTreatment.<>.presMaxUFVolumeML:600.0 +TEST.VALUE:ModeTreatment.<>.presUFRate:10.0 +TEST.VALUE:ModeTreatment.<>.pendingParamChangesTimer:1000000 +TEST.VALUE:ModeTreatment.<>.pendingUFVolumeChange:1200.0 +TEST.VALUE:ModeTreatment.<>.pendingUFRateChange:20.0 +TEST.VALUE:ModeTreatment.<>.pendingTreatmentTimeChange:120 +TEST.VALUE:ModeTreatment.verifyUFSettingsConfirmation.uFVolume:1200.0 +TEST.VALUE:ModeTreatment.verifyUFSettingsConfirmation.adjustment:MACRO=UF_ADJ_UF_RATE +TEST.VALUE:uut_prototype_stubs.getDialysisState.return:DIALYSIS_UF_STATE +TEST.VALUE:uut_prototype_stubs.getUltrafiltrationState.return:UF_RUNNING_STATE +TEST.VALUE:uut_prototype_stubs.getCurrentOperationMode.return:MODE_TREA +TEST.VALUE:uut_prototype_stubs.sendChangeUFSettingsOptionResponse.return:1 +TEST.VALUE:uut_prototype_stubs.getMSTimerCount.return:1061000 +TEST.VALUE:uut_prototype_stubs.didTimeout.return:0 +TEST.EXPECTED:ModeTreatment.<>.presTreatmentTimeSecs:3600 +TEST.EXPECTED:ModeTreatment.<>.presMaxUFVolumeML:1200.0 +TEST.EXPECTED:ModeTreatment.<>.presUFRate:20.0 +TEST.EXPECTED:ModeTreatment.verifyUFSettingsConfirmation.return:0 +TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsOptionResponse.accepted:0 +TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsOptionResponse.reason:0 +TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsOptionResponse.volume_mL:1200.0 +TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsOptionResponse.time_min:60 +TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsOptionResponse.ufRate_mL_min:20.0 +TEST.EXPECTED:uut_prototype_stubs.didTimeout.timeoutPeriod:60000 +TEST.END + +-- Test Case: verifyUFSettingsConfirmation_adjust_tx_dialysis_state +TEST.UNIT:ModeTreatment +TEST.SUBPROGRAM:verifyUFSettingsConfirmation +TEST.NEW +TEST.NAME:verifyUFSettingsConfirmation_adjust_tx_dialysis_state +TEST.STUB:ModeTreatment.broadcastTreatmentSettingsRanges +TEST.VALUE:ModeTreatment.<>.currentTreatmentState:TREATMENT_DIALYSIS_STATE +TEST.VALUE:ModeTreatment.<>.presTreatmentTimeSecs:3600 +TEST.VALUE:ModeTreatment.<>.presMaxUFVolumeML:600.0 +TEST.VALUE:ModeTreatment.<>.presUFRate:10.0 +TEST.VALUE:ModeTreatment.<>.pendingParamChangesTimer:1000000 +TEST.VALUE:ModeTreatment.<>.pendingUFVolumeChange:1200.0 +TEST.VALUE:ModeTreatment.<>.pendingUFRateChange:20.0 +TEST.VALUE:ModeTreatment.<>.pendingTreatmentTimeChange:120 +TEST.VALUE:ModeTreatment.verifyUFSettingsConfirmation.uFVolume:1200.0 +TEST.VALUE:ModeTreatment.verifyUFSettingsConfirmation.adjustment:MACRO=UF_ADJ_UF_RATE +TEST.VALUE:uut_prototype_stubs.getDialysisState.return:DIALYSIS_UF_STATE +TEST.VALUE:uut_prototype_stubs.getUltrafiltrationState.return:UF_PAUSED_STATE +TEST.VALUE:uut_prototype_stubs.getCurrentOperationMode.return:MODE_TREA +TEST.VALUE:uut_prototype_stubs.sendChangeUFSettingsOptionResponse.return:1 +TEST.VALUE:uut_prototype_stubs.getMSTimerCount.return:1061000 +TEST.VALUE:uut_prototype_stubs.didTimeout.return:0 +TEST.EXPECTED:ModeTreatment.<>.presTreatmentTimeSecs:3600 +TEST.EXPECTED:ModeTreatment.<>.presMaxUFVolumeML:1200.0 +TEST.EXPECTED:ModeTreatment.<>.presUFRate:20.0 +TEST.EXPECTED:ModeTreatment.verifyUFSettingsConfirmation.return:0 +TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsOptionResponse.accepted:0 +TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsOptionResponse.reason:0 +TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsOptionResponse.volume_mL:1200.0 +TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsOptionResponse.time_min:60 +TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsOptionResponse.ufRate_mL_min:20.0 +TEST.EXPECTED:uut_prototype_stubs.didTimeout.timeoutPeriod:60000 +TEST.END + +-- Test Case: verifyUFSettingsConfirmation_adjust_tx_not_dialysis_state +TEST.UNIT:ModeTreatment +TEST.SUBPROGRAM:verifyUFSettingsConfirmation +TEST.NEW +TEST.NAME:verifyUFSettingsConfirmation_adjust_tx_not_dialysis_state +TEST.STUB:ModeTreatment.broadcastTreatmentSettingsRanges +TEST.VALUE:ModeTreatment.<>.currentTreatmentState:TREATMENT_RINSEBACK_STATE +TEST.VALUE:ModeTreatment.<>.presTreatmentTimeSecs:3600 +TEST.VALUE:ModeTreatment.<>.presMaxUFVolumeML:600.0 +TEST.VALUE:ModeTreatment.<>.presUFRate:10.0 +TEST.VALUE:ModeTreatment.<>.pendingParamChangesTimer:1000000 +TEST.VALUE:ModeTreatment.<>.pendingUFVolumeChange:1200.0 +TEST.VALUE:ModeTreatment.<>.pendingUFRateChange:20.0 +TEST.VALUE:ModeTreatment.<>.pendingTreatmentTimeChange:120 +TEST.VALUE:ModeTreatment.verifyUFSettingsConfirmation.uFVolume:1200.0 +TEST.VALUE:ModeTreatment.verifyUFSettingsConfirmation.adjustment:UF_ADJ_TREATMENT_TIME +TEST.VALUE:uut_prototype_stubs.getDialysisState.return:DIALYSIS_UF_STATE +TEST.VALUE:uut_prototype_stubs.getUltrafiltrationState.return:UF_PAUSED_STATE +TEST.VALUE:uut_prototype_stubs.getCurrentOperationMode.return:MODE_TREA +TEST.VALUE:uut_prototype_stubs.sendChangeUFSettingsOptionResponse.return:1 +TEST.VALUE:uut_prototype_stubs.getMSTimerCount.return:1061000 +TEST.VALUE:uut_prototype_stubs.didTimeout.return:0 +TEST.EXPECTED:ModeTreatment.<>.presTreatmentTimeSecs:3600 +TEST.EXPECTED:ModeTreatment.<>.presMaxUFVolumeML:1200.0 +TEST.EXPECTED:ModeTreatment.<>.presUFRate:10.0 +TEST.EXPECTED:ModeTreatment.verifyUFSettingsConfirmation.return:0 +TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsOptionResponse.accepted:0 +TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsOptionResponse.reason:0 +TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsOptionResponse.volume_mL:1200.0 +TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsOptionResponse.time_min:60 +TEST.EXPECTED:uut_prototype_stubs.sendChangeUFSettingsOptionResponse.ufRate_mL_min:10.0 +TEST.EXPECTED:uut_prototype_stubs.didTimeout.timeoutPeriod:60000 +TEST.END