Index: firmware/App/Modes/Prime.c =================================================================== diff -u -rb1a931b5151819b00f9bc7bb7d352805c3e55015 -r249024dec91cfbdb4c864cc38affac4fb4af9b06 --- firmware/App/Modes/Prime.c (.../Prime.c) (revision b1a931b5151819b00f9bc7bb7d352805c3e55015) +++ firmware/App/Modes/Prime.c (.../Prime.c) (revision 249024dec91cfbdb4c864cc38affac4fb4af9b06) @@ -47,8 +47,8 @@ #define BLOOD_PUMP_FAST_FLOW_RATE_CIRC_BLOOD_CIRCUIT_ML_MIN 300 ///< Blood pump fast flow rate during prime recirculate blood circuit state. #define BLOOD_PUMP_FLOW_RATE_SALINE_DIALYZER_ML_MIN 300 ///< Blood pump flow rate during prime the saline dialyzer dialysate state. #define DIALYSATE_PUMP_PRIME_FLOW_RATE_ML_MIN 300 ///< Dialysate pump flow rate during priming fluid path. -#define DIALYSATE_PUMP_FAST_PRIME_FLOW_RATE_ML_MIN 600 ///< Dialysate pump flow rate during priming fluid path. -#define DPO_PUMP_PRIME_FLOW_RATE_ML_MIN 225 ///< Dialysate outlet pump flow rate needed to match PWM with dialysate inlet pump @ 300 mL/min in open loop mode. +#define DIALYSATE_PUMP_FAST_PRIME_FLOW_RATE_ML_MIN 600 ///< Dialysate pump faster flow rate during priming fluid path. +#define DPO_PUMP_PRIME_FLOW_RATE_ML_MIN 240 ///< Dialysate outlet pump flow rate needed to match PWM with dialysate inlet pump @ 300 mL/min in open loop mode. #define DIALYSATE_DIALYZER_TUBE_VOLUME_ML 115 ///< This total tube volume is used to calculate the Dpi & Dpo time out in the dialysate dialyzer state. #define DIALYSATE_DIALYZER_BYPASS_TUBE_VOLUME_ML 75 ///< This volume is used to calculate the DPi pump time out in the dialyzer bypass state.