Index: firmware/App/Controllers/DialInFlow.c =================================================================== diff -u -rd0c3528f9ce49e8f24176de6bd49f4b5088e78b8 -r323df1eba09cd27a00a56e5e38b1fffcbff3b170 --- firmware/App/Controllers/DialInFlow.c (.../DialInFlow.c) (revision d0c3528f9ce49e8f24176de6bd49f4b5088e78b8) +++ firmware/App/Controllers/DialInFlow.c (.../DialInFlow.c) (revision 323df1eba09cd27a00a56e5e38b1fffcbff3b170) @@ -826,14 +826,14 @@ deltaFlow = 0.5 * estimatedFlow; // 50% estimated flow if ( TRUE == isDialInPumpRunning() ) { - if ( ( TRUE == isPersistentAlarmTriggered( ALARM_ID_HD_DIAL_IN_FLOW_CHECK_FAILURE, deltaFlow < fabs( estimatedFlow - filteredDialInFlowMeterReading ) ) ) ) - { - SET_ALARM_WITH_2_F32_DATA( ALARM_ID_HD_DIAL_IN_FLOW_CHECK_FAILURE, filteredDialInFlowMeterReading, estimatedFlow ); - } + if ( ( TRUE == isPersistentAlarmTriggered( ALARM_ID_HD_DIAL_IN_FLOW_CHECK_FAILURE, deltaFlow < fabs( estimatedFlow - filteredDialInFlowMeterReading ) ) ) ) // DN-31OCT2022 + { + SET_ALARM_WITH_2_F32_DATA( ALARM_ID_HD_DIAL_IN_FLOW_CHECK_FAILURE, filteredDialInFlowMeterReading, estimatedFlow ); + } } else { - isPersistentAlarmTriggered( ALARM_ID_HD_DIAL_IN_FLOW_CHECK_FAILURE, FALSE ); + isPersistentAlarmTriggered( ALARM_ID_HD_DIAL_IN_FLOW_CHECK_FAILURE, FALSE ); } return estimatedFlow;