Index: firmware/App/Controllers/DialInFlow.c =================================================================== diff -u -r85a94d154c634f0bb7b85a3a7312a77fdb9708df -r39b5287d113697ddf9b1624a42aa1f30dabe11a4 --- firmware/App/Controllers/DialInFlow.c (.../DialInFlow.c) (revision 85a94d154c634f0bb7b85a3a7312a77fdb9708df) +++ firmware/App/Controllers/DialInFlow.c (.../DialInFlow.c) (revision 39b5287d113697ddf9b1624a42aa1f30dabe11a4) @@ -821,18 +821,16 @@ //{ // SET_ALARM_WITH_2_U32_DATA( ALARM_ID_HD_DIAL_IN_FLOW_CHECK_FAILURE, filteredDialInFlowMeterReading, estimatedFlow ); //} - - // Check measured rotor speed vs. measured motor speed while controlling to target if ( TRUE == isDialInPumpRunning() ) { - if ( ( TRUE == isPersistentAlarmTriggered( ALARM_ID_HD_DIAL_IN_FLOW_CHECK_FAILURE, deltaFlow < fabs( estimatedFlow - filteredDialInFlowMeterReading ) ) ) ) - { - SET_ALARM_WITH_2_F32_DATA( ALARM_ID_HD_DIAL_IN_FLOW_CHECK_FAILURE, filteredDialInFlowMeterReading, estimatedFlow ); - } + if ( ( TRUE == isPersistentAlarmTriggered( ALARM_ID_HD_DIAL_IN_FLOW_CHECK_FAILURE, deltaFlow < fabs( estimatedFlow - filteredDialInFlowMeterReading ) ) ) ) // DN-31OCT2022 + { + SET_ALARM_WITH_2_F32_DATA( ALARM_ID_HD_DIAL_IN_FLOW_CHECK_FAILURE, filteredDialInFlowMeterReading, estimatedFlow ); + } } else { - isPersistentAlarmTriggered( ALARM_ID_HD_DIAL_IN_FLOW_CHECK_FAILURE, FALSE ); + isPersistentAlarmTriggered( ALARM_ID_HD_DIAL_IN_FLOW_CHECK_FAILURE, FALSE ); } return estimatedFlow;