Index: firmware/App/Modes/Prime.c =================================================================== diff -u -ree505bda9bc7191211b1aa99a5e0c79e99834d64 -r4db65dc16d162be7336b271b8cecf3dea5892b3d --- firmware/App/Modes/Prime.c (.../Prime.c) (revision ee505bda9bc7191211b1aa99a5e0c79e99834d64) +++ firmware/App/Modes/Prime.c (.../Prime.c) (revision 4db65dc16d162be7336b271b8cecf3dea5892b3d) @@ -793,7 +793,6 @@ primeSalineDialyzerTimeLimit = (U32)( ( ( dialyzerDialysateVolume * DIALYZER_VOLUME_SCALE_FACTOR + DIALYZER_DVI_PATH_VOLUME_ML ) * SEC_PER_MIN * MS_PER_SECOND ) / BLOOD_PUMP_FLOW_RATE_SALINE_DIALYZER_ML_MIN ); primeSalineDialyzerStartTime = getMSTimerCount(); -// primeSalineDialyzerBubbleClearStartTime = getMSTimerCount(); if ( TRUE == doesAlarmStatusIndicateStop() ) { @@ -814,15 +813,11 @@ *************************************************************************/ static HD_PRE_TREATMENT_PRIME_STATE_T handlePrimeSalineDialyzerState( void ) { -// static BOOL bubbleClearActive = FALSE; - HD_PRE_TREATMENT_PRIME_STATE_T state = HD_PRIME_SALINE_DIALYZER_STATE; if ( TRUE == didTimeout( primeSalineDialyzerStartTime, primeSalineDialyzerTimeLimit ) ) { state = HD_PRIME_RESERVOIR_TWO_FILL_COMPLETE_STATE; - // Release pressure - setValvePosition( VDI, VALVE_POSITION_A_INSERT_EJECT ); } // TODO: Rework alarm and timeout logic here. PRIME_SALINE_DIALYZER_TIME_OUT_COUNT may not be reached. if ( TRUE == didTimeout( primeSalineDialyzerStartTime, PRIME_SALINE_DIALYZER_TIME_OUT_COUNT ) )