Index: firmware/App/Modes/Dialysis.c =================================================================== diff -u -r81e7649d3fe44642515332103d0d24de28154331 -r4fdd94336f97f71d0ba2918260125ecb46d52e75 --- firmware/App/Modes/Dialysis.c (.../Dialysis.c) (revision 81e7649d3fe44642515332103d0d24de28154331) +++ firmware/App/Modes/Dialysis.c (.../Dialysis.c) (revision 4fdd94336f97f71d0ba2918260125ecb46d52e75) @@ -1305,13 +1305,10 @@ } } - // TODO Set proper limits and use separate Alarm Number // TODO How does this alarm get cleared? if ( driftDiff > LOAD_CELL_PRIMARY_BACKUP_MAX_ALLOWED_DRIFT_GRAMS ) { - //checkPersistentAlarm( ALARM_ID_DG_LOAD_CELL_PRIMARY_BACKUP_DRIFT_OUT_OF_RANGE, isDriftOutOfRange, driftDiff, - // LOAD_CELL_PRIMARY_BACKUP_MAX_ALLOWED_DRIFT_GRAMS ); - SET_ALARM_WITH_2_F32_DATA( ALARM_ID_DG_LOAD_CELL_PRIMARY_BACKUP_DRIFT_OUT_OF_RANGE, loadCellCurrentDrift, loadCellPreviousDrift ) + SET_ALARM_WITH_2_F32_DATA( ALARM_ID_HD_LOAD_CELL_PRIMARY_BACKUP_DRIFT_OUT_OF_RANGE, loadCellCurrentDrift, loadCellPreviousDrift ) } }