Index: firmware/App/HDCommon.h =================================================================== diff -u -r808bf52687216f0b8d81753a83983514b287238a -r56dcc6329c1da53e0da6949bba7568fa82f3a24c --- firmware/App/HDCommon.h (.../HDCommon.h) (revision 808bf52687216f0b8d81753a83983514b287238a) +++ firmware/App/HDCommon.h (.../HDCommon.h) (revision 56dcc6329c1da53e0da6949bba7568fa82f3a24c) @@ -54,11 +54,11 @@ // #define DISABLE_PUMP_FLOW_CHECKS 1 // Do not error on HD pump flow checks // #define DISABLE_PUMP_SPEED_CHECKS 1 // Do not error on HD pump speed checks // #define DISABLE_PUMP_DIRECTION_CHECKS 1 // Do not error on HD pump direction checks - #define DISABLE_SYRINGE_PUMP 1 // Disable syringe pump functionality +// #define DISABLE_SYRINGE_PUMP 1 // Disable syringe pump functionality #define ALWAYS_ALLOW_SYRINGE_PUMP_CMDS 1 // Allow syringe pump commands at any time except when pump is busy #define DISABLE_PRESSURE_CHECKS 1 // Do not error on HD pressure checks - #define DISABLE_ARTERIAL_PRESSURE_CHECK 1 // Do not error on arterial pressure out of range - #define DISABLE_VENOUS_PRESSURE_CHECK 1 // Do not error on venous pressure out of range +// #define DISABLE_ARTERIAL_PRESSURE_CHECK 1 // Do not error on arterial pressure out of range +// #define DISABLE_VENOUS_PRESSURE_CHECK 1 // Do not error on venous pressure out of range // #define DISABLE_UF_ALARMS 1 // Do not error on HD ultrafiltration checks #define DISABLE_DIALYSATE_TEMP_CHECK 1 // Disable dialysate temperature check @@ -76,7 +76,7 @@ // #define DISABLE_FPGA_COUNTER_CHECKS 1 // Disable alarms associated with FPGA read/error counters #define DISABLE_VOLTAGE_MONITOR 1 // Disable voltage monitoring/alarms #define ALLOW_1_MIN_TREATMENT_DURATION 1 // Allow user to change treatment duration to as low as 1 minute - #define DISABLE_SYRINGE_PUMP_ALARMS 1 // Disable some syringe pump alarms that are triggering intermittently +// #define DISABLE_SYRINGE_PUMP_ALARMS 1 // Disable some syringe pump alarms that are triggering intermittently // #define NO_PUMP_FLOW_LIMITS 1 // Allow any commanded flow rate for peristaltic pumps #define DISABLE_BUBBLE_ALARMS 1 // Disable bubble alarms #define DISABLE_UI_POST_TEST 1 // Disable the UI POST