Index: firmware/App/Modes/SelfTests.c =================================================================== diff -u -r80944744742c64276e19ace023836a2db7c9502e -r6ad2e9157e1c75b8a239440f96ab086303b0efe8 --- firmware/App/Modes/SelfTests.c (.../SelfTests.c) (revision 80944744742c64276e19ace023836a2db7c9502e) +++ firmware/App/Modes/SelfTests.c (.../SelfTests.c) (revision 6ad2e9157e1c75b8a239440f96ab086303b0efe8) @@ -941,14 +941,14 @@ previousNormalVenousPressure = getFilteredVenousPressure(); // Check to see if sensor is within normal ranges before we execute pressure sensor tests - if (previousNormalArterialPressure <= NORMAL_PRESSURE_LOWER_BOUNDARY || previousNormalArterialPressure >= NORMAL_PRESSURE_UPPER_BOUNDARY ) + if ( previousNormalArterialPressure <= NORMAL_PRESSURE_LOWER_BOUNDARY || previousNormalArterialPressure >= NORMAL_PRESSURE_UPPER_BOUNDARY ) { - SET_ALARM_WITH_1_F32_DATA( ALARM_ID_HD_ARTERIAL_PRESSURE_OUT_OF_RANGE, previousNormalArterialPressure ); + SET_ALARM_WITH_1_F32_DATA( ALARM_ID_PRE_TREATMENT_DRY_PRESSURE_TEST_FAILURE, previousNormalArterialPressure ); } - if (previousNormalVenousPressure <= NORMAL_PRESSURE_LOWER_BOUNDARY || previousNormalVenousPressure >= NORMAL_PRESSURE_UPPER_BOUNDARY ) + if ( previousNormalVenousPressure <= NORMAL_PRESSURE_LOWER_BOUNDARY || previousNormalVenousPressure >= NORMAL_PRESSURE_UPPER_BOUNDARY ) { - SET_ALARM_WITH_1_F32_DATA( ALARM_ID_HD_VENOUS_PRESSURE_OUT_OF_RANGE, previousNormalVenousPressure ); + SET_ALARM_WITH_1_F32_DATA( ALARM_ID_PRE_TREATMENT_DRY_PRESSURE_TEST_FAILURE, previousNormalVenousPressure ); } setValvePosition( VDI, VALVE_POSITION_C_CLOSE ); @@ -979,7 +979,7 @@ // End the test when reaching target pressure or time out if ( ( TRUE == didTimeout( pressureSelfTestBloodPumpRunStartTime, BLOOD_PUMP_RUN_TIME_PRESSURE_SELF_TEST ) ) || - ( ARTERIAL_PRESSURE_SELF_TEST_LOW_LIMIT_MMHG >= arterialPressure ) && ( VENOUS_PRESSURE_SELF_TEST_HIGH_LIMIT_MMHG <= venousPressure ) ) + ( ( ARTERIAL_PRESSURE_SELF_TEST_LOW_LIMIT_MMHG >= arterialPressure ) && ( VENOUS_PRESSURE_SELF_TEST_HIGH_LIMIT_MMHG <= venousPressure ) ) ) { signalBloodPumpHardStop(); // Test pass when reading positive arterial pressure and negative venous pressure