Index: firmware/App/Services/AlarmMgmt.c =================================================================== diff -u -rc9890bc0bc9a1d779be9fec7e0478cfc6fba1cfb -r849fffb475b8be4c24963e1c694e382a0a96ba1b --- firmware/App/Services/AlarmMgmt.c (.../AlarmMgmt.c) (revision c9890bc0bc9a1d779be9fec7e0478cfc6fba1cfb) +++ firmware/App/Services/AlarmMgmt.c (.../AlarmMgmt.c) (revision 849fffb475b8be4c24963e1c694e382a0a96ba1b) @@ -1358,8 +1358,8 @@ noEndTreatment = TRUE; } - // If in Treatment-Stop state or Fault/Service/Standby Mode, allow user to minimize the alarm window - if ( ( MODE_FAUL == currentMode ) || ( MODE_SERV == currentMode ) || ( MODE_STAN == currentMode ) ) + // If in Treatment/Fault/Service/Standby Mode, allow user to minimize the alarm window + if ( ( MODE_FAUL == currentMode ) || ( MODE_SERV == currentMode ) || ( MODE_STAN == currentMode ) || ( MODE_TREA == currentMode ) ) { noMinimize = FALSE; }