Index: firmware/App/Controllers/SyringePump.c =================================================================== diff -u -r7e86f5b6c5da2de71efd58a2d9bfcab43c426a08 -r9f39dcb0a661d2623ce48a3793c898524f6dbc6e --- firmware/App/Controllers/SyringePump.c (.../SyringePump.c) (revision 7e86f5b6c5da2de71efd58a2d9bfcab43c426a08) +++ firmware/App/Controllers/SyringePump.c (.../SyringePump.c) (revision 9f39dcb0a661d2623ce48a3793c898524f6dbc6e) @@ -1005,8 +1005,7 @@ // Check if syringe pump is on while BP is off { - BOOL runWhileOff = ( ( ( SYRINGE_PUMP_HEP_BOLUS_STATE == syringePumpState ) || ( SYRINGE_PUMP_HEP_CONTINUOUS_STATE == syringePumpState ) ) && - ( isBloodPumpRunning() != TRUE ) ); + BOOL runWhileOff = ( ( SYRINGE_PUMP_HEP_CONTINUOUS_STATE == syringePumpState ) && ( isBloodPumpRunning() != TRUE ) ); if ( TRUE == isPersistentAlarmTriggered( ALARM_ID_HD_SYRINGE_PUMP_RUNNING_WHILE_BP_OFF_ERROR, runWhileOff ) ) { @@ -1897,7 +1896,6 @@ } } - /*********************************************************************//** * @brief * The getHeprinBolusTargetRate function gets the heprin bolus target