Index: firmware/App/Controllers/Voltages.c =================================================================== diff -u -r0637f7a14c2a48a7c8af853af3559d68afbd462d -ra0cbd00ac6d6f19b724bf5e57ca43ad22e45f56d --- firmware/App/Controllers/Voltages.c (.../Voltages.c) (revision 0637f7a14c2a48a7c8af853af3559d68afbd462d) +++ firmware/App/Controllers/Voltages.c (.../Voltages.c) (revision a0cbd00ac6d6f19b724bf5e57ca43ad22e45f56d) @@ -241,13 +241,13 @@ HD_OP_MODE_T opMode = getCurrentOperationMode(); // Pre-Treatment, Treatment, or Post-Treatment mode - if ( ( MODE_PRET == opMode ) || ( MODE_TREA == opMode ) || ( MODE_POST == opMode ) ) // TODO - restore alarms when CPLDs updated in lab + if ( ( MODE_PRET == opMode ) || ( MODE_TREA == opMode ) || ( MODE_POST == opMode ) ) { -// SET_ALARM_WITH_2_F32_DATA( ALARM_ID_HD_AC_POWER_LOST_IN_TREATMENT, getMonitoredLineLevel( MONITORED_LINE_24V ), getMonitoredLineLevel( MONITORED_LINE_24V_REGEN ) ); + SET_ALARM_WITH_2_F32_DATA( ALARM_ID_HD_AC_POWER_LOST_IN_TREATMENT, getMonitoredLineLevel( MONITORED_LINE_24V ), getMonitoredLineLevel( MONITORED_LINE_24V_REGEN ) ); } else { -// SET_ALARM_WITH_2_F32_DATA( ALARM_ID_HD_AC_POWER_LOST, getMonitoredLineLevel( MONITORED_LINE_24V ), getMonitoredLineLevel( MONITORED_LINE_24V_REGEN ) ); + SET_ALARM_WITH_2_F32_DATA( ALARM_ID_HD_AC_POWER_LOST, getMonitoredLineLevel( MONITORED_LINE_24V ), getMonitoredLineLevel( MONITORED_LINE_24V_REGEN ) ); } } else