Index: firmware/App/Modes/Dialysis.c =================================================================== diff -u -rf760ffc4b10556e5186e9ceb90294262063440ca -ra9c1eee2bed2b569a6f32cbc5e80c4631559d683 --- firmware/App/Modes/Dialysis.c (.../Dialysis.c) (revision f760ffc4b10556e5186e9ceb90294262063440ca) +++ firmware/App/Modes/Dialysis.c (.../Dialysis.c) (revision a9c1eee2bed2b569a6f32cbc5e80c4631559d683) @@ -40,7 +40,7 @@ // ********** private definitions ********** #define MAX_UF_RATE_ML_PER_HOUR 2750.0F ///< Maximum ultrafiltration rate in mL/hour -#define MAX_UF_ACCURACY_ERROR_ML 250.0F ///< Maximum ultrafiltration accuracy error in mL over the entire treatment. +#define MAX_UF_ACCURACY_ERROR_ML 50.0F ///< Maximum ultrafiltration accuracy error in mL over the entire treatment. /// Saline bolus data broadcast interval (ms/task time) count. static const U32 SALINE_BOLUS_DATA_PUB_INTERVAL = ( MS_PER_SECOND / TASK_GENERAL_INTERVAL ); /// Ultrafiltration rate accuracy check interval count. @@ -1061,7 +1061,7 @@ } // Check total UF volume error - if ( ( fabs( refUFVolume - measUFVolume ) ) > (F32)MAX_UF_ACCURACY_ERROR_ML ) + if ( ( fabs( refUFVolume - measUFVolume ) ) > MAX_UF_ACCURACY_ERROR_ML ) { #ifndef _RELEASE_ if ( getSoftwareConfigStatus( SW_CONFIG_DISABLE_ULTRAFILTRATION_ALARMS ) != SW_CONFIG_ENABLE_VALUE )