Index: firmware/App/Controllers/PresOccl.c =================================================================== diff -u -racb7643f67efbc5f954c7f14704fe303196abfd8 -rb3d4ca79dc7533d0f20dd137d0f71b233219d2fe --- firmware/App/Controllers/PresOccl.c (.../PresOccl.c) (revision acb7643f67efbc5f954c7f14704fe303196abfd8) +++ firmware/App/Controllers/PresOccl.c (.../PresOccl.c) (revision b3d4ca79dc7533d0f20dd137d0f71b233219d2fe) @@ -516,7 +516,7 @@ else if ( MODE_TREA == getCurrentOperationMode() ) // All treatment states not covered by the if above { F32 artLowLimit = (F32)getS32TreatmentParamLowerRangeLimit( TREATMENT_PARAM_ART_PRESSURE_LOW_LIMIT ); - F32 artHighLimit = (F32)getS32TreatmentParamLowerRangeLimit( TREATMENT_PARAM_ART_PRESSURE_HIGH_LIMIT ); + F32 artHighLimit = (F32)getS32TreatmentParamUpperRangeLimit( TREATMENT_PARAM_ART_PRESSURE_HIGH_LIMIT ); BOOL artPresLow = ( artPres < artLowLimit ? TRUE : FALSE ); BOOL artPresHigh = ( artPres > artHighLimit ? TRUE : FALSE ); @@ -613,7 +613,7 @@ } if ( TRUE == isPersistentAlarmTriggered( ALARM_ID_VENOUS_PRESSURE_LOW, venPresLow ) ) { - SET_ALARM_WITH_2_F32_DATA( ALARM_ID_VENOUS_PRESSURE_LOW, venPres, venHighLimit ); + SET_ALARM_WITH_2_F32_DATA( ALARM_ID_VENOUS_PRESSURE_LOW, venPres, venLowLimit ); } } else