Index: firmware/App/Modes/Prime.c =================================================================== diff -u -r69ccdb2725e7aed091392a5b7d8d59afcd7afd33 -rba5306539f60ac7d27220fdc3173a22a5368f2ff --- firmware/App/Modes/Prime.c (.../Prime.c) (revision 69ccdb2725e7aed091392a5b7d8d59afcd7afd33) +++ firmware/App/Modes/Prime.c (.../Prime.c) (revision ba5306539f60ac7d27220fdc3173a22a5368f2ff) @@ -46,6 +46,7 @@ #define BLOOD_PUMP_FAST_FLOW_RATE_CIRC_BLOOD_CIRCUIT_ML_MIN 300 ///< Blood pump fast flow rate during prime recirculate blood circuit state. #define BLOOD_PUMP_FLOW_RATE_SALINE_DIALYZER_ML_MIN 300 ///< Blood pump flow rate during prime the saline dialyzer dialysate state. #define DIALYSATE_PUMP_PRIME_FLOW_RATE_ML_MIN 300 ///< Dialysate pump flow rate during priming fluid path. +#define DPO_PUMP_PRIME_FLOW_RATE_ML_MIN 225 ///< Dialysate outlet pump flow rate needed to match PWM with dialysate inlet pump @ 300 mL/min in open loop mode. #define DIALYSATE_DIALYZER_TUBE_VOLUME_ML 115 ///< This total tube volume is used to calculate the Dpi & Dpo time out in the dialysate dialyzer state. #define DIALYSATE_DIALYZER_BYPASS_TUBE_VOLUME_ML 75 ///< This volume is used to calculate the DPi pump time out in the dialyzer bypass state. @@ -606,7 +607,7 @@ signalBloodPumpHardStop(); setDialInPumpTargetFlowRate( DIALYSATE_PUMP_PRIME_FLOW_RATE_ML_MIN, MOTOR_DIR_FORWARD, PUMP_CONTROL_MODE_OPEN_LOOP ); - setDialOutPumpTargetRate( 225, MOTOR_DIR_FORWARD, PUMP_CONTROL_MODE_OPEN_LOOP ); // 225 target flow rate matches PWM duty cycle for DPi pump @ 300. + setDialOutPumpTargetRate( DPO_PUMP_PRIME_FLOW_RATE_ML_MIN, MOTOR_DIR_FORWARD, PUMP_CONTROL_MODE_OPEN_LOOP ); // Calculate the time out value that must passed prior to checking for the steady state volume in the reservoir primeDialysateDialyzerTimeLimit = (U32)( ( ( DIALYSATE_DIALYZER_TUBE_VOLUME_ML + dialyzerDialysateVolume ) * SEC_PER_MIN * MS_PER_SECOND ) / DIALYSATE_PUMP_PRIME_FLOW_RATE_ML_MIN );