Index: firmware/App/Modes/SelfTests.c =================================================================== diff -u -r86ccf9565595418e7424e5cd26b5abdd63482f04 -rbbd57f3327fd5835994a4ea7afc7cbeb36ce36de --- firmware/App/Modes/SelfTests.c (.../SelfTests.c) (revision 86ccf9565595418e7424e5cd26b5abdd63482f04) +++ firmware/App/Modes/SelfTests.c (.../SelfTests.c) (revision bbd57f3327fd5835994a4ea7afc7cbeb36ce36de) @@ -48,8 +48,8 @@ #define STABILTY_PRESSURE_SELF_TEST_TIME ( 4 * MS_PER_SECOND ) ///< Time to wait for pressure to stabilize in ms. #define DECAY_PRESSURE_SELF_TEST_TIME ( 2 * MS_PER_SECOND ) ///< time to wait for pressure to decay in ms. -#define ARTERIAL_PRESSURE_SELF_TEST_LOW_LIMIT_MMHG -50.0F ///< Arterial pressure low limit after running blood pump. -#define VENOUS_PRESSURE_SELF_TEST_HIGH_LIMIT_MMHG 400 ///< Venous pressure high limit after running blood pump. +#define ARTERIAL_PRESSURE_SELF_TEST_LOW_LIMIT_MMHG -80.0F ///< Arterial pressure low limit after running blood pump. +#define VENOUS_PRESSURE_SELF_TEST_HIGH_LIMIT_MMHG 40.0F ///< Venous pressure high limit after running blood pump. #define DECAY_PRESSURE_DIFF_TOLERANCE_MMHG 5.0F ///< Difference in pressure readings after the pump stops (in mmHg). #define STABILITY_PRESSURE_DIFF_TOLERANCE_MMHG 2.0F ///< Difference in pressure readings while in a stable pressured state (in mmHg).