Index: firmware/App/Services/AlarmMgmt.c =================================================================== diff -u -re12fd5b499b9d32bd21b43cd4d178dd0241aa1e3 -rbf7d9287c229b4b921ed3f0fb3b2f112f4fa1ba5 --- firmware/App/Services/AlarmMgmt.c (.../AlarmMgmt.c) (revision e12fd5b499b9d32bd21b43cd4d178dd0241aa1e3) +++ firmware/App/Services/AlarmMgmt.c (.../AlarmMgmt.c) (revision bf7d9287c229b4b921ed3f0fb3b2f112f4fa1ba5) @@ -1390,8 +1390,8 @@ noEndTreatment = TRUE; } - // If in Treatment-Stop state or Fault/Service/Standby Mode, allow user to minimize the alarm window - if ( ( MODE_FAUL == currentMode ) || ( MODE_SERV == currentMode ) || ( MODE_STAN == currentMode ) ) + // If in Treatment/Fault/Service/Standby Mode, allow user to minimize the alarm window + if ( ( MODE_FAUL == currentMode ) || ( MODE_SERV == currentMode ) || ( MODE_STAN == currentMode ) || ( MODE_TREA == currentMode ) ) { noMinimize = FALSE; }