Index: firmware/App/Controllers/PresOccl.c =================================================================== diff -u -r727f30438c094ec62abe1fad1f5a8962cd304a4d -rced4d5be0d18c9b1001143206d10e09fd15f8efb --- firmware/App/Controllers/PresOccl.c (.../PresOccl.c) (revision 727f30438c094ec62abe1fad1f5a8962cd304a4d) +++ firmware/App/Controllers/PresOccl.c (.../PresOccl.c) (revision ced4d5be0d18c9b1001143206d10e09fd15f8efb) @@ -81,6 +81,7 @@ #define PSI_TO_MMHG ( 51.7149F ) ///< Conversion factor for converting PSI to mmHg. +// The new arterial pressure sensor is the same as the venous pressure sensor #define VENOUS_PRESSURE_NORMAL_OP 0 ///< Venous pressure status bits indicate normal operation. #define VENOUS_PRESSURE_CMD_MODE 1 ///< Venous pressure status bits indicate sensor in command mode. #define VENOUS_PRESSURE_STALE_DATA 2 ///< Venous pressure status bits indicate data is stale (no new data since last fpga read). @@ -874,7 +875,6 @@ } } } - // Check for occlusion in Treatment modes where pumps are moving else if ( MODE_TREA == getCurrentOperationMode() ) { Index: firmware/App/Modes/SelfTests.c =================================================================== diff -u -r727f30438c094ec62abe1fad1f5a8962cd304a4d -rced4d5be0d18c9b1001143206d10e09fd15f8efb --- firmware/App/Modes/SelfTests.c (.../SelfTests.c) (revision 727f30438c094ec62abe1fad1f5a8962cd304a4d) +++ firmware/App/Modes/SelfTests.c (.../SelfTests.c) (revision ced4d5be0d18c9b1001143206d10e09fd15f8efb) @@ -117,6 +117,7 @@ static BOOL wetSelfTestsResult; ///< Result of wet self-tests. static WET_SELF_TESTS_STATE_T currentWetSelfTestsState; ///< Current state of the wet self-tests state machine. +static WET_SELF_TESTS_STATE_T checkpointWetSelfTestsState; ///< Checkpoint state of the wet self tests state machine. static U32 settleStartTime; ///< Wait for reservoir to settle start time. static U32 displacementStartTime; ///< Dialysate displacement starting time. static F32 fmdIntegratedVolume; ///< FMD integrated volume over displacement time. @@ -584,6 +585,7 @@ { wetSelfTestsResult = FALSE; currentWetSelfTestsState = WET_SELF_TESTS_START_STATE; + checkpointWetSelfTestsState = WET_SELF_TESTS_START_STATE; settleStartTime = 0; displacementStartTime = 0; fmdIntegratedVolume = 0.0; @@ -1620,7 +1622,7 @@ * The handleWetSelfTestSetupState function setup for bubble self-test * check. * @details Inputs: none - * @details Outputs: bubble detector self-test requested + * @details Outputs: bubble detector self-test requested, checkpointWetSelfTestsState * @return the next state of wet self-tests state machine *************************************************************************/ static WET_SELF_TESTS_STATE_T handleWetSelfTestSetupState( void ) @@ -1631,6 +1633,7 @@ if ( TRUE == doesAlarmStatusIndicateStop() ) { + checkpointWetSelfTestsState = WET_SELF_TESTS_BUBBLE_CHECK_STATE; state = WET_SELF_TESTS_STOPPED_STATE; setupForSelfTestsStop(); } @@ -1643,7 +1646,7 @@ * The handleWetSelfTestBubblesState function waiting for air bubble detectors * self-tests to finish. * @details Inputs: bubbleSelfTestStatus - * @details Outputs: none + * @details Outputs: checkpointWetSelfTestsState * @return the next state of wet self-tests state machine *************************************************************************/ static WET_SELF_TESTS_STATE_T handleWetSelfTestBubblesState( void ) @@ -1657,6 +1660,7 @@ if ( TRUE == doesAlarmStatusIndicateStop() ) { + checkpointWetSelfTestsState = WET_SELF_TESTS_BUBBLE_CHECK_STATE; state = WET_SELF_TESTS_STOPPED_STATE; setupForSelfTestsStop(); } @@ -1669,7 +1673,7 @@ * The handleWetSelfTestPrimeCheckState function checks arterial and venous * lines to make sure they are primed. * @details Inputs: Arterial and venous bubble detectors status - * @details Outputs: Test pass/fail + * @details Outputs: Test pass/fail, checkpointWetSelfTestsState * @return the next state of wet self-tests state machine *************************************************************************/ static WET_SELF_TESTS_STATE_T handleWetSelfTestPrimeCheckState( void ) @@ -1691,6 +1695,7 @@ if ( TRUE == doesAlarmStatusIndicateStop() ) { + checkpointWetSelfTestsState = WET_SELF_TESTS_PRIME_CHECK_STATE; state = WET_SELF_TESTS_STOPPED_STATE; setupForSelfTestsStop(); } @@ -1739,7 +1744,7 @@ * self-test for blood leak detector. * @details Inputs: none * @details Outputs: bloodLeakDebubbleStartTimeMS, settleStartTime, - * isValvesSettingSent + * isValvesSettingSent, checkpointWetSelfTestsState * @return the next state of wet self-tests state machine *************************************************************************/ static WET_SELF_TESTS_STATE_T handleWetSelfTestBloodLeakDetectorState( void ) @@ -1776,6 +1781,7 @@ if ( TRUE == doesAlarmStatusIndicateStop() ) { + checkpointWetSelfTestsState = WET_SELF_TESTS_BLOOD_LEAK_DETECTOR_STATE; state = WET_SELF_TESTS_STOPPED_STATE; setupForSelfTestsStop(); } @@ -1788,7 +1794,7 @@ * The handleWetSelfTestStartFirstDisplacementState function setups the valves * and pumps to start first dialysate displacement. * @details Inputs: settleStartTime, isValvesSettingSent - * @details Outputs: isValvesSettingSent + * @details Outputs: isValvesSettingSent, checkpointWetSelfTestsState * @return the next state of wet self-tests state machine *************************************************************************/ static WET_SELF_TESTS_STATE_T handleWetSelfTestFirstDisplacementSetupState( void ) @@ -1821,6 +1827,7 @@ if ( TRUE == doesAlarmStatusIndicateStop() ) { + checkpointWetSelfTestsState = WET_SELF_TESTS_FIRST_DISPLACEMENT_SETUP_STATE; state = WET_SELF_TESTS_STOPPED_STATE; setupForSelfTestsStop(); } @@ -1833,7 +1840,7 @@ * The handleWetSelfTestFirstDisplacementState function handles the first * dialysate displacement from reservoir 1 to reservoir 2. * @details Inputs: displacementStartTime, fmdIntegratedVolume, settleStartTime - * @details Outputs: settleStartTime, displacementStartTime + * @details Outputs: settleStartTime, displacementStartTime, checkpointWetSelfTestsState * @return the next state of wet self-tests state machine *************************************************************************/ static WET_SELF_TESTS_STATE_T handleWetSelfTestFirstDisplacementState( void ) @@ -1866,6 +1873,7 @@ if ( TRUE == doesAlarmStatusIndicateStop() ) { + checkpointWetSelfTestsState = WET_SELF_TESTS_FIRST_DISPLACEMENT_STATE; state = WET_SELF_TESTS_STOPPED_STATE; setupForSelfTestsStop(); } @@ -1878,7 +1886,7 @@ * The handleWetSelfTestFirstDisplacementVerifyState function checks the load cell * readings and FMD integrated volume after the first dialysate displacement. * @details Inputs: settleStartTime, fmdIntegratedVolume - * @details Outputs: none + * @details Outputs: checkpointWetSelfTestsState * @return the next state of wet self-tests state machine *************************************************************************/ static WET_SELF_TESTS_STATE_T handleWetSelfTestFirstDisplacementVerifyState( void ) @@ -1928,6 +1936,7 @@ if ( TRUE == doesAlarmStatusIndicateStop() ) { + checkpointWetSelfTestsState = WET_SELF_TESTS_FIRST_DISPLACEMENT_VERIFY_STATE; state = WET_SELF_TESTS_STOPPED_STATE; setupForSelfTestsStop(); } @@ -1940,7 +1949,7 @@ * The handleWetSelfTestSecondDisplacementSetupState function setups the valves * and pumps to start second dialysate displacement. * @details Inputs: settleStartTime, isValvesSettingSent - * @details Outputs: isValvesSettingSent + * @details Outputs: isValvesSettingSent, checkpointWetSelfTestsState * @return the next state of wet self-tests state machine *************************************************************************/ static WET_SELF_TESTS_STATE_T handleWetSelfTestSecondDisplacementSetupState( void ) @@ -1973,6 +1982,7 @@ if ( TRUE == doesAlarmStatusIndicateStop() ) { + checkpointWetSelfTestsState = WET_SELF_TESTS_SECOND_DISPLACEMENT_SETUP_STATE; state = WET_SELF_TESTS_STOPPED_STATE; setupForSelfTestsStop(); } @@ -1985,7 +1995,7 @@ * The handleWetSelfTestSecondDisplacementState function handles the first * dialysate displacement from reservoir 2 to reservoir 1. * @details Inputs: displacementStartTime, fmdIntegratedVolume, settleStartTime - * @details Outputs: displacementStartTime + * @details Outputs: displacementStartTime, checkpointWetSelfTestsState * @return the next state of wet self-tests state machine *************************************************************************/ static WET_SELF_TESTS_STATE_T handleWetSelfTestSecondDisplacementState( void ) @@ -2016,6 +2026,7 @@ if ( TRUE == doesAlarmStatusIndicateStop() ) { + checkpointWetSelfTestsState = WET_SELF_TESTS_SECOND_DISPLACEMENT_STATE; state = WET_SELF_TESTS_STOPPED_STATE; setupForSelfTestsStop(); } @@ -2028,7 +2039,7 @@ * The handleWetSelfTestSecondDisplacementVerifyState function checks the load cell * readings and FMD integrated volume after the second dialysate displacement. * @details Inputs: settleStartTime, reservoirVolume[], reservoirs' weights - * @details Outputs: verify correctness of dialysate flow meter and load cell + * @details Outputs: verify correctness of dialysate flow meter and load cell, checkpointWetSelfTestsState * @return the next state of wet self-tests state machine *************************************************************************/ static WET_SELF_TESTS_STATE_T handleWetSelfTestSecondDisplacementVerifyState( void ) @@ -2078,6 +2089,7 @@ if ( TRUE == doesAlarmStatusIndicateStop() ) { + checkpointWetSelfTestsState = WET_SELF_TESTS_SECOND_DISPLACEMENT_VERIFY_STATE; state = WET_SELF_TESTS_STOPPED_STATE; setupForSelfTestsStop(); } @@ -2089,7 +2101,7 @@ * @brief * The handleWetSelfTestStoppedState function handles the stopped wet self-tests * operation. - * @details Inputs: selfTestsResumeRequested + * @details Inputs: selfTestsResumeRequested, checkpointWetSelfTestsState * @details Outputs: selfTestsResumeRequested, selfTestStartTime * @return the next state of wet self-tests state machine *************************************************************************/ @@ -2106,7 +2118,37 @@ selfTestStartTime = getMSTimerCount(); doorClosedRequired( TRUE, TRUE ); selfTestsResumeRequested = FALSE; - state = WET_SELF_TESTS_START_STATE; // TODO - should start state make sure there is sufficient/appropriate res volumes to restart wet self-tests? and if not, should it signal res mgmt to drain and fill R1 and R2 before resuming wet self-tests? + + switch( checkpointWetSelfTestsState ) + { + case WET_SELF_TESTS_BUBBLE_CHECK_SETUP_STATE: + case WET_SELF_TESTS_BUBBLE_CHECK_STATE: + case WET_SELF_TESTS_PRIME_CHECK_STATE: + case WET_SELF_TESTS_BLOOD_LEAK_DETECTOR_STATE: + state = WET_SELF_TESTS_START_STATE; + break; + + case WET_SELF_TESTS_FIRST_DISPLACEMENT_SETUP_STATE: + case WET_SELF_TESTS_FIRST_DISPLACEMENT_STATE: + state = WET_SELF_TESTS_FIRST_DISPLACEMENT_SETUP_STATE; + break; + + case WET_SELF_TESTS_FIRST_DISPLACEMENT_VERIFY_STATE: + state = WET_SELF_TESTS_SECOND_DISPLACEMENT_SETUP_STATE; + break; + + case WET_SELF_TESTS_SECOND_DISPLACEMENT_SETUP_STATE: + case WET_SELF_TESTS_SECOND_DISPLACEMENT_STATE: + state = WET_SELF_TESTS_SECOND_DISPLACEMENT_SETUP_STATE; + break; + + case WET_SELF_TESTS_SECOND_DISPLACEMENT_VERIFY_STATE: + case WET_SELF_TESTS_COMPLETE_STATE: + state = WET_SELF_TESTS_COMPLETE_STATE; + + default: + state = WET_SELF_TESTS_START_STATE; + } } return state;