Index: firmware/App/Modes/Prime.c =================================================================== diff -u -r486f68528804a67c4632bb7fd530fae4cb63cd36 -rd367cda979e9684a25b0d4d5df52bbcf4f3c0acf --- firmware/App/Modes/Prime.c (.../Prime.c) (revision 486f68528804a67c4632bb7fd530fae4cb63cd36) +++ firmware/App/Modes/Prime.c (.../Prime.c) (revision d367cda979e9684a25b0d4d5df52bbcf4f3c0acf) @@ -57,7 +57,6 @@ #define NO_AIR_DETECTED_COUNT ( 40 * MS_PER_SECOND ) ///< No air detected time period count. #define PURGE_AIR_TIME_OUT_COUNT ( 240 * MS_PER_SECOND ) ///< Time period count for purge air time out. -#define PRIME_SALINE_DIALYZER_TIME_OUT_COUNT ( 60 * MS_PER_SECOND ) ///< Time period count for prime saline dialyzer time out. #define LOAD_CELL_STEADY_VOLUME_SAMPLING_TIME ( 1 * MS_PER_SECOND ) ///< Time load cell reading steady state detection sampling time in seconds. #define PRIME_DIALYSATE_BYPASS_TIME_LIMIT ( 8 * MS_PER_SECOND ) ///< Time limit for priming dialysate bypass circuit. #define STEADY_VOLUME_COUNT_SEC ( 10000 / LOAD_CELL_STEADY_VOLUME_SAMPLING_TIME ) ///< Counter must be greater than 10 seconds before steady volume is true. @@ -864,11 +863,6 @@ { state = HD_PRIME_RESERVOIR_TWO_FILL_COMPLETE_STATE; } - // TODO: Rework alarm and timeout logic here. PRIME_SALINE_DIALYZER_TIME_OUT_COUNT may not be reached. - if ( TRUE == didTimeout( primeSalineDialyzerStartTime, PRIME_SALINE_DIALYZER_TIME_OUT_COUNT ) ) - { - SET_ALARM_WITH_1_U32_DATA( ALARM_ID_HD_PRIME_SALINE_DIALYZER_TIME_OUT, PRIME_SALINE_DIALYZER_TIME_OUT_COUNT ); // Trigger HD prime saline dialyzer time out alarm. - } if ( TRUE == doesAlarmStatusIndicateStop() ) {