Index: firmware/App/Modes/Prime.c =================================================================== diff -u -r80d2387215e98241cbadf36eb8d42e79812ec578 -rd4a4438fece666d728dc9825157af6ca99be8eb9 --- firmware/App/Modes/Prime.c (.../Prime.c) (revision 80d2387215e98241cbadf36eb8d42e79812ec578) +++ firmware/App/Modes/Prime.c (.../Prime.c) (revision d4a4438fece666d728dc9825157af6ca99be8eb9) @@ -87,7 +87,7 @@ { 120, 257 }, // DIALYZER_TYPE_BBRAUN_PRO_19H { 87, 233 }, // DIALYZER_TYPE_FRESENIUS_OPTIFLUX_F160NRE { 102, 280 }, // DIALYZER_TYPE_FRESENIUS_OPTIFLUX_F180NRE - { 117, 327 } }; // DIALYZER_TYPE_FRESENIUS_OPTIFLUX_F200NRE + { 113, 300 } }; // DIALYZER_TYPE_FRESENIUS_OPTIFLUX_F200NRE static U32 primeDialysateDialyzerTimeLimit; ///< Time limit in msec for priming dialysate dialyzer path. static U32 primeSalineDialyzerTimeLimit; ///< Time limit in msec for priming saline dialyzer path. Index: firmware/App/Modes/Rinseback.c =================================================================== diff -u -rcbaa57803fa8359407a799fca0cde6ae2c486bd7 -rd4a4438fece666d728dc9825157af6ca99be8eb9 --- firmware/App/Modes/Rinseback.c (.../Rinseback.c) (revision cbaa57803fa8359407a799fca0cde6ae2c486bd7) +++ firmware/App/Modes/Rinseback.c (.../Rinseback.c) (revision d4a4438fece666d728dc9825157af6ca99be8eb9) @@ -210,8 +210,8 @@ static void setupForRinsebackDelivery( U32 rate ) { // Open VBA and VBV valves to allow flow from saline bag and to patient venous line - setValvePosition( VBA, VALVE_POSITION_C_CLOSE ); // draw from saline back instead of patient - setValvePosition( VBV, VALVE_POSITION_B_OPEN ); // return to patient + setValvePosition( VBA, VALVE_POSITION_B_OPEN ); // patient will have removed arterial line from fistula and connected it to RB port to saline - so we will get saline from arterial line by opening + setValvePosition( VBV, VALVE_POSITION_B_OPEN ); // return to patient // Start blood pump at rinseback flow rate setBloodPumpTargetFlowRate( rate, MOTOR_DIR_FORWARD, PUMP_CONTROL_MODE_CLOSED_LOOP ); // Start air trap leveling control