Index: firmware/App/Modes/Dialysis.c =================================================================== diff -u -r61716bc97ecca8af1ec560333844a8cf602eccb0 -rd9c9c94bf66bc7030f0528ec18608eb6863804eb --- firmware/App/Modes/Dialysis.c (.../Dialysis.c) (revision 61716bc97ecca8af1ec560333844a8cf602eccb0) +++ firmware/App/Modes/Dialysis.c (.../Dialysis.c) (revision d9c9c94bf66bc7030f0528ec18608eb6863804eb) @@ -47,7 +47,7 @@ static const U32 UF_ACCURACY_CHECK_INTERVAL = ((1 * MIN_PER_HOUR * SEC_PER_MIN * MS_PER_SECOND) / TASK_GENERAL_INTERVAL); #define MAX_SALINE_VOLUME_DELIVERED 800 ///< Maximum saline volume delivered for a treatment. -#define SALINE_BOLUS_RATE_ML_MIN 150 ///< Fixed rate for saline bolus delivery. +#define SALINE_BOLUS_RATE_ML_MIN 300 ///< Fixed rate for saline bolus delivery. #define MAX_BOLUS_ERROR_PCT 0.2 ///< Maximum error in saline bolus volume delivered (as a percentage of target). #define MAX_ACTIVE_LOAD_CELL_CHANGE_G 50.0 ///< Maximum delta between new and previous measured UF volume.