Index: firmware/App/Modes/Prime.c =================================================================== diff -u -r68e6a5f41d4e9decee2658b3421f34892c9b9ac7 -rdde8219d4ba764374366e204cbf391b5e901e43f --- firmware/App/Modes/Prime.c (.../Prime.c) (revision 68e6a5f41d4e9decee2658b3421f34892c9b9ac7) +++ firmware/App/Modes/Prime.c (.../Prime.c) (revision dde8219d4ba764374366e204cbf391b5e901e43f) @@ -830,10 +830,8 @@ signalDialOutPumpHardStop(); signalDialInPumpHardStop(); setBloodPumpTargetFlowRate( BLOOD_PUMP_FLOW_RATE_SALINE_DIALYZER_ML_MIN, MOTOR_DIR_FORWARD, PUMP_CONTROL_MODE_OPEN_LOOP ); - setBloodPumpTargetFlowRate( 100, MOTOR_DIR_FORWARD, PUMP_CONTROL_MODE_OPEN_LOOP ); primeSalineDialyzerTimeLimit = (U32)( ( ( dialyzerDialysateVolume * DIALYZER_VOLUME_SCALE_FACTOR + DIALYZER_DVI_PATH_VOLUME_ML ) * SEC_PER_MIN * MS_PER_SECOND ) / BLOOD_PUMP_FLOW_RATE_SALINE_DIALYZER_ML_MIN ); - primeSalineDialyzerTimeLimit = (U32)( ( ( dialyzerDialysateVolume * DIALYZER_VOLUME_SCALE_FACTOR + DIALYZER_DVI_PATH_VOLUME_ML ) * SEC_PER_MIN * MS_PER_SECOND ) / 100.0F ); primeSalineDialyzerStartTime = getMSTimerCount(); if ( TRUE == doesAlarmStatusIndicateStop() )