Index: firmware/App/Controllers/PresOccl.c =================================================================== diff -u -rc589a601707829f3886efe4d18fba31067639e2b -re7d8cb4e5190a4183ef39d48f240ce23c623f3a7 --- firmware/App/Controllers/PresOccl.c (.../PresOccl.c) (revision c589a601707829f3886efe4d18fba31067639e2b) +++ firmware/App/Controllers/PresOccl.c (.../PresOccl.c) (revision e7d8cb4e5190a4183ef39d48f240ce23c623f3a7) @@ -396,7 +396,7 @@ // Check for stale venous pressure reading if ( TRUE == isPersistentAlarmTriggered( ALARM_ID_HD_ARTERIAL_PRESSURE_READ_TIMEOUT_ERROR, ( lastArterialPressureReadCtr == artReadCtr ) ) ) { - activateAlarmNoData( ALARM_ID_HD_ARTERIAL_PRESSURE_READ_TIMEOUT_ERROR ); + //activateAlarmNoData( ALARM_ID_HD_ARTERIAL_PRESSURE_READ_TIMEOUT_ERROR ); } // Record arterial pressure sensor read counter for next time around lastArterialPressureReadCtr = artReadCtr; @@ -420,7 +420,7 @@ } else { - SET_ALARM_WITH_1_U32_DATA( ALARM_ID_HD_ARTERIAL_PRESSURE_READ_TIMEOUT_ERROR, (U32)artErrorCtr ); + //SET_ALARM_WITH_1_U32_DATA( ALARM_ID_HD_ARTERIAL_PRESSURE_READ_TIMEOUT_ERROR, (U32)artErrorCtr ); } // Record arterial pressure sensor read counter for next time around lastArterialPressureReadCtr = artReadCtr;