Index: firmware/App/Modes/Prime.c =================================================================== diff -u -r645a316a9cda57a4d7561080add53dd12bbe4770 -rf9d86272e3f352766e48ecc71189c0b6c8106bc1 --- firmware/App/Modes/Prime.c (.../Prime.c) (revision 645a316a9cda57a4d7561080add53dd12bbe4770) +++ firmware/App/Modes/Prime.c (.../Prime.c) (revision f9d86272e3f352766e48ecc71189c0b6c8106bc1) @@ -39,7 +39,7 @@ #define MAX_PRIME_TIME ( 10 * SEC_PER_MIN ) ///< Maximum prime time (in seconds). #define PRIME_DATA_PUB_INTERVAL ( MS_PER_SECOND / TASK_GENERAL_INTERVAL ) ///< Interval (ms/task time) at which the prime data is published on the CAN bus. -#define BLOOD_PUMP_FAST_FLOW_RATE_PURGE_AIR_ML_MIN 400 ///< Blood pump fast flow rate to fill fluid. +#define BLOOD_PUMP_FAST_FLOW_RATE_PURGE_AIR_ML_MIN 300 ///< Blood pump fast flow rate to fill fluid. #define BLOOD_PUMP_SLOW_FLOW_RATE_PURGE_AIR_ML_MIN 150 ///< Blood pump slow flow rate after fluid reach lower level of air trap sensor. #define BLOOD_PUMP_SALINE_FLOW_RATE_PURGE_AIR_ML_MIN 200 ///< Blood pump very slow flow rate during prime saline dialyzer state #define BLOOD_PUMP_FLOW_RATE_CIRC_BLOOD_CIRCUIT_ML_MIN 300 ///< Blood pump flow rate during prime recirculate blood circuit state. @@ -628,10 +628,10 @@ } else { - state = HD_PRIME_RESERVOIR_TWO_FILL_COMPLETE_STATE; + state = HD_PRIME_SALINE_DIALYZER_SETUP_STATE; } #else - state = HD_PRIME_RESERVOIR_TWO_FILL_COMPLETE_STATE; + state = HD_PRIME_SALINE_DIALYZER_SETUP_STATE; #endif } }