Index: etc/reports/alarms/AlarmMapping.csv =================================================================== diff -u -r1244c4ea38fdf2236ac2e4ce052e9da2200b43fa -rcad35b1aa4f7428a64cca2cfb8f1d2ef44645933 --- etc/reports/alarms/AlarmMapping.csv (.../AlarmMapping.csv) (revision 1244c4ea38fdf2236ac2e4ce052e9da2200b43fa) +++ etc/reports/alarms/AlarmMapping.csv (.../AlarmMapping.csv) (revision cad35b1aa4f7428a64cca2cfb8f1d2ef44645933) @@ -206,9 +206,9 @@ 205,"TD UI POST failed." 206,"TD and UI software builds are not compatible." 207,"TD POST Safety Shutdown failure." - 208,"FW reserved for use." - 209,"FW reserved for use." - 210,"FW reserved for use." + 208,"TD initial rev blood circuit prime limit reached." + 209,"TD initial blood circuit prime timeout." + 210,"TD prime complete." 211,"FW reserved for use." 212,"FW reserved for use." 213,"TD blood pressure systolic low." @@ -254,19 +254,19 @@ 253,"FW reserved for use." 254,"FW reserved for use." 255,"FW reserved for use." - 256,"FW reserved for use." - 257,"FW reserved for use." - 258,"FW reserved for use." - 259,"FW reserved for use." - 260,"FW reserved for use." - 261,"FW reserved for use." - 262,"FW reserved for use." - 263,"FW reserved for use." - 264,"FW reserved for use." - 265,"FW reserved for use." - 266,"FW reserved for use." - 267,"FW reserved for use." - 268,"FW reserved for use." + 256,"TD invalid tubing set." + 257,"TD install new blood tubing set." + 258,"TD no tubing set detected." + 259,"TD tubing set authentication timeout." + 260,"TD pressure too high to zero." + 261,"TD venous pressurization timeout." + 262,"TD venous pressure decay failure." + 263,"TD venous pressure leak." + 264,"TD venous pressure relief failure." + 265,"TD arterial pressurization timeout." + 266,"TD arterial pressure decay failure." + 267,"TD arterial pressure leak." + 268,"TD arterial pressure relief failure." 269,"FW reserved for use." 270,"FW reserved for use." 271,"FW reserved for use."