Index: etc/reports/alarms/AlarmMapping.py =================================================================== diff -u -r1244c4ea38fdf2236ac2e4ce052e9da2200b43fa -rcad35b1aa4f7428a64cca2cfb8f1d2ef44645933 --- etc/reports/alarms/AlarmMapping.py (.../AlarmMapping.py) (revision 1244c4ea38fdf2236ac2e4ce052e9da2200b43fa) +++ etc/reports/alarms/AlarmMapping.py (.../AlarmMapping.py) (revision cad35b1aa4f7428a64cca2cfb8f1d2ef44645933) @@ -207,9 +207,9 @@ 205 : "TD UI POST failed." 206 : "TD and UI software builds are not compatible." 207 : "TD POST Safety Shutdown failure." - 208 : "FW reserved for use." - 209 : "FW reserved for use." - 210 : "FW reserved for use." + 208 : "TD initial rev blood circuit prime limit reached." + 209 : "TD initial blood circuit prime timeout." + 210 : "TD prime complete." 211 : "FW reserved for use." 212 : "FW reserved for use." 213 : "TD blood pressure systolic low." @@ -255,19 +255,19 @@ 253 : "FW reserved for use." 254 : "FW reserved for use." 255 : "FW reserved for use." - 256 : "FW reserved for use." - 257 : "FW reserved for use." - 258 : "FW reserved for use." - 259 : "FW reserved for use." - 260 : "FW reserved for use." - 261 : "FW reserved for use." - 262 : "FW reserved for use." - 263 : "FW reserved for use." - 264 : "FW reserved for use." - 265 : "FW reserved for use." - 266 : "FW reserved for use." - 267 : "FW reserved for use." - 268 : "FW reserved for use." + 256 : "TD invalid tubing set." + 257 : "TD install new blood tubing set." + 258 : "TD no tubing set detected." + 259 : "TD tubing set authentication timeout." + 260 : "TD pressure too high to zero." + 261 : "TD venous pressurization timeout." + 262 : "TD venous pressure decay failure." + 263 : "TD venous pressure leak." + 264 : "TD venous pressure relief failure." + 265 : "TD arterial pressurization timeout." + 266 : "TD arterial pressure decay failure." + 267 : "TD arterial pressure leak." + 268 : "TD arterial pressure relief failure." 269 : "FW reserved for use." 270 : "FW reserved for use." 271 : "FW reserved for use."