Index: sources/model/hd/alarm/MAlarmMapping.cpp =================================================================== diff -u -r994d6d453820ba52adc987d81a2dbd54356a40eb -r22c9f4d0c1430dae4faf45961aadc7fd35fd10aa --- sources/model/hd/alarm/MAlarmMapping.cpp (.../MAlarmMapping.cpp) (revision 994d6d453820ba52adc987d81a2dbd54356a40eb) +++ sources/model/hd/alarm/MAlarmMapping.cpp (.../MAlarmMapping.cpp) (revision 22c9f4d0c1430dae4faf45961aadc7fd35fd10aa) @@ -61,11 +61,11 @@ /*0025*/case GuiAlarmID::ALARM_ID_DD_VALVE_CONTROL_FAILURE : { result = QObject::tr("DD FPGA not accepting commanded valve states." ); break; } /* 25*/ /*0026*/case GuiAlarmID::ALARM_ID_TD_ARTERIAL_SENSOR_TIMEOUT_FAULT : { result = QObject::tr("TD arterial pressure sensor freshness/error timeout fault." ); break; } /* 26*/ /*0027*/case GuiAlarmID::ALARM_ID_TD_VENOUS_SENSOR_TIMEOUT_FAULT : { result = QObject::tr("TD venous pressure sensor freshness/error timeout fault." ); break; } /* 27*/ -/*0028*/case GuiAlarmID::ALARM_ID_DD_HYD_OUTLET_PRES_TIMEOUT_FAULT : { result = QObject::tr("DD hydraulics outlet pressure sensor freshness/error timeout fault." ); break; } /* 28*/ -/*0029*/case GuiAlarmID::ALARM_ID_DD_BIBAG_PRES_TIMEOUT_FAULT : { result = QObject::tr("DD bibag pressure sensor freshness/error timeout fault." ); break; } /* 29*/ -/*0030*/case GuiAlarmID::ALARM_ID_DD_SPENT_DIALYSATE_PRES_TIMEOUT_FAULT : { result = QObject::tr("DD spent dialysate pressure sensor freshness/error timeout fault." ); break; } /* 30*/ -/*0031*/case GuiAlarmID::ALARM_ID_DD_FRESH_DIALYSATE_PRES_TIMEOUT_FAULT : { result = QObject::tr("DD fresh dialysate pressure sensor freshness/error timeout fault." ); break; } /* 31*/ -/*0032*/case GuiAlarmID::ALARM_ID_DD_TRANSMEMB_PRES_TIMEOUT_FAULT : { result = QObject::tr("DD transmembrane pressure sensor freshness/error timeout fault." ); break; } /* 32*/ +/*0028*/case GuiAlarmID::ALARM_ID_AVAILABLE_28 : { result = QObject::tr("Available alarms." ); break; } /* 28*/ +/*0029*/case GuiAlarmID::ALARM_ID_AVAILABLE_29 : { result = QObject::tr("Available alarms." ); break; } /* 29*/ +/*0030*/case GuiAlarmID::ALARM_ID_AVAILABLE_30 : { result = QObject::tr("Available alarms." ); break; } /* 30*/ +/*0031*/case GuiAlarmID::ALARM_ID_AVAILABLE_31 : { result = QObject::tr("Available alarms." ); break; } /* 31*/ +/*0032*/case GuiAlarmID::ALARM_ID_AVAILABLE_32 : { result = QObject::tr("Available alarms." ); break; } /* 32*/ /*0033*/case GuiAlarmID::ALARM_ID_TD_VOLTAGE_OUT_OF_RANGE : { result = QObject::tr("TD voltage out of range fault." ); break; } /* 33*/ /*0034*/case GuiAlarmID::ALARM_ID_TD_AC_POWER_LOST : { result = QObject::tr("TD AC power lost." ); break; } /* 34*/ /*0035*/case GuiAlarmID::ALARM_ID_TD_AC_POWER_LOST_IN_TREATMENT : { result = QObject::tr("TD AC power lost during treatment." ); break; } /* 35*/ @@ -91,7 +91,7 @@ /*0055*/case GuiAlarmID::ALARM_ID_DD_CONCENTRATE_PUMP_FAULT : { result = QObject::tr("DD concentrate pump fault." ); break; } /* 55*/ /*0056*/case GuiAlarmID::ALARM_ID_DD_TEMPERATURE_SENSOR_OUT_OF_RANGE : { result = QObject::tr("DD temperature sensor out of range." ); break; } /* 56*/ /*0057*/case GuiAlarmID::ALARM_ID_DD_RTD_SENSORS_FPGA_FAULT : { result = QObject::tr("DD temperature sensors freshness timeout fault." ); break; } /* 57*/ -/*0058*/case GuiAlarmID::ALARM_ID_DD_BARO_SENSOR_FPGA_FAULT : { result = QObject::tr("DD baro sensors freshness timeout fault." ); break; } /* 58*/ +/*0058*/case GuiAlarmID::ALARM_ID_AVAILABLE_58 : { result = QObject::tr("DD Available alarms." ); break; } /* 58*/ /*0059*/case GuiAlarmID::ALARM_ID_DD_BAROMETRIC_SENSOR_COEFFS_BAD_CRC : { result = QObject::tr("DD barometric sensor coefficients bad CRC." ); break; } /* 59*/ /*0060*/case GuiAlarmID::ALARM_ID_DD_D12_PUMP_RPM_OUT_OF_RANGE : { result = QObject::tr("DD fresh dialysate pump RPM out of range." ); break; } /* 60*/ /*0061*/case GuiAlarmID::ALARM_ID_DD_D12_PUMP_OFF_FAULT : { result = QObject::tr("DD fresh dialysate pump running when it is not commanded to run." ); break; } /* 61*/ @@ -105,8 +105,8 @@ /*0069*/case GuiAlarmID::ALARM_ID_DD_D48_PUMP_DIRECTION_FPGA_FAULT : { result = QObject::tr("DD spent dialysate pump direction error count FPGA fault." ); break; } /* 69*/ /*0070*/case GuiAlarmID::ALARM_ID_DD_FLUID_TOO_LOW_WHILE_D5_HEAT_IS_ON : { result = QObject::tr("DD primary heater is On when no fluid is present." ); break; } /* 70*/ /*0071*/case GuiAlarmID::ALARM_ID_DD_FLUID_TOO_LOW_WHILE_D45_HEAT_IS_ON : { result = QObject::tr("DD trimmer heater is On when no fluid is present." ); break; } /* 71*/ -/*0072*/case GuiAlarmID::ALARM_ID_DD_WATER_INLET_INPUT_PRES_TIMEOUT_FAULT : { result = QObject::tr("DD water inlet input pressure sensor freshness/error timeout fault." ); break; } /* 72*/ -/*0073*/case GuiAlarmID::ALARM_ID_DD_WATER_INLET_OUTPUT_PRES_TIMEOUT_FAULT : { result = QObject::tr("DD water inlet output pressure sensor freshness/error timeout fault." ); break; } /* 73*/ +/*0072*/case GuiAlarmID::ALARM_ID_AVAILABLE_72 : { result = QObject::tr("Available alarms." ); break; } /* 72*/ +/*0073*/case GuiAlarmID::ALARM_ID_AVAILABLE_73 : { result = QObject::tr("Available alarms." ); break; } /* 73*/ /*0074*/case GuiAlarmID::ALARM_ID_DD_D76_PUMP_SPEED_CONTROL_ERROR : { result = QObject::tr("DD UF pump speed control error." ); break; } /* 74*/ /*0075*/case GuiAlarmID::ALARM_ID_DD_SC_STATE1_FILL_START_PRESSURE_OUT_OF_RANGE: { result = QObject::tr("DD spent chamber state 1 fill pressure out of range." ); break; } /* 75*/ /*0076*/case GuiAlarmID::ALARM_ID_DD_SC_STATE2_FILL_START_PRESSURE_OUT_OF_RANGE: { result = QObject::tr("DD spent chamber state 2 fill pressure out of range." ); break; } /* 76*/