Index: AlarmDefs.h =================================================================== diff -u -r04b3024e6888c3f3fe6ebfcc594d02287cb22c60 -rbf2726e707c57ae7acd604d1c108a598be5d5c37 --- AlarmDefs.h (.../AlarmDefs.h) (revision 04b3024e6888c3f3fe6ebfcc594d02287cb22c60) +++ AlarmDefs.h (.../AlarmDefs.h) (revision bf2726e707c57ae7acd604d1c108a598be5d5c37) @@ -7,8 +7,8 @@ * * @file AlarmDefs.h * -* @author (last) Sameer Kalliadan Poyil -* @date (last) 02-Mar-2026 +* @author (last) Varshini Nagabooshanam +* @date (last) 13-Mar-2026 * * @author (original) Sean Nash * @date (original) 01-Aug-2024 @@ -58,7 +58,7 @@ ALARM_ID_TD_VENOUS_SENSOR_TIMEOUT_FAULT = 27, ///< TD venous pressure sensor freshness/error timeout fault ALARM_ID_TD_TMP_PRESSURE_LOW = 28, ///< TD TMP pressure low ALARM_ID_TD_TMP_PRESSURE_HIGH = 29, ///< TD TMP pressure high - ALARM_ID_AVAILABLE_30 = 30, ///< Available alarms + ALARM_ID_TD_VENOUS_BUBBLE_SELF_TEST_FAILURE = 30, ///< TD venous bubble self-test failure. ALARM_ID_AVAILABLE_31 = 31, ///< Available alarms ALARM_ID_AVAILABLE_32 = 32, ///< Available alarms ALARM_ID_TD_VOLTAGE_OUT_OF_RANGE = 33, ///< TD voltage out of range fault @@ -86,7 +86,7 @@ ALARM_ID_DD_CONCENTRATE_PUMP_FAULT = 55, ///< DD concentrate pump fault ALARM_ID_DD_TEMPERATURE_SENSOR_OUT_OF_RANGE = 56, ///< DD temperature sensor out of range ALARM_ID_DD_RTD_SENSORS_FPGA_FAULT = 57, ///< DD temperature sensors freshness timeout fault - ALARM_ID_AVAILABLE_58 = 58, ///< DD Available alarms + ALARM_ID_DD_BC_FILL_TIMEOUT_FAULT = 58, ///< DD balancing chamber fill timeout fault ALARM_ID_DD_BAROMETRIC_SENSOR_COEFFS_BAD_CRC = 59, ///< DD barometric sensor coefficients bad CRC ALARM_ID_DD_D12_PUMP_RPM_OUT_OF_RANGE = 60, ///< DD fresh dialysate pump RPM out of range ALARM_ID_DD_D12_PUMP_OFF_FAULT = 61, ///< DD fresh dialysate pump running when it is not commanded to run @@ -362,9 +362,9 @@ // Priority Rank Source Fault DDFault Stops ClrIm NoClr NoRes NoRin NoEnd BlkRB BlkET NoBRcr NoDRcr AutoRes ClrOnly TxLog ID { ALARM_PRIORITY_HIGH, 10, ALM_SRC_TD, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_TD_ARTERIAL_SENSOR_TIMEOUT_FAULT }, { ALARM_PRIORITY_HIGH, 10, ALM_SRC_TD, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_TD_VENOUS_SENSOR_TIMEOUT_FAULT }, - { ALARM_PRIORITY_HIGH, 204, ALM_SRC_TD, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, TRUE, ALARM_ID_TD_TMP_PRESSURE_LOW }, - { ALARM_PRIORITY_HIGH, 205, ALM_SRC_TD, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, TRUE, ALARM_ID_TD_TMP_PRESSURE_HIGH }, - { ALARM_PRIORITY_LOW, 999, ALM_SRC_TD, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_AVAILABLE_30 }, + { ALARM_PRIORITY_HIGH, 204, ALM_SRC_TD, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, TRUE , ALARM_ID_TD_TMP_PRESSURE_LOW }, + { ALARM_PRIORITY_HIGH, 205, ALM_SRC_TD, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, TRUE , ALARM_ID_TD_TMP_PRESSURE_HIGH }, + { ALARM_PRIORITY_LOW, 711, ALM_SRC_TD, FALSE, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_TD_VENOUS_BUBBLE_SELF_TEST_FAILURE }, { ALARM_PRIORITY_LOW, 999, ALM_SRC_TD, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_AVAILABLE_31 }, { ALARM_PRIORITY_LOW, 999, ALM_SRC_TD, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_AVAILABLE_32 }, { ALARM_PRIORITY_HIGH, 10, ALM_SRC_TD, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_TD_VENOUS_SENSOR_TIMEOUT_FAULT }, @@ -393,7 +393,7 @@ { ALARM_PRIORITY_HIGH, 110, ALM_SRC_DD, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_CONCENTRATE_PUMP_FAULT }, { ALARM_PRIORITY_HIGH, 110, ALM_SRC_DD, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_TEMPERATURE_SENSOR_OUT_OF_RANGE }, { ALARM_PRIORITY_HIGH, 110, ALM_SRC_DD, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_RTD_SENSORS_FPGA_FAULT }, - { ALARM_PRIORITY_LOW, 999, ALM_SRC_TD, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_AVAILABLE_58 }, + { ALARM_PRIORITY_HIGH, 110, ALM_SRC_DD, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_BC_FILL_TIMEOUT_FAULT }, { ALARM_PRIORITY_HIGH, 110, ALM_SRC_DD, FALSE, TRUE, TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_BAROMETRIC_SENSOR_COEFFS_BAD_CRC }, { ALARM_PRIORITY_HIGH, 110, ALM_SRC_DD, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_D12_PUMP_RPM_OUT_OF_RANGE }, { ALARM_PRIORITY_HIGH, 110, ALM_SRC_DD, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_D12_PUMP_OFF_FAULT }, @@ -574,6 +574,7 @@ { 110, ALARM_ID_DD_D27_COND_SENSOR_FPGA_FAULT }, { 110, ALARM_ID_DD_D29_COND_SENSOR_FPGA_FAULT }, { 110, ALARM_ID_DD_D43_COND_SENSOR_FPGA_FAULT }, + { 110, ALARM_ID_DD_BC_FILL_TIMEOUT_FAULT }, { 110, ALARM_ID_FP_FPGA_COMM_TIMEOUT }, { 110, ALARM_ID_DD_D11_PUMP_SPEED_CONTROL_ERROR }, { 110, ALARM_ID_DD_D10_PUMP_SPEED_CONTROL_ERROR }, @@ -654,12 +655,11 @@ { 700, ALARM_ID_TD_ARTERIAL_PRESSURE_SELF_TEST_FAILURE }, { 700, ALARM_ID_TD_VENOUS_PRESSURE_SELF_TEST_FAILURE }, { 710, ALARM_ID_TD_PRE_TREATMENT_DRY_PRESSURE_TEST_FAILURE }, + { 711, ALARM_ID_TD_VENOUS_BUBBLE_SELF_TEST_FAILURE }, { 850, ALARM_ID_BATTERY_LOW }, { 852, ALARM_ID_TD_AC_POWER_LOST }, - { 999, ALARM_ID_AVAILABLE_30 }, { 999, ALARM_ID_AVAILABLE_31 }, { 999, ALARM_ID_AVAILABLE_32 }, - { 999, ALARM_ID_AVAILABLE_58 }, { 999, ALARM_ID_UI_RESERVED_116 }, { 999, ALARM_ID_UI_RESERVED_117 }, { 999, ALARM_ID_UI_RESERVED_118 }, @@ -729,7 +729,7 @@ { 27 , ALARM_ID_TD_VENOUS_SENSOR_TIMEOUT_FAULT , "TD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | TD Venous Pressure Sensor Error | TD venous pressure sensor reporting error or not being read successfully | Venous pressure sensor is reporting an error or FPGA is failing to read sensor. | "}, { 28 , ALARM_ID_TD_TMP_PRESSURE_LOW , "TD | Low TMP Pressure | TMP pressure is low | | | | "}, { 29 , ALARM_ID_TD_TMP_PRESSURE_HIGH , "TD | High TMP Pressure | TMP pressure is high | | | | "}, - { 30 , ALARM_ID_AVAILABLE_30 , " | | | | | | "}, + { 30 , ALARM_ID_TD_VENOUS_BUBBLE_SELF_TEST_FAILURE , "TD | Self Test: Venous Air Detector | A problem has been detected with the venous air detector.\n- Check cartridge installation.\n- If problem persists, terminate treatment preparation and \n begin again using a new cartridge. | Process: Pre- Tx Wet Self- Test | TD venous air bubble detector self-test failure | If the venous air bubble self test did not detect bubble for more than 500 msec | "}, { 31 , ALARM_ID_AVAILABLE_31 , " | | | | | | "}, { 32 , ALARM_ID_AVAILABLE_32 , " | | | | | | "}, { 33 , ALARM_ID_TD_VOLTAGE_OUT_OF_RANGE , "TD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | TD Fault: Voltage Monitor | TD monitored voltage is out of range | If the TD monitored voltages are out of range for a certain period of time. | "}, @@ -758,7 +758,7 @@ { 55 , ALARM_ID_DD_CONCENTRATE_PUMP_FAULT , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: Pump Fault | DD concentrate pump fault | If the concentrate pumps fault is reported by the FPGA. | "}, { 56 , ALARM_ID_DD_TEMPERATURE_SENSOR_OUT_OF_RANGE , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: Temp Sensor Range | DD temperature sensor out of range | If the measured temperature < 0 C or > 80 C for the non-fluid inline sensors and > 120 C for the fluid sensors. | "}, { 57 , ALARM_ID_DD_RTD_SENSORS_FPGA_FAULT , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: Temperature RTD Comm | DD temperature sensors not being read successfully | If the FPGA does not report fresh data for a certain period of time | "}, - { 58 , ALARM_ID_AVAILABLE_58 , " | | | | | | "}, + { 58 , ALARM_ID_DD_BC_FILL_TIMEOUT_FAULT , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD BC fill timeout Error | DD Balancing chamber fill timeout error | If BC fill time exceeds the allowed timeout based on the balancing chamber switching period. | "}, { 59 , ALARM_ID_DD_BAROMETRIC_SENSOR_COEFFS_BAD_CRC , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: Baro temp Cal CRC | DD barometric pressure sensor bad coefficients CRC | If the following happens: 1. The stored CRC in the sensor does not match the calculated CRC. 2. The CRC and coefficients have not been received from the sensor after 20 seconds from the power up. | "}, { 60 , ALARM_ID_DD_D12_PUMP_RPM_OUT_OF_RANGE , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: Fresh Dial Pump Speed Fault | DD fresh dialysate pump RPM out of range | If the fresh dialysate pump RPM > 300.0 of the target RPM. | "}, { 61 , ALARM_ID_DD_D12_PUMP_OFF_FAULT , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: Fresh Dial Pump Speed Not Off | DD fresh dialysate pump off fault | If the pump is in off state and the RPM is greater than the minimum RPM of 300.0 RPM. | "},