Index: AlarmDefs.h =================================================================== diff -u -r1d7224948bc9a9f1393e846166d832c4efebcebe -rd1d088e2daa86f02ce1adee7c837e72e94e548f1 --- AlarmDefs.h (.../AlarmDefs.h) (revision 1d7224948bc9a9f1393e846166d832c4efebcebe) +++ AlarmDefs.h (.../AlarmDefs.h) (revision d1d088e2daa86f02ce1adee7c837e72e94e548f1) @@ -119,7 +119,7 @@ ALARM_ID_DD_BC_STATE1_FILL_END_PRESSURE_OUT_OF_RANGE = 88, ///< DD balance chamber state1 fill end pressure not in range ALARM_ID_DD_BC_STATE2_FILL_START_PRESSURE_OUT_OF_RANGE = 89, ///< DD balance chamber state2 fill start pressure not in range ALARM_ID_DD_BC_STATE2_FILL_END_PRESSURE_OUT_OF_RANGE = 90, ///< DD balance chamber state2 fill end pressure not in range - ALARM_ID_AVAILABLE_91 = 91, ///< Available alarms + ALARM_ID_DD_GEND_HYD_CHAMBER_PRESS_OUT_OF_RANGE = 91, ///< DD generate dialysate hydraulics chamber ngative pressure out of range ALARM_ID_AVAILABLE_92 = 92, ///< Available alarms ALARM_ID_AVAILABLE_93 = 93, ///< Available alarms ALARM_ID_DD_BC_STATE1_FILL_PRESSURE_DROP_OUT_OF_RANGE = 94, ///< DD balance chamber state1 fill initiated and pressure drop is not in range @@ -391,7 +391,7 @@ { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DD_BC_STATE1_FILL_END_PRESSURE_OUT_OF_RANGE }, { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DD_BC_STATE2_FILL_START_PRESSURE_OUT_OF_RANGE }, { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DD_BC_STATE2_FILL_END_PRESSURE_OUT_OF_RANGE }, - { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_AVAILABLE_91 }, + { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DD_GEND_HYD_CHAMBER_PRESS_OUT_OF_RANGE }, { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_AVAILABLE_92 }, { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_AVAILABLE_93 }, { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DD_BC_STATE1_FILL_PRESSURE_DROP_OUT_OF_RANGE }, @@ -527,6 +527,7 @@ { 110, ALARM_ID_DD_BC_STATE1_FILL_END_PRESSURE_OUT_OF_RANGE }, { 110, ALARM_ID_DD_BC_STATE2_FILL_START_PRESSURE_OUT_OF_RANGE }, { 110, ALARM_ID_DD_BC_STATE2_FILL_END_PRESSURE_OUT_OF_RANGE }, + { 110, ALARM_ID_DD_GEND_HYD_CHAMBER_PRESS_OUT_OF_RANGE }, { 110, ALARM_ID_DD_BC_STATE1_FILL_PRESSURE_DROP_OUT_OF_RANGE }, { 110, ALARM_ID_DD_BC_STATE2_FILL_PRESSURE_DROP_OUT_OF_RANGE }, { 110, ALARM_ID_DD_D74_COND_SENSOR_FPGA_FAULT }, @@ -550,26 +551,25 @@ { 710, ALARM_ID_TD_PRE_TREATMENT_DRY_PRESSURE_TEST_FAILURE }, { 850, ALARM_ID_BATTERY_LOW }, { 852, ALARM_ID_TD_AC_POWER_LOST }, - { 999, ALARM_ID_AVAILABLE_91 }, - { 999, ALARM_ID_AVAILABLE_92 }, - { 999, ALARM_ID_AVAILABLE_93 }, - { 999, ALARM_ID_AVAILABLE_98 }, - { 999, ALARM_ID_AVAILABLE_99 }, - { 999, ALARM_ID_UI_RESERVED_116 }, - { 999, ALARM_ID_UI_RESERVED_117 }, - { 999, ALARM_ID_UI_RESERVED_118 }, - { 999, ALARM_ID_UI_RESERVED_119 }, - { 999, ALARM_ID_UI_RESERVED_120 }, - { 999, ALARM_ID_UI_RESERVED_121 }, - { 999, ALARM_ID_UI_RESERVED_122 }, - { 999, ALARM_ID_UI_RESERVED_123 }, - { 999, ALARM_ID_UI_RESERVED_124 }, - { 999, ALARM_ID_UI_RESERVED_125 }, - { 999, ALARM_ID_UI_RESERVED_126 }, - { 999, ALARM_ID_UI_RESERVED_127 }, - { 999, ALARM_ID_UI_RESERVED_128 }, - { 999, ALARM_ID_UI_RESERVED_129 }, - { 999, ALARM_ID_UI_RESERVED_130 }, + { 999, ALARM_ID_AVAILABLE_92 }, + { 999, ALARM_ID_AVAILABLE_93 }, + { 999, ALARM_ID_AVAILABLE_98 }, + { 999, ALARM_ID_AVAILABLE_99 }, + { 999, ALARM_ID_UI_RESERVED_116 }, + { 999, ALARM_ID_UI_RESERVED_117 }, + { 999, ALARM_ID_UI_RESERVED_118 }, + { 999, ALARM_ID_UI_RESERVED_119 }, + { 999, ALARM_ID_UI_RESERVED_120 }, + { 999, ALARM_ID_UI_RESERVED_121 }, + { 999, ALARM_ID_UI_RESERVED_122 }, + { 999, ALARM_ID_UI_RESERVED_123 }, + { 999, ALARM_ID_UI_RESERVED_124 }, + { 999, ALARM_ID_UI_RESERVED_125 }, + { 999, ALARM_ID_UI_RESERVED_126 }, + { 999, ALARM_ID_UI_RESERVED_127 }, + { 999, ALARM_ID_UI_RESERVED_128 }, + { 999, ALARM_ID_UI_RESERVED_129 }, + { 999, ALARM_ID_UI_RESERVED_130 }, { 999, ALARM_ID_NO_ALARM }, }; @@ -687,7 +687,7 @@ { 88 , ALARM_ID_DD_BC_STATE1_FILL_END_PRESSURE_OUT_OF_RANGE , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: BC Pressure not in range | DD balancing chamber state 1 fill end pressure is not in range | If the fresh dialysate pressure (< 21 psig or > 27 psig) and spent dialysate pressure (< 27 psig or > 31 psig) not in range for balancing chamber state1 fill end. | "}, { 89 , ALARM_ID_DD_BC_STATE2_FILL_START_PRESSURE_OUT_OF_RANGE , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: BC Pressure not in range | DD balancing chamber state 2 fill start pressure is not in range | If the fresh dialysate pressure (< 21 psig or > 27 psig) and spent dialysate pressure (< 27 psig or > 31 psig) not in range for balancing chamber state2 fill start. | "}, { 90 , ALARM_ID_DD_BC_STATE2_FILL_END_PRESSURE_OUT_OF_RANGE , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: BC Pressure not in range | DD balancing chamber state 2 fill end pressure is not in range | If the fresh dialysate pressure (< 21 psig or > 27 psig) and spent dialysate pressure (< 27 psig or > 31 psig) not in range for balancing chamber state2 fill end. | "}, - { 91 , ALARM_ID_AVAILABLE_91 , " | | | | | | "}, + { 91 , ALARM_ID_DD_GEND_HYD_CHAMBER_PRESS_OUT_OF_RANGE , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: Hyd Cham Neg Pres not in range | DD hydraulics chamber degassing pressure is not in range | If the hydraulics chamber pressure is not less than or equal to calculated target degas pressure psig. | "} { 92 , ALARM_ID_AVAILABLE_92 , " | | | | | | "}, { 93 , ALARM_ID_AVAILABLE_93 , " | | | | | | "}, { 94 , ALARM_ID_DD_BC_STATE1_FILL_PRESSURE_DROP_OUT_OF_RANGE , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: BC Pressure not in range | DD balancing chamber state 1 fill initiated and pressure drop is not in range | If the fresh dialysate pressure (< 9 psig or > 13 psig) and spent dialysate pressure (< 9 psig or > 13 psig) not in range for balancing chamber state1 fill in progress. | "},