Index: AlarmDefs.h =================================================================== diff -u -r3b43acb32a0253d05fd98e65cdcb5e7fb7a41870 -rf09251af9e269f3db48265dc5d6e4b9215c9f1f2 --- AlarmDefs.h (.../AlarmDefs.h) (revision 3b43acb32a0253d05fd98e65cdcb5e7fb7a41870) +++ AlarmDefs.h (.../AlarmDefs.h) (revision f09251af9e269f3db48265dc5d6e4b9215c9f1f2) @@ -7,8 +7,8 @@ * * @file AlarmDefs.h * -* @author (last) Praneeth Bunne -* @date (last) 04-Jun-2026 +* @author (last) Varshini Nagabooshanam +* @date (last) 18-Jun-2026 * * @author (original) Sean Nash * @date (original) 01-Aug-2024 @@ -226,7 +226,10 @@ ALARM_ID_TD_INTEGRITY_POST_TEST_FAILED = 194, ///< TD integrity post test fault ALARM_ID_DD_INTEGRITY_POST_TEST_FAILED = 195, ///< TD integrity post test fault ALARM_ID_TD_EMPTY_SALINE_BAG = 196, ///< TD empty saline bag alarm - ALARM_ID_TD_TUBE_SET_MISMATCH = 197, ///< TD tubeset mismatch alarm + ALARM_ID_DD_D92_PUMP_RPM_OUT_OF_RANGE = 197, ///< DD D92 pump rpm out of range + ALARM_ID_TD_PRIME_COMPLETED_MEDIUM = 198, ///< TD prime completed medium priority alarm + ALARM_ID_TD_REPEATED_AIR_TRAP_LOWER_EVENTS = 199, ///< TD repeated air trap lower events + ALARM_ID_TD_TUBE_SET_MISMATCH = 200, ///< TD tubeset mismatch alarm NUM_OF_ALARM_IDS ///< Number of alarms }; typedef enum Alarm_List ALARM_ID_T; ///< Type for alarm list enumeration @@ -571,6 +574,9 @@ { ALARM_PRIORITY_HIGH, 1, ALM_SRC_TD, TRUE , FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_TD_INTEGRITY_POST_TEST_FAILED }, { ALARM_PRIORITY_HIGH, 1, ALM_SRC_DD, FALSE, FALSE, TRUE , TRUE , TRUE , TRUE , TRUE , TRUE , FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_INTEGRITY_POST_TEST_FAILED }, { ALARM_PRIORITY_HIGH, 202, ALM_SRC_TD, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, ALARM_ID_TD_EMPTY_SALINE_BAG }, + { ALARM_PRIORITY_HIGH, 110, ALM_SRC_DD, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, ALARM_ID_DD_D92_PUMP_RPM_OUT_OF_RANGE }, + { ALARM_PRIORITY_MEDIUM, 590, ALM_SRC_TD, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, ALARM_ID_TD_PRIME_COMPLETED_MEDIUM }, + { ALARM_PRIORITY_HIGH, 17, ALM_SRC_TD, FALSE, FALSE, TRUE , TRUE , FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, TRUE , ALARM_ID_TD_REPEATED_AIR_TRAP_LOWER_EVENTS }, { ALARM_PRIORITY_MEDIUM, 300, ALM_SRC_TD, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, ALARM_ID_TD_TUBE_SET_MISMATCH }, }; // Priority Rank Source Fault DDFault Stops ClrIm NoClr NoRes NoRin NoEnd BlkRB BlkET NoBRcr NoDRcr AutoRes ClrOnly TxLog ID @@ -648,6 +654,7 @@ { 12, ALARM_ID_TD_VENOUS_BUBBLE_DETECTED }, { 13, ALARM_ID_TD_AIR_TRAP_FILL_TIMEOUT }, { 13, ALARM_ID_TD_AIR_TRAP_LOWER_TIMEOUT }, + { 17, ALARM_ID_TD_REPEATED_AIR_TRAP_LOWER_EVENTS }, { 100, ALARM_ID_TD_AC_POWER_LOST_IN_TREATMENT }, { 110, ALARM_ID_DD_CRITICAL_DATA_ERROR }, { 110, ALARM_ID_DD_WATCHDOG_EXPIRED }, @@ -683,6 +690,7 @@ { 110, ALARM_ID_DD_D48_PUMP_CURRENT_OUT_OF_RANGE }, { 110, ALARM_ID_DD_D48_PUMP_DIRECTION_INVALID }, { 110, ALARM_ID_DD_D48_PUMP_DIRECTION_FPGA_FAULT }, + { 110, ALARM_ID_DD_D92_PUMP_RPM_OUT_OF_RANGE }, { 110, ALARM_ID_DD_FLUID_TOO_LOW_WHILE_D5_HEAT_IS_ON }, { 110, ALARM_ID_DD_FLUID_TOO_LOW_WHILE_D45_HEAT_IS_ON }, { 110, ALARM_ID_FP_COMM_TOO_MANY_BAD_CRCS }, @@ -746,6 +754,7 @@ { 435, ALARM_ID_FP_INLET_PRESSURE_OUT_HIGH_RANGE }, { 436, ALARM_ID_FP_INLET_CONDUCTIVITY_OUT_HIGH_RANGE }, { 437, ALARM_ID_FP_INLET_CONDUCTIVITY_OUT_LOW_RANGE }, + { 590, ALARM_ID_TD_PRIME_COMPLETED_MEDIUM }, { 698, ALARM_ID_DD_BLOOD_LEAK_RECOVERING_PLEASE_WAIT }, { 700, ALARM_ID_TD_ARTERIAL_PRESSURE_SELF_TEST_FAILURE }, { 700, ALARM_ID_TD_VENOUS_PRESSURE_SELF_TEST_FAILURE }, @@ -1001,7 +1010,10 @@ { 194 , ALARM_ID_TD_INTEGRITY_POST_TEST_FAILED , "TD | Service Required: Hemodialysis Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | TD Firmware Image Integrity Failure | TD firmware image integrity POST test failed | If the calculated CRC of the TD image does not match the expected CRC. | "}, { 195 , ALARM_ID_DD_INTEGRITY_POST_TEST_FAILED , "DD | Service Required: Hemodialysis Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Firmware Image Integrity Failure | DD firmware image integrity POST test failed | If the calculated CRC of the DD image does not match the expected CRC. | "}, { 196 , ALARM_ID_TD_EMPTY_SALINE_BAG , "TD | Blood Pump Stopped: Saline Bag Empty | The saline bag appears empty.\n- Check the saline bag and replace if empty. | Blood Stop: Saline Bag Empty | TD Saline bag is empty | If the measured arterial pressure < -300.0 mmHg for more than 250 msec. | "}, - { 197, ALARM_ID_TD_TUBE_SET_MISMATCH , "TD | Wrong Tube Set Installed | The installed tube set does not match the selected treatment type. \n- Remove the current tube set.\n- Install the correct tube set for the selected treatment:\n HD or HD with Online Fluid requires an HD tube set.\n HDF requires an HDF tube set. | TD Tube Set Mismatch | TD installed tube set type does not match the selected treatment modality | If the scanned tube set type does not match the tube set type required for the selected treatment modality. | "}, + { 197 , ALARM_ID_DD_D92_PUMP_RPM_OUT_OF_RANGE , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: Subtitution Pump Fault | DD substitution pump RPM out of range | If the substitution pump RPM > 20.0 of the target RPM. | "}, + { 198 , ALARM_ID_TD_PRIME_COMPLETED_MEDIUM , "TD | Prime Complete - 30 MIN WARNING | The priming process completed 90 minutes ago. \n- Treatment must begin within 30 minutes or the\npre-treatment process will be terminated.\n- Press Resume/OK to continue. | Process: Pre-Tx Prime Complete | TD prime completed medium priority alarm | In Pre-Treatment Recirculation for 90 minutes | "}, + { 199 , ALARM_ID_TD_REPEATED_AIR_TRAP_LOWER_EVENTS , "TD | Blood Pump Stopped: Venous Air Detector | The air trap level required repeated lowering events. \n- Confirm blood tubing connections are secure.\n- Inspect the air trap for abnormal fluid levels. \n- Resume treatment if the condition has been corrected. | Blood Stop: Repeated Air Trap Lower | Repeated air trap lower events detected during treatment | If 5 air trap lower events occur within 5 minutes during treatment. | "}, + { 200 , ALARM_ID_TD_TUBE_SET_MISMATCH , "TD | Wrong Tube Set Installed | The installed tube set does not match the selected treatment type. \n- Remove the current tube set.\n- Install the correct tube set for the selected treatment:\n HD or HD with Online Fluid requires an HD tube set.\n HDF requires an HDF tube set. | TD Tube Set Mismatch | TD installed tube set type does not match the selected treatment modality | If the scanned tube set type does not match the tube set type required for the selected treatment modality. | "}, }; #endif // manual alarms