Index: AlarmDefs.h =================================================================== diff -u -r4709c4af2b480b2e4d7e68dc30df1f509837849e -raebe9aa2b6e24169026b4d5d9ea08d6718a7b5fc --- AlarmDefs.h (.../AlarmDefs.h) (revision 4709c4af2b480b2e4d7e68dc30df1f509837849e) +++ AlarmDefs.h (.../AlarmDefs.h) (revision aebe9aa2b6e24169026b4d5d9ea08d6718a7b5fc) @@ -119,9 +119,9 @@ ALARM_ID_DD_BC_STATE1_FILL_END_PRESSURE_OUT_OF_RANGE = 88, ///< DD balance chamber state1 fill end pressure not in range ALARM_ID_DD_BC_STATE2_FILL_START_PRESSURE_OUT_OF_RANGE = 89, ///< DD balance chamber state2 fill start pressure not in range ALARM_ID_DD_BC_STATE2_FILL_END_PRESSURE_OUT_OF_RANGE = 90, ///< DD balance chamber state2 fill end pressure not in range - ALARM_ID_DD_GEND_HYD_CHAMBER_PRESS_OUT_OF_RANGE = 91, ///< DD generate dialysate hydraulics chamber pressure out of range - ALARM_ID_DD_GEND_FRESH_DIAL_PRESS_OUT_OF_RANGE = 92, ///< DD generate dialysate fresh dialysate pressure out of range - ALARM_ID_DD_GEND_SPENT_DIAL_PRESS_OUT_OF_RANGE = 93, ///< DD generate dialysate spent dialysate pressure out of range + ALARM_ID_AVAILABLE_91 = 91, ///< Available alarms + ALARM_ID_AVAILABLE_92 = 92, ///< Available alarms + ALARM_ID_AVAILABLE_93 = 93, ///< Available alarms ALARM_ID_DD_BC_STATE1_FILL_PRESSURE_DROP_OUT_OF_RANGE = 94, ///< DD balance chamber state1 fill initiated and pressure drop is not in range ALARM_ID_DD_BC_STATE2_FILL_PRESSURE_DROP_OUT_OF_RANGE = 95, ///< DD balance chamber state2 fill initiated and pressure drop is not in range ALARM_ID_DD_D74_COND_SENSOR_FPGA_FAULT = 96, ///< DD D74 bicarb conductivity sensor freshness/error timeout fault = 96, ///< Available alarms @@ -386,9 +386,9 @@ { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DD_BC_STATE1_FILL_END_PRESSURE_OUT_OF_RANGE }, { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DD_BC_STATE2_FILL_START_PRESSURE_OUT_OF_RANGE }, { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DD_BC_STATE2_FILL_END_PRESSURE_OUT_OF_RANGE }, - { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DD_GEND_HYD_CHAMBER_PRESS_OUT_OF_RANGE }, - { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DD_GEND_FRESH_DIAL_PRESS_OUT_OF_RANGE }, - { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DD_GEND_SPENT_DIAL_PRESS_OUT_OF_RANGE }, + { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_AVAILABLE_91 }, + { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_AVAILABLE_92 }, + { ALARM_PRIORITY_LOW, 999, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , TRUE , FALSE, FALSE, ALARM_ID_AVAILABLE_93 }, { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DD_BC_STATE1_FILL_PRESSURE_DROP_OUT_OF_RANGE }, { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DD_BC_STATE2_FILL_PRESSURE_DROP_OUT_OF_RANGE }, { ALARM_PRIORITY_HIGH, 110, FALSE, TRUE , TRUE , TRUE , FALSE, TRUE , FALSE, FALSE, FALSE, FALSE, FALSE, TRUE , FALSE, FALSE, ALARM_ID_DD_D74_COND_SENSOR_FPGA_FAULT }, @@ -519,9 +519,6 @@ { 110, ALARM_ID_DD_BC_STATE1_FILL_END_PRESSURE_OUT_OF_RANGE }, { 110, ALARM_ID_DD_BC_STATE2_FILL_START_PRESSURE_OUT_OF_RANGE }, { 110, ALARM_ID_DD_BC_STATE2_FILL_END_PRESSURE_OUT_OF_RANGE }, - { 110, ALARM_ID_DD_GEND_HYD_CHAMBER_PRESS_OUT_OF_RANGE }, - { 110, ALARM_ID_DD_GEND_FRESH_DIAL_PRESS_OUT_OF_RANGE }, - { 110, ALARM_ID_DD_GEND_SPENT_DIAL_PRESS_OUT_OF_RANGE }, { 110, ALARM_ID_DD_BC_STATE1_FILL_PRESSURE_DROP_OUT_OF_RANGE }, { 110, ALARM_ID_DD_BC_STATE2_FILL_PRESSURE_DROP_OUT_OF_RANGE }, { 110, ALARM_ID_DD_D74_COND_SENSOR_FPGA_FAULT }, @@ -543,6 +540,9 @@ { 700, ALARM_ID_TD_VENOUS_PRESSURE_SELF_TEST_FAILURE }, { 710, ALARM_ID_TD_PRE_TREATMENT_DRY_PRESSURE_TEST_FAILURE }, { 852, ALARM_ID_TD_AC_POWER_LOST }, + { 999, ALARM_ID_AVAILABLE_91 }, + { 999, ALARM_ID_AVAILABLE_92 }, + { 999, ALARM_ID_AVAILABLE_93 }, { 999, ALARM_ID_AVAILABLE_97 }, { 999, ALARM_ID_AVAILABLE_98 }, { 999, ALARM_ID_AVAILABLE_99 }, @@ -678,9 +678,9 @@ { 88 , ALARM_ID_DD_BC_STATE1_FILL_END_PRESSURE_OUT_OF_RANGE , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: BC Pressure not in range | DD balancing chamber state 1 fill end pressure is not in range | If the fresh dialysate pressure (< 21 psig or > 27 psig) and spent dialysate pressure (< 27 psig or > 31 psig) not in range for balancing chamber state1 fill end. | "}, { 89 , ALARM_ID_DD_BC_STATE2_FILL_START_PRESSURE_OUT_OF_RANGE , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: BC Pressure not in range | DD balancing chamber state 2 fill start pressure is not in range | If the fresh dialysate pressure (< 21 psig or > 27 psig) and spent dialysate pressure (< 27 psig or > 31 psig) not in range for balancing chamber state2 fill start. | "}, { 90 , ALARM_ID_DD_BC_STATE2_FILL_END_PRESSURE_OUT_OF_RANGE , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: BC Pressure not in range | DD balancing chamber state 2 fill end pressure is not in range | If the fresh dialysate pressure (< 21 psig or > 27 psig) and spent dialysate pressure (< 27 psig or > 31 psig) not in range for balancing chamber state2 fill end. | "}, - { 91 , ALARM_ID_DD_GEND_HYD_CHAMBER_PRESS_OUT_OF_RANGE , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: Hyd Cham Press not in range | DD hydraulics chamber degassed pressure is not in range | If the hydraulics chamber pressure is not between -11.7877 psig and -12.2789 psig. | "}, - { 92 , ALARM_ID_DD_GEND_FRESH_DIAL_PRESS_OUT_OF_RANGE , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: Fresh dial Press not in range | DD fresh dialysate pressure is not in range | If the fresh dialysate pressure (< 21 psig or > 27 psig) not in range. | "}, - { 93 , ALARM_ID_DD_GEND_SPENT_DIAL_PRESS_OUT_OF_RANGE , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: Spent dial Press not in range | DD spent dialysate pressure is not in range | If the spent dialysate pressure (< 27 psig or > 31 psig) not in range. | "}, + { 91 , ALARM_ID_AVAILABLE_91 , " | | | | | | "}, + { 92 , ALARM_ID_AVAILABLE_92 , " | | | | | | "}, + { 93 , ALARM_ID_AVAILABLE_93 , " | | | | | | "}, { 94 , ALARM_ID_DD_BC_STATE1_FILL_PRESSURE_DROP_OUT_OF_RANGE , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: BC Pressure not in range | DD balancing chamber state 1 fill initiated and pressure drop is not in range | If the fresh dialysate pressure (< 9 psig or > 13 psig) and spent dialysate pressure (< 9 psig or > 13 psig) not in range for balancing chamber state1 fill in progress. | "}, { 95 , ALARM_ID_DD_BC_STATE2_FILL_PRESSURE_DROP_OUT_OF_RANGE , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: BC Pressure not in range | DD balancing chamber state 2 fill initiated and pressure drop is not in range | If the fresh dialysate pressure (< 9 psig or > 13 psig) and spent dialysate pressure (< 9 psig or > 13 psig) not in range for balancing chamber state2 fill in progress. | "}, { 96 , ALARM_ID_DD_D74_COND_SENSOR_FPGA_FAULT , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Conductivity Sensor Error | DD conductivity sensor (D74) located before balancing chamber reporting error or not being read successfully | Conductvity Sensor (D74) is reporting an error or FPGA is failing to read sensor. | "},