Index: AlarmDefs.h =================================================================== diff -u -r57cf24a496d5c1e2987fde2a925957e15dd1b018 -r466c854c5ea54c2983ba4e6a209e32e98e7b594a --- AlarmDefs.h (.../AlarmDefs.h) (revision 57cf24a496d5c1e2987fde2a925957e15dd1b018) +++ AlarmDefs.h (.../AlarmDefs.h) (revision 466c854c5ea54c2983ba4e6a209e32e98e7b594a) @@ -811,6 +811,7 @@ { 147 , ALARM_ID_FP_PRESSURE_RELIEF_OUT_HIGH_RANGE , "FP | Service Required: FP Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | FP Fault: Pressure Relief Valve High | FP pressure relief is too High | If P17 is > 17 PSI | "}, { 148 , ALARM_ID_FP_INLET_PRESSURE_REGULATOR_OUT_OF_RANGE , "FP | Service Required: FP Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | FP Fault: Pressure regulator not in range| FP pressure regulator is out of bounds | If M3 is not within 30 +/- 2 PSI | "}, { 149 , ALARM_ID_FP_POWER_BUDGET_OUT_OF_RANGE , "FP | Service Required: FP Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | FP Fault: Power usage has exceeded budget| FP device is using too much power | If the device current is greater that its allowed percentage | "}, + // Num Alarm ID Device Display Title Instructions Alarm List Title Description Trigger Condition { 150 , ALARM_ID_FP_FILTER_HEALTH_PRESSURE_OUT_OF_RANGE , "FP | FP: Filter Health | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | FP Filters: Pressure difference too high | FP pressure difference between the filters is too high | If the difference between M3 and P8 is ≥ 10 PSI after 2 minutes | "}, { 151 , ALARM_ID_FP_PERMEATE_FLUSH_CONDUCTIVITY_TIMEOUT , "FP | FP: Filter Health | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | FP Filters: Permeate conductivity high | FP conductivity after flushing the system is too high | If P18 conductivity > 200 uS/cm after 120 seconds ± 150ms of flushing. | "}, { 152 , ALARM_ID_FP_CONCENTRATE_FLUSH_VOLUME_OUT_OF_RANGE , "FP | FP: Filter Health | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | FP Filters: Volume difference too high | FP volume difference between integrated flow and expected volume | If P7 is <90% of the target integrated volume ( 1L ) after 30 seconds ± 150ms of flushing. | "},