Index: AlarmDefs.h =================================================================== diff -u -r5f6f19f4bdd2abcfb405622426420b9e8596f954 -rd653d7fd2ce503b9264b4fef0405dc9a8dccda3e --- AlarmDefs.h (.../AlarmDefs.h) (revision 5f6f19f4bdd2abcfb405622426420b9e8596f954) +++ AlarmDefs.h (.../AlarmDefs.h) (revision d653d7fd2ce503b9264b4fef0405dc9a8dccda3e) @@ -509,10 +509,10 @@ { 33 , ALARM_ID_TD_VOLTAGE_OUT_OF_RANGE , "TD | Service Required: Hemodialysis Device | A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | TD Fault: Voltage Monitor | TD monitored voltage is out of range | If the TD monitored voltages are out of range for a certain period of time. | "}, { 34 , ALARM_ID_TD_AC_POWER_LOST , "TD | Power Loss | AC Power has been lost.\n- Check all electrical connections.\n- Press Resume/OK to return to welcome screen. | Loss of AC Power | TD AC power lost alarm | The TD Software shall trigger an alarm if an AC power loss is detected while the TD system is in standby, service, or treatment parameters mode. | "}, { 35 , ALARM_ID_TD_AC_POWER_LOST_IN_TREATMENT , "TD | Power Loss | Treatment must be terminated due to a power loss.\n- Follow clinic instructions on when to initiate manual rinseback.\n- Check all electrical connections.\n- If power returns within 5 minutes of outage, \n- rinseback will be allowed. | Blood Stop: Loss of AC Power | TD AC power lost alarm | The TD Software shall trigger an alarm if an AC power loss is detected while the TD system is in pre-treatment, treatment, or post-treatment mode. | "}, - { 36 , ALARM_ID_DD_D17_COND_SENSOR_FPGA_FAULT , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Conductivity Sensor (D17) Error | DD conductivity sensor located before balancing chamber reporting error or not being read successfully | Conductvity Sensor (D17) is reporting an error or FPGA is failing to read sensor. | "}, - { 37 , ALARM_ID_DD_D27_COND_SENSOR_FPGA_FAULT , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Conductivity Sensor (D27) Error | DD conductivity sensor located after balancing chamber reporting error or not being read successfully | Conductvity Sensor (D27) is reporting an error or FPGA is failing to read sensor. | "}, - { 38 , ALARM_ID_DD_D29_COND_SENSOR_FPGA_FAULT , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Conductivity Sensor (D29) Error | DD conductivity sensor located after balancing chamber reporting error or not being read successfully | Conductvity Sensor (D29) is reporting an error or FPGA is failing to read sensor. | "}, - { 39 , ALARM_ID_DD_D43_COND_SENSOR_FPGA_FAULT , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Conductivity Sensor (D43) Error | DD conductivity sensor located in spent dialysate path reporting error or not being read successfully | Conductvity Sensor (D43) is reporting an error or FPGA is failing to read sensor. | "}, + { 36 , ALARM_ID_DD_D17_COND_SENSOR_FPGA_FAULT , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Conductivity Sensor Error | DD conductivity sensor (D17) located before balancing chamber reporting error or not being read successfully | Conductvity Sensor (D17) is reporting an error or FPGA is failing to read sensor. | "}, + { 37 , ALARM_ID_DD_D27_COND_SENSOR_FPGA_FAULT , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Conductivity Sensor Error | DD conductivity sensor (D27) located after balancing chamber reporting error or not being read successfully | Conductvity Sensor (D27) is reporting an error or FPGA is failing to read sensor. | "}, + { 38 , ALARM_ID_DD_D29_COND_SENSOR_FPGA_FAULT , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Conductivity Sensor Error | DD conductivity sensor (D29) located after balancing chamber reporting error or not being read successfully | Conductvity Sensor (D29) is reporting an error or FPGA is failing to read sensor. | "}, + { 39 , ALARM_ID_DD_D43_COND_SENSOR_FPGA_FAULT , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Conductivity Sensor Error | DD conductivity sensor (D43) located in spent dialysate path reporting error or not being read successfully | Conductvity Sensor (D43) is reporting an error or FPGA is failing to read sensor. | "}, { 40 , ALARM_ID_TD_VENOUS_BUBBLE_DETECTED , "TD | Venous Air Detected | Air detected in the venous air trap.\n- Confirm arterial access placement and confirm\nconnections are secure.\n- Treatment must be terminated without rinseback. | Blood Stop: Venous Air Bubble | TD venous air bubble detected alarm | If bubbles are detected during the treatment. | "}, { 41 , ALARM_ID_TD_CARTRIDGE_DOOR_OPENED , "TD | Blood Pump Stopped: Door Open | The device door is open.\n Close the device door. \n- Press Resume/OK to continue. | Blood Stop: Door Open | TD cartridge door opened alarm | If the cartridge door is opened as reported by the FPGA. | "}, { 42 , ALARM_ID_TD_STUCK_BUTTON_TEST_FAILED , "TD | Stuck Button | A button was pressed during start up and appears stuck. \n- Press and release the power and stop buttons on the\nfront of the device.\n- Turn device off, then back on.\n- If problem continues, call service to schedule a repair. | TD POST: Stuck Button | TD Stuck button POST failure. Stop or Off button detected to be pressed for at least 1 second during test shortly after power up | Stop or Off button are pressed for more than 1 sec during POST. | "}, @@ -526,10 +526,10 @@ { 50 , ALARM_ID_TD_VENOUS_PRESSURE_SELF_TEST_FAILURE , "TD | Self Test: Cartridge | A cartridge was detected during self- test.\n- Open door\n- Remove cartridge\n- Close door\n- Press Resume/OK | Process: Pre- Tx Remove Cartridge | TD venous pressure self-test failure alarm | If the venous pressure < -100 or > 600 mmHg | "}, // Num Alarm ID Device Display Title Instructions Alarm List Title Description Trigger Condition { 51 , ALARM_ID_TD_PRE_TREATMENT_DRY_PRESSURE_TEST_FAILURE , "TD | Self Test: Cartridge Check | A problem has been detected during self- test. \n- Confirm cartridge is properly installed without kinks or clamps.\n- Ensure all connections are secure. \n- Press Resume/OK to resume. \n- If problem persists, install a new cartridge. | Process: Pre- Tx Dry Self- Test | TD pre-treatment mode dry self-tests failure | If the following dry self tests have failed: 1. handleDrySelfTestPressureSensorsState 2. handleDrySelfTestPressureSensorsNormalState | "}, - { 52 , ALARM_ID_DD_D11_PUMP_SPEED_CONTROL_ERROR , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: D11_PUMP Speed Fault | DD concentrate pump D11_PUMP speed control error | If the concentrate pump D11_PUMP speed > 0.02 tolerance for a certain period of time. | "}, - { 53 , ALARM_ID_DD_D10_PUMP_SPEED_CONTROL_ERROR , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: D10_PUMP Speed Fault | DD concentrate pump D10_PUMP speed control error | If the concentrate pump D11_PUMP speed > 0.02 tolerance for a certain period of time. | "}, - { 54 , ALARM_ID_DD_CONC_PUMP_HALL_SENSOR_OUT_OF_RANGE , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: D11_PUMP D10_PUMP Speed Range | DD concentrate pump hall sensor out of range. | Hall sensor pulse to pulse timing is < 1ms (speed too fast). | "}, - { 55 , ALARM_ID_DD_CONCENTRATE_PUMP_FAULT , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: D11_PUMP/D10_PUMP Fault | DD concentrate pump fault | If the concentrate pumps fault is reported by the FPGA. | "}, + { 52 , ALARM_ID_DD_D11_PUMP_SPEED_CONTROL_ERROR , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: PUMP Speed Fault | DD concentrate pump D11_PUMP speed control error | If the concentrate pump D11_PUMP speed > 0.02 tolerance for a certain period of time. | "}, + { 53 , ALARM_ID_DD_D10_PUMP_SPEED_CONTROL_ERROR , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: PUMP Speed Fault | DD concentrate pump D10_PUMP speed control error | If the concentrate pump D11_PUMP speed > 0.02 tolerance for a certain period of time. | "}, + { 54 , ALARM_ID_DD_CONC_PUMP_HALL_SENSOR_OUT_OF_RANGE , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: PUMP Speed Fault | DD concentrate pump hall sensor out of range. | Hall sensor pulse to pulse timing is < 1ms (speed too fast). | "}, + { 55 , ALARM_ID_DD_CONCENTRATE_PUMP_FAULT , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: PUMP Fault | DD concentrate pump fault | If the concentrate pumps fault is reported by the FPGA. | "}, { 56 , ALARM_ID_DD_TEMPERATURE_SENSOR_OUT_OF_RANGE , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: Temp Sensor Range | DD temperature sensor out of range | If the measured temperature < 0 C or > 80 C for the non-fluid inline sensors and > 120 C for the fluid sensors. | "}, { 57 , ALARM_ID_DD_RTD_SENSORS_FPGA_FAULT , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: Temperature RTD Comm | DD temperature sensors not being read successfully | If the FPGA does not report fresh data for a certain period of time | "}, { 58 , ALARM_ID_DD_BARO_SENSOR_FPGA_FAULT , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: Baro temp Comm | DD barometric sensor not being read successfully | If the FPGA does not report fresh data for a certain period of time | "}, @@ -544,8 +544,8 @@ { 67 , ALARM_ID_DD_D48_PUMP_CURRENT_OUT_OF_RANGE , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: Spent Dial Pump Current | DD spent dialysate pump current out of range | If the spent dialysate pump current > 2.2 A when it is running or if the current > 0.1 A when it is off. | "}, { 68 , ALARM_ID_DD_D48_PUMP_DIRECTION_INVALID , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: Spent Dial Pump Dir Fault | DD spent dialysate pump invalid direction | If the spent dialysate pump's direction is not in forward direction. | "}, { 69 , ALARM_ID_DD_D48_PUMP_DIRECTION_FPGA_FAULT , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: Spent Dial pump Dir Err Cnt | DD spent dialysate pump direction FPGA fault | If the FPGA indicates an direction error for a certain period of time. | "}, - { 70 , ALARM_ID_DD_FLUID_TOO_LOW_WHILE_D5_HEAT_IS_ON , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: D5 On With No Fluid | DD primary heater is On with low fluid level | If the Hydraulics floater sensing low level fluid for a certain period of time. | "}, - { 71 , ALARM_ID_DD_FLUID_TOO_LOW_WHILE_D45_HEAT_IS_ON , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: D45 On With No Fluid | DD trimmer heater is On with low fluid level | If the Balance air separation chamber fluid level is low for a certain period of time. | "}, + { 70 , ALARM_ID_DD_FLUID_TOO_LOW_WHILE_D5_HEAT_IS_ON , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: Heater On With No Fluid | DD D5 heater is On with low fluid level | If the Hydraulics floater sensing low level fluid for a certain period of time. | "}, + { 71 , ALARM_ID_DD_FLUID_TOO_LOW_WHILE_D45_HEAT_IS_ON , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Fault: Heater On With No Fluid | DD D45 heater is On with low fluid level | If the Balance air separation chamber fluid level is low for a certain period of time. | "}, { 72 , ALARM_ID_DD_WATER_INLET_INPUT_PRES_TIMEOUT_FAULT , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Water Inlet Input Press Sensor Error | DD water inlet input pressure sensor reporting error or not being read successfully | Water Inlet Input pressure sensor is reporting an error or FPGA is failing to read sensor. | "}, { 73 , ALARM_ID_DD_WATER_INLET_OUTPUT_PRES_TIMEOUT_FAULT , "DD | Service Required: Dialysate Device | A problem was detected with the dialysate device. \n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. | DD Water Inlet Output Press Sensor Error | DD water inlet output pressure sensor reporting error or not being read successfully | Water Inlet Output pressure sensor is reporting an error or FPGA is failing to read sensor. | "}, { 74 , ALARM_ID_DD_INLET_WATER_PRESSURE_IN_LOW_RANGE , "DD | Incoming Water Pressure Input Low | Water pressure of incoming water is low.\n- Treatment is currently paused.\n- Check the incoming water supply.\n- When the pressure is within target range\nResume/OK button will be available.\n- Press Resume/OK to continue treatment. | DD: Inlet Water Pres Input Low | DD Inlet water pressure input too low | Inlet water pressure input < 25 PSI. | "},